Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness

被引:20
|
作者
Li, Xuepei [1 ]
Zhou, Yijing [1 ]
Young, Charlotte Aimee [2 ]
Chen, Aiming [3 ]
Jin, Guangming [1 ]
Zheng, Danying [1 ]
机构
[1] Sun Yat Sen Univ, State Key Lab Ophthalmol, Zhongshan Ophthalm Ctr, Guangzhou, Peoples R China
[2] Nanchang Univ, Affiliated Hosp 3, Dept Ophthalmol, Nanchang, Jiangxi, Peoples R China
[3] Sun Yat Sen Univ, Affiliated Hosp 5, Zhuhai, Peoples R China
基金
中国国家自然科学基金;
关键词
Anterior chamber depth (ACD); corneal thickness; swept source anterior segment optical coherence tomography; Oculus Pentacam; FOURIER-DOMAIN; SCHEIMPFLUG; ULTRASOUND; REPRODUCIBILITY; REPEATABILITY; KERATOCONUS; PARAMETERS; CURVATURE; AGREEMENT; BIOMETRY;
D O I
10.21037/atm-20-187
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Accurate measurements of anterior chamber depth (ACD) and regional corneal thickness are especially important for the diagnosis and treatment of many ocular disease. This study aimed to evaluate the repeatability and reliability of a new swept source anterior segment optical coherence tomography (AS-OCT) and its agreement with Oculus Pentacam for measurements of ACD and corneal thickness (CT). Methods: The central corneal thickness (CCT), superior corneal thickness (SCT), inferior corneal thickness (ICT), nasal corneal thickness (NCT), temporal corneal thickness (TCT) and ACD of the right eye from forty-nine adults aged 18 to 36 years (24.78 +/- 4.36 years old) were measured with Pentacam and AS-OCT (CASIA2). All subjects were measured twice with each device. The repeatability was determined using the coefficient of repeatability (COR), the relative COR and the limits of agreement (LOA). Bland-Altman plot was also used for evaluating the agreement between parameters from the two devices. Results: For the repeatability of CASIA2, the COR of the two measurements of ACD, CCT, SCT, ICT, NCT and TCT was 0.31 mm, 18.58, 25.83, 28.32, 26.71 and 22.09 mu m respectively. There were no statistically significant differences between the CT and ACD measurements by CASIA2 (P>0.05). For measurements with Pentacam and CASIA2, the COR of ACD, CCT, SCT, NCT, ICT and TCT was 0.294 mm, 13.10, 51.57, 48.06, 56.21 and 47.69 mu m respectively. No significant differences were found between the values measured by CAISA2 and Pentacam for CT and ACD (P>0.05). The Bland-Altman analysis also suggested high consistency between measurements obtained by Pentacam and CASIA2. Conclusions: Our results suggest that Pentacam and CASIA2 have good agreement in CT and ACD measurements. The two devices can be considered interchangeable for these parameters' measurements in healthy subjects when monitoring corneal conditions or planning ocular surgery. However, subtle differences between CASIA2 and Pentacam should also be kept in mind for certain specific clinical or research purposes.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Comparison of Oculus Pentacam and Stratus Optical Coherence Tomography for Measurement of Central Corneal Thickness
    Ceylan, Osman Melih
    Turk, Adem
    Erdurman, Cuneyt
    Mumcuoglu, Tarkan
    Erdem, Uzeyir
    Gokce, Gokcen
    Dagli, Selim
    [J]. CORNEA, 2011, 30 (06) : 670 - 674
  • [2] Comparative study of anterior segment measurement with Pentacam and anterior segment optical coherence tomography
    Fu, Jing
    Wang, Xiaozhen
    Li, Shuning
    Wu, Gewei
    Wang, Ningli
    [J]. CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2010, 45 (06): : 627 - 631
  • [3] Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
    Huang, Jinhai
    Lu, Weicong
    Savini, Giacomo
    Chen, Hao
    Wang, Chengfang
    Yu, Xinxin
    Bao, Fangjun
    Wang, Qinmei
    [J]. JOURNAL OF OPHTHALMOLOGY, 2016, 2016
  • [4] Comparison of the Scanning Peripheral Anterior Chamber Depth Analyzer and the Anterior Segment Optical Coherence Tomography in the Assessment of Anterior Ocular Segment
    Kashiwagi, K.
    Furuya, T.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [5] Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography
    Lavanya, Raghavan
    Teo, Livia
    Friedman, David S.
    Aung, Han T.
    Baskaran, Mani
    Gao, Hong
    Alfred, Tamuno
    Seah, Steve K.
    Kashiwagi, Kenji
    Foster, Paul J.
    Aung, Tin
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (08) : 1023 - 1026
  • [6] Comparison of Corneal Thickness Measured with Swept Source Anterior Segment Optical Coherence Tomographer, Pentacam and Ultrasound
    Haines, Lacey
    Carter, Kirstie
    Marschall, Sebastian
    Gawish, Ahmed
    Wong, Alex
    Fieguth, Paul
    Sorbara, Luigina
    Bizheva, Kostadinka K.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [7] Assessment and reproducibility of anterior chamber depth measurement with anterior segment optical coherence tomography compared with immersion ultrasonography
    Nemeth, Gabor
    Vajas, Attila
    Tsorbatzoglou, Alexis
    Kolozsvari, Bence
    Modis, Laszlo, Jr.
    Berta, Andras
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (03): : 443 - 447
  • [8] Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography
    Li, Haitao
    Leung, Christopher Kai Shun
    Cheung, Carol Yim Lui
    Wong, Lee
    Pang, Chi Pui
    Weinreb, Robert Neal
    Lam, Dennis Shun Chiu
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (11) : 1490 - 1492
  • [9] Comparison of corneal thickness in patients with dry eye disease using the Pentacam rotating Scheimpflug camera and anterior segment optical coherence tomography
    Fujimoto, Keiichi
    Inomata, Takenori
    Okumura, Yuichi
    Iwata, Nanami
    Fujio, Kenta
    Eguchi, Atsuko
    Nagino, Ken
    Shokirova, Hurramhon
    Karasawa, Maria
    Murakami, Akira
    [J]. PLOS ONE, 2020, 15 (02):
  • [10] Cycloplegic Anterior Chamber Depth Measurements Using Anterior Segment Optical Coherence Tomography (VisanteTM)
    Aschbrenner, M. W.
    Moss, H. B.
    Lyles, G. W.
    Cohen, K. L.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)