FLOW CYTOMETRIC ANALYSIS OF WHOLE-BLOOD LYSIS, 3 ANTICOAGULANTS, AND 5 CELL PREPARATIONS

被引:51
|
作者
CARTER, PH
RESTORUIZ, S
WASHINGTON, GC
ETHRIDGE, S
PALINI, A
VOGT, R
WAXDAL, M
FLEISHER, T
NOGUCHI, PD
MARTI, GE
机构
[1] CTR DIS CONTROL,DIV ENVIRONM HLTH LAB SCI,ATLANTA,GA 30333
[2] FAST SYST INC,GAITHERSBURG,MD 20877
[3] NIH,CC,CPD,CLIN IMMUNOL SERV,BETHESDA,MD 20892
来源
CYTOMETRY | 1992年 / 13卷 / 01期
关键词
ETHYLENEDIAMINETETRACETIC ACID (EDTA); HEPARIN; ACD; ACID CITRATE DEXTROSE; IMMUNOPHENOTYPING; CELL VIABILITY;
D O I
10.1002/cyto.990130111
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We studied the effects of anticoagulants and cell preparation methods on lymphocyte forward-angle scatter (FSC), autofluorescence, and immunofluorescent staining for CD45, CD14, and CD13. Blood samples collected in ethylenediaminetetraacetic acid (EDTA), heparin, and acid citrate dextrose (ACD) were processed by using conventional Hypaque-Ficoll (HF) separation and four whole blood (WB) lysis techniques: Immuno-lyse, Q-Prep, FACS Lyse, and Gen Trak Lysis. Lymphocytes prepared by using three of the four whole blood methods gave FCS values comparable to those isolated by HF, while one method (FACS Lyse) gave consistently lower values. Autofluorescence values were comparable by all methods except Immuno-lyse, which showed consistently higher values in blood stored for 24 h with any anticoagulant. Immunofluorescent values for CD45-stained cells were quite consistent across all methods, and among the whole blood methods, FACS Lyse and Q-Prep uniformly gave the highest purity of CD45-positive cells in the lymphocyte light scatter gates. Additionally, propidium iodide (PI) analyses of CD45-stained whole blood, and analyzed without lysis, confirmed that ACD and heparin were superior to EDTA for maintaining viable leucocytes overnight. Future studies should focus on other commonly used reagents, a wide variety of abnormal samples, and cell viability.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 50 条