OPTIMIZATION OF BRACHYTHERAPY DOSE DISTRIBUTIONS BY SIMULATED ANNEALING

被引:47
|
作者
SLOBODA, RS
机构
[1] Medical Physics Department, Cross Cancer Institute, Edmonton, Alberta, T6G 1Z2
关键词
BRACHYTHERAPY; OPTIMIZATION; REMOTE AFTERLOADING; SIMULATED ANNEALING; TREATMENT PLANNING;
D O I
10.1118/1.596783
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An algorithm based on the method of simulated annealing is presented for optimizing brachytherapy dose distributions. The algorithm accommodates either static configurations of multiple sources or single stepping sources, hence in principle can be used to optimize both low- and high-dose rate treatments delivered with remote afterloading equipment. Required inputs include the specification of target dose rates and dose rate limits, expressed in absolute or relative terms, at operator selected points near the treatment site. The influence of the dose rate limits can be adjusted continuously through the use of one or more penalty factors. The algorithm generates a set of integer weights, one for each available source position, which are interpreted in terms of configuration occupancy numbers for static source arrangements and relative dwell times for stepping sources. Application is made to several variations of a hypothetical low-dose rate vaginal vault planning problem involving one rectal and six applicator calculation points. The algorithm's performance for different source strengths, annealing schedules, target dose rates, dose rate limits, and values of a single penalty factor-lambda was examined. With a simple annealing schedule and value of lambda = 25, the algorithm found solutions of high quality for all problem variants. The CPU time required for optimization on a Vax 11/750 computer ranged from 2 min for a single configuration to 25 min for a solution consisting of four configurations. These results support the use of simulated annealing for clinical planning of low dose rate vaginal treatments, and encourage investigation of other applications in brachytherapy.
引用
收藏
页码:955 / 964
页数:10
相关论文
共 50 条
  • [21] OPTIMIZATION USING SIMULATED ANNEALING
    BROOKS, SP
    MORGAN, BJT
    [J]. STATISTICIAN, 1995, 44 (02): : 241 - 257
  • [22] Dynamic optimization with simulated annealing
    Faber, R
    Jockenhövel, T
    Tsatsaronis, G
    [J]. COMPUTERS & CHEMICAL ENGINEERING, 2005, 29 (02) : 273 - 290
  • [23] Robust optimization with simulated annealing
    Dimitris Bertsimas
    Omid Nohadani
    [J]. Journal of Global Optimization, 2010, 48 : 323 - 334
  • [24] GLOBAL OPTIMIZATION AND SIMULATED ANNEALING
    DEKKERS, A
    AARTS, E
    [J]. MATHEMATICAL PROGRAMMING, 1991, 50 (03) : 367 - 393
  • [25] GEOMETRY OPTIMIZATION BY SIMULATED ANNEALING
    DONNELLY, RA
    [J]. CHEMICAL PHYSICS LETTERS, 1987, 136 (3-4) : 274 - 278
  • [26] Simulated annealing for convex optimization
    Kalai, Adam Tauman
    Vempala, Santosh
    [J]. MATHEMATICS OF OPERATIONS RESEARCH, 2006, 31 (02) : 253 - 266
  • [27] Robust optimization with simulated annealing
    Bertsimas, Dimitris
    Nohadani, Omid
    [J]. JOURNAL OF GLOBAL OPTIMIZATION, 2010, 48 (02) : 323 - 334
  • [28] Topological Optimization and Simulated Annealing
    Goto, Tiago
    Najafabadi, Hossein R.
    Falheiro, Mizael
    Martins, Thiago C.
    Barari, Ahmad
    Tsuzuki, Marcos S. G.
    [J]. IFAC PAPERSONLINE, 2021, 54 (01): : 205 - 210
  • [29] A Simulated Annealing optimization technique to obtain uniform dose distribution in gamma irradiators
    Chakrabarty, Shatabdi
    Vishwakarma, R. S.
    Selvam, T. Palani
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2023, 209
  • [30] Anatomy-based inverse planning simulated annealing optimization in high-dose-rate prostate brachytherapy: Significant dosimetric advantage over other optimization techniques
    Jacob, Dayee
    Raben, Adam
    Sarkar, Abhirup
    Grimm, Jimm
    Simpson, Larry
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (03): : 820 - 827