Effect of light source instability on uniformity of 3D reconstructions from a cone beam optical CT scanner

被引:0
|
作者
J. Begg
M. L. Taylor
L. Holloway
T. Kron
R. D. Franich
机构
[1] RMIT University,School of Applied Sciences and Health Innovations Research Institute
[2] Liverpool and Macarthur Cancer Therapy Centres and Ingham Institute,Institute of Medical Physics, School of Physics
[3] University of Sydney,Centre for Medical Radiation Physics
[4] University of Wollongong,South West Clinical School, School of Medicine
[5] University of New South Wales,Physical Sciences
[6] Peter MacCallum Cancer Centre,undefined
来源
Australasian Physical & Engineering Sciences in Medicine | 2014年 / 37卷
关键词
Optical CT; 3D dosimetry; Gel dosimetry; LED light source; Uncertainty; Error;
D O I
暂无
中图分类号
学科分类号
摘要
Temporally varying light intensity during acquisition of projection images in an optical CT scanner can potentially be misinterpreted as physical properties of the sample. This work investigated the impact of LED light source intensity instability on measured attenuation coefficients. Different scenarios were investigated by conducting one or both of the reference and data scans in a ‘cold’ scanner, where the light source intensity had not yet stabilised. Uniform samples were scanned to assess the impact on measured uniformity. The orange (590 nm) light source decreased in intensity by 29 % over the first 2 h, while the red (633 nm) decreased by 9 %. The rates of change of intensity at 2 h were 0.1 and 0.03 % respectively over a 5 min period—corresponding to the scan duration. The normalisation function of the reconstruction software does not fully account for the intensity differences and discrepancies remain. Attenuation coefficient inaccuracies of up to 8 % were observed for data reconstructed from projection images acquired with a cold scanner. Increased noise was observed for most cases where one or both of the scans was acquired without sufficient warm-up. The decrease in accuracy and increase in noise were most apparent for data reconstructed from reference and data scans acquired with a cold scanner on different days.
引用
收藏
页码:791 / 798
页数:7
相关论文
共 50 条
  • [41] Imaging Study of Pseudo-CT Synthesized From Cone-Beam CT Based on 3D CycleGAN in Radiotherapy
    Sun, Hongfei
    Fan, Rongbo
    Li, Chunying
    Lu, Zhengda
    Xie, Kai
    Ni, Xinye
    Yang, Jianhua
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [42] Sex determination of 3D facial reconstructions from CT data sets
    Verhoff, M. A.
    Gehl, A.
    Kettner, M.
    Kreutz, K.
    Heinemann, A.
    Ramsthaler, F.
    RECHTSMEDIZIN, 2009, 19 (06) : 441 - 444
  • [43] Studies on circular isocentric cone-beam trajectories for 3D image reconstructions using FDK algorithm
    Soimu, Delia
    Buliev, Ivan
    Pallikarakis, Nicolas
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2008, 32 (03) : 210 - 220
  • [44] Measurements on 3D models of human skulls derived from two different cone beam CT scanners
    van Vlijmen, Olivier J. C.
    Rangel, Frits A.
    Berge, Stefaan J.
    Bronkhorst, Ewald M.
    Becking, Alfred G.
    Kuijpers-Jagtman, Anne Marie
    CLINICAL ORAL INVESTIGATIONS, 2011, 15 (05) : 721 - 727
  • [45] Craniomaxillofacial surgery planning based on 3D models derived from Cone-Beam CT data
    Adolphs, Nicolai
    Liu, Weichen
    Keeve, Erwin
    Hoffmeister, Bodo
    COMPUTER AIDED SURGERY, 2013, 18 (5-6) : 101 - 108
  • [46] Measurements on 3D models of human skulls derived from two different cone beam CT scanners
    Olivier J. C. van Vlijmen
    Frits A. Rangel
    Stefaan J. Bergé
    Ewald M. Bronkhorst
    Alfred G. Becking
    Anne Marie Kuijpers-Jagtman
    Clinical Oral Investigations, 2011, 15 : 721 - 727
  • [47] Validation of conventional 2D lateral cephalometry using 3D cone beam CT
    Shaw, Karen
    McIntyre, Grant
    Mossey, Peter
    Menhinick, Alison
    Thomson, Donald
    JOURNAL OF ORTHODONTICS, 2013, 40 (01) : 22 - 28
  • [48] Quantification of prostate motion based on 3D/3D image registration using daily cone beam CT (CBCT)
    Hammoud, R.
    Kim, J.
    Li, S.
    Patel, S.
    Pradhan, D.
    Chen, Q.
    Wen, N.
    Lord, B.
    Ajlouni, M.
    Movsas, B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S648 - S649
  • [49] A method for mandibular dental arch superimposition using 3D cone beam CT and orthodontic 3D digital model
    Park, Tae-Joon
    Lee, Sang-Hyun
    Lee, Ki-Soo
    KOREAN JOURNAL OF ORTHODONTICS, 2012, 42 (04) : 169 - 181
  • [50] GPU-based 3D cone-beam CT image reconstruction: application to micro CT
    Zhao, Xing
    Bian, Junguo
    Sidky, Emil Y.
    Cho, Seungryong
    Zhang, Peng
    Pan, Xiaochuan
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3922 - +