Effect of light source instability on uniformity of 3D reconstructions from a cone beam optical CT scanner

被引:0
|
作者
J. Begg
M. L. Taylor
L. Holloway
T. Kron
R. D. Franich
机构
[1] RMIT University,School of Applied Sciences and Health Innovations Research Institute
[2] Liverpool and Macarthur Cancer Therapy Centres and Ingham Institute,Institute of Medical Physics, School of Physics
[3] University of Sydney,Centre for Medical Radiation Physics
[4] University of Wollongong,South West Clinical School, School of Medicine
[5] University of New South Wales,Physical Sciences
[6] Peter MacCallum Cancer Centre,undefined
来源
Australasian Physical & Engineering Sciences in Medicine | 2014年 / 37卷
关键词
Optical CT; 3D dosimetry; Gel dosimetry; LED light source; Uncertainty; Error;
D O I
暂无
中图分类号
学科分类号
摘要
Temporally varying light intensity during acquisition of projection images in an optical CT scanner can potentially be misinterpreted as physical properties of the sample. This work investigated the impact of LED light source intensity instability on measured attenuation coefficients. Different scenarios were investigated by conducting one or both of the reference and data scans in a ‘cold’ scanner, where the light source intensity had not yet stabilised. Uniform samples were scanned to assess the impact on measured uniformity. The orange (590 nm) light source decreased in intensity by 29 % over the first 2 h, while the red (633 nm) decreased by 9 %. The rates of change of intensity at 2 h were 0.1 and 0.03 % respectively over a 5 min period—corresponding to the scan duration. The normalisation function of the reconstruction software does not fully account for the intensity differences and discrepancies remain. Attenuation coefficient inaccuracies of up to 8 % were observed for data reconstructed from projection images acquired with a cold scanner. Increased noise was observed for most cases where one or both of the scans was acquired without sufficient warm-up. The decrease in accuracy and increase in noise were most apparent for data reconstructed from reference and data scans acquired with a cold scanner on different days.
引用
收藏
页码:791 / 798
页数:7
相关论文
共 50 条
  • [1] Effect of light source instability on uniformity of 3D reconstructions from a cone beam optical CT scanner
    Begg, J.
    Taylor, M. L.
    Holloway, L.
    Kron, T.
    Franich, R. D.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (04) : 791 - 798
  • [2] Cone beam optical CT scanner for 3D dosimetry
    Miller, J
    Adamovics, J
    Dietrich, J
    MEDICAL PHYSICS, 2005, 32 (06) : 2138 - 2138
  • [3] Effect of voxel size on the accuracy of 3D reconstructions with cone beam CT
    Maret, D.
    Telmon, N.
    Peters, O. A.
    Lepage, B.
    Treil, J.
    Inglese, J. M.
    Peyre, A.
    Kahn, J. L.
    Sixou, M.
    DENTOMAXILLOFACIAL RADIOLOGY, 2012, 41 (08) : 649 - 655
  • [4] DEVELOPMENT OF A 3D CT SCANNER WITH A CONE-BEAM AND VIDEOFLUOROSCOPIC SYSTEM
    ENDO, M
    SATOH, K
    OKAZAKI, T
    HATTORI, Y
    KOBAYASHI, S
    JIMBO, M
    RADIOLOGY, 1995, 197 : 292 - 292
  • [5] Study on reconstructed image of 3D cone-beam CT scanner
    Nakamori, N
    Sudou, T
    Kanamori, H
    Endou, M
    Kusakabe, M
    PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 : 696 - 705
  • [6] A prototype fan-beam optical CT scanner for 3D dosimetry
    Campbell, Warren G.
    Rudko, D. A.
    Braam, Nicolas A.
    Wells, Derek M.
    Jirasek, Andrew
    MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Accuracy Comparison of the Optical 3D Scanner and CT Scanner
    Mendricky, Radomir
    Sobotka, Jiri
    MANUFACTURING TECHNOLOGY, 2020, 20 (06): : 791 - 801
  • [8] Accuracy of 3D Reconstructions Based on Cone Beam Computed Tomography
    Maret, D.
    Molinier, F.
    Braga, J.
    Peters, O. A.
    Telmon, N.
    Treil, J.
    Inglese, J. M.
    Cossie, A.
    Kahn, J. L.
    Sixou, M.
    JOURNAL OF DENTAL RESEARCH, 2010, 89 (12) : 1465 - 1469
  • [9] Projection-based improvement of 3D reconstructions from motion-impaired dental cone beam CT data
    Niebler, Stefan
    Schoemer, Elmar
    Tjaden, Henning
    Schwanecke, Ulrich
    Schulze, Ralf
    MEDICAL PHYSICS, 2019, 46 (10) : 4470 - 4480
  • [10] Cervical vertebrae maturation index estimates on cone beam CT: 3D reconstructions vs sagittal sections
    Bonfim, Marco A. E.
    Costa, Andre L. F.
    Fuziy, Acacio
    Ximenez, Michel E. L.
    Cotrim-Ferreira, Flavio A.
    Ferreira-Santos, Rivea I.
    DENTOMAXILLOFACIAL RADIOLOGY, 2016, 45 (01)