Biomechanical changes in the proximal femur before and after removal of femoral neck system

被引:0
|
作者
Nan, Chong [1 ]
Liu, Yuxiu [1 ]
Zhang, Di [2 ]
Qin, Yazhuo [1 ]
Yu, Hetong [1 ]
Ma, Zhanbei [1 ]
机构
[1] Baoding 1 Cent Hosp, Dept Orthoped, Baoding 071000, Hebei, Peoples R China
[2] Hebei Med Univ, Hosp 3, Dept Spinal Surg, Shijiazhuang 050000, Hebei, Peoples R China
来源
关键词
Femoral neck system; Cannulated screw removal; Biomechanical effect; Finite element analysis; FINITE-ELEMENT-ANALYSIS; FRACTURE;
D O I
10.1186/s13018-024-04769-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background As an innovative internal fixation system, FNS (femoral neck system) is increasingly being utilized by surgeons for the treatment of femoral neck fractures. At present, there have been numerous finite element analysis experiments studying the immediate stability of FNS and CSS in treating femoral neck fractures. However, there is scarce mechanical analysis available regarding the effects post internal fixation removal. This study aimed to investigate the alterations in mechanical parameters of the proximal femur before and after the removal of FNS (femoral neck system), and to assess potential distinctions in indicators following the extraction of CSS (Cannulated Screws).Methods A proximal femur model was reconstructed using finite element numerical techniques. The models for CSS and FNS were formulated utilizing characteristics and parametric definitions. The internal fixation was combined with a normal proximal femur model to simulate the healing state after fracture surgery. Within the framework of static analysis, consistent stress burdens were applied across the entirety of the models. The total deformation and equivalent stress of the proximal femur were recorded before and after the removal of internal fixation.Results Under the standing condition, the total deformation of the model before and after removing CSS was 0.99 mm and 1.10 mm, respectively, indicating an increase of 12%. The total deformation of the model before and after removing FNS was 0.65 mm and 0.76 mm, respectively, indicating an increase of 17%. The equivalent stress for CSS and FNS were 55.21 MPa and 250.67 MPa, respectively. The average equivalent stress on the cross-section of the femoral neck before and after removal of CSS was 7.76 MPa and 6.11 MPa, respectively. The average equivalent stress on the cross-section of the femoral neck before and after removal of FNS was 9.89 MPa and 8.79 MPa, respectively.Conclusions The retention of internal fixation may contribute to improved stability of the proximal femur. However, there still existed risks of stress concentration in internal fixation and stress shielding in the proximal femur. Compared to CSS, the removal of FNS results in larger bone tunnels and insufficient model stability. Further clinical interventions are recommended to address this issue.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] A Controlled Variable Study of the Biomechanical Properties of the Proximal Femur before and after Cancellous Bone Removal
    Wang, Haicheng
    Ding, Kai
    Zhang, Yifan
    Ren, Chuan
    Huo, Haoyu
    Zhu, Yanbin
    Zhang, Qi
    Chen, Wei
    ORTHOPAEDIC SURGERY, 2024, 16 (05) : 1215 - 1229
  • [2] The Analysis of Biomechanical Properties of Proximal Femur after Implant Removal
    Yang, Jae Hyuk
    Jung, Tae Gon
    Honnurappa, Arjun Rupanagudi
    Cha, Jae Min
    Ham, Chang Hwa
    Kim, Tae Yoon
    Suh, Seung Woo
    APPLIED BIONICS AND BIOMECHANICS, 2016, 2016
  • [3] Femoral neck fracture after removal of the standard gamma interlocking nail: a cadaveric study to determine factors influencing the biomechanical properties of the proximal femur
    Kukla, C
    Pichl, W
    Prokesch, R
    Jacyniak, W
    Heinze, G
    Gatterer, R
    Heinz, T
    JOURNAL OF BIOMECHANICS, 2001, 34 (12) : 1519 - 1526
  • [4] Femoral neck lengthening osteotomy after growth disturbance of the proximal femur
    Hasler, CC
    Morscher, EW
    JOURNAL OF PEDIATRIC ORTHOPAEDICS-PART B, 1999, 8 (04): : 271 - 275
  • [5] Impact of femoral neck system removal after femoral neck fracture healing on biomechanical stability and screw stripping risk
    Lee, Se-Won
    Pak, Jeongah
    Lim, Dohyung
    BIOMEDICAL ENGINEERING LETTERS, 2025, 15 (02) : 349 - 355
  • [6] Photoelastic and thermoelastic measurement of stress on the proximal femur before and after implantation of a hip prosthesis with retention of the femoral neck
    Refior, HJ
    Schidlo, C
    Plitz, W
    Heining, S
    ORTHOPEDICS, 2002, 25 (05) : 505 - 511
  • [7] Biomechanical Study of Femoral Bone after Proximal Bone Tumor Removal
    Aroonjarattham, Panya
    Kijthaworn, Peeranat
    Jamda, Promchai
    Supamongkol, Tanakorn
    Suwanboriboon, Samart
    Aroonjarattham, Kitti
    Somtua, Chompunut
    9TH THAI SOCIETY OF MECHANICAL ENGINEERS, INTERNATIONAL CONFERENCE ON MECHANICAL ENGINEERING (TSME-ICOME 2018), 2019, 501
  • [8] INTRAMEDULLARY PRESSURE CHANGES IN PROXIMAL FEMUR BEFORE AND AFTER TREATMENT OF CONGENITAL DISLOCATION OF THE HIP - EVALUATION OF THE DEVELOPMENT OF THE FEMORAL-HEAD
    MA, RX
    JI, SJ
    ZHOU, YD
    LIU, WD
    CHINESE MEDICAL JOURNAL, 1991, 104 (11) : 945 - 948
  • [9] The Analysis of Biomechanical Properties of Proximal Femur after Implant Removal (vol 2016, 4987831, 2016)
    Yang, Jae Hyuk
    Jung, Tae Gon
    Honnurappa, Arjun Rupanagudi
    Cha, Jae Min
    Ham, Chang Hwa
    Kim, Tae Yoon
    Suh, Seung Woo
    APPLIED BIONICS AND BIOMECHANICS, 2020, 2020
  • [10] Patterns of stress distribution at the proximal femur after implantation of a modular neck prosthesis. A biomechanical study
    Politis, Angelos N.
    Siogkas, George K.
    Gelalis, Ioannis D.
    Xenakis, Theodore A.
    CLINICAL BIOMECHANICS, 2013, 28 (04) : 415 - 422