Real time noninvasive estimation of work of breathing using facemask leak-corrected tidal volume during noninvasive pressure support: validation study

被引:0
|
作者
Michael J. Banner
Carl G. Tams
Neil R. Euliano
Paul J. Stephan
Trevor J. Leavitt
A. Daniel Martin
Nawar Al-Rawas
Andrea Gabrielli
机构
[1] University of Florida College of Medicine,Department of Anesthesiology
[2] University of Florida,Department of Physical Therapy, College of Public Health and Health Professions
[3] Santa Fe College,undefined
[4] Convergent Engineering,undefined
关键词
Work of breathing; Noninvasive; Facemask; Real-time corrections; Tidal volume;
D O I
暂无
中图分类号
学科分类号
摘要
We describe a real time, noninvasive method of estimating work of breathing (esophageal balloon not required) during noninvasive pressure support (PS) that uses an artificial neural network (ANN) combined with a leak correction (LC) algorithm, programmed to ignore asynchronous breaths, that corrects for differences in inhaled and exhaled tidal volume (VT) from facemask leaks (WOBANN,LC/min). Validation studies of WOBANN,LC/min were performed. Using a dedicated and popular noninvasive ventilation ventilator (V60, Philips), in vitro studies using PS (5 and 10 cm H2O) at various inspiratory flow rate demands were simulated with a lung model. WOBANN,LC/min was compared with the actual work of breathing, determined under conditions of no facemask leaks and estimated using an ANN (WOBANN/min). Using the same ventilator, an in vivo study of healthy adults (n = 8) receiving combinations of PS (3–10 cm H2O) and expiratory positive airway pressure was done. WOBANN,LC/min was compared with physiologic work of breathing/min (WOBPHYS/min), determined from changes in esophageal pressure and VT applied to a Campbell diagram. For the in vitro studies, WOBANN,LC/min and WOBANN/min ranged from 2.4 to 11.9 J/min and there was an excellent relationship between WOBANN,LC/breath and WOBANN/breath, r = 0.99, r2 = 0.98 (p < 0.01). There were essentially no differences between WOBANN,LC/min and WOBANN/min. For the in vivo study, WOBANN,LC/min and WOBPHYS/min ranged from 3 to 12 J/min and there was an excellent relationship between WOBANN,LC/breath and WOBPHYS/breath, r = 0.93, r2 = 0.86 (p < 0.01). An ANN combined with a facemask LC algorithm provides noninvasive and valid estimates of work of breathing during noninvasive PS. WOBANN,LC/min, automatically and continuously estimated, may be useful for assessing inspiratory muscle loads and guiding noninvasive PS settings as in a decision support system to appropriately unload inspiratory muscles.
引用
收藏
页码:285 / 294
页数:9
相关论文
共 22 条
  • [21] Real-time Noninvasive Monitoring of Intracranial Fluid Shifts During Dialysis Using Volumetric Integral Phase-Shift Spectroscopy (VIPS): A Proof-of-Concept Study
    Chethan P. Venkatasubba Rao
    Eric M. Bershad
    Eusebia Calvillo
    Nelson Maldonado
    Rahul Damani
    Sreedhar Mandayam
    Jose I. Suarez
    Neurocritical Care, 2018, 28 : 117 - 126
  • [22] BEDSIDE MEASUREMENT OF THE 3RD VENTRICLES DIAMETER DURING EPISODES OF ARISING INTRACRANIAL-PRESSURE AFTER HEAD TRAUMA - USING TRANSCRANIAL REAL-TIME SONOGRAPHY FOR A NONINVASIVE EXAMINATION OF INTRACRANIAL COMPENSATION MECHANISMS
    MURSCH, K
    VOGELSLANG, JP
    ZIMMERER, B
    LUDWIG, HC
    BEHNKE, J
    MARKAKIS, E
    ACTA NEUROCHIRURGICA, 1995, 137 (1-2) : 19 - 24