In vivo dosimetry using MOSkin detector during Cobalt-60 high-dose-rate (HDR) brachytherapy of skin cancer

被引:0
|
作者
Z. Jamalludin
W. L. Jong
G. F. Ho
A. B. Rosenfeld
N. M. Ung
机构
[1] University of Malaya Medical Centre,Medical Physics Unit
[2] University of Malaya Medical Centre,Department of Clinical Oncology
[3] University of Malaya,Clinical Oncology Unit, Faculty of Medicine
[4] University of Wollongong,Centre for Medical Radiation Physics
关键词
Mo; In vivo dosimetry; Skin dose; Co-60 HDR brachytherapy;
D O I
暂无
中图分类号
学科分类号
摘要
The MOSkin, a metal–oxide semiconductor field-effect transistor based detector, is suitable for evaluating skin dose due to its water equivalent depth (WED) of 0.07 mm. This study evaluates doses received by target area and unavoidable normal skin during a the case of skin brachytherapy. The MOSkin was evaluated for its feasibility as detector of choice for in vivo dosimetry during skin brachytherapy. A high-dose rate Cobalt-60 brachytherapy source was administered to the tumour located at the medial aspect of the right arm, complicated with huge lymphedema thus limiting the arm motion. The source was positioned in the middle of patients’ right arm with supine, hands down position. A 5 mm lead and 5 mm bolus were sandwiched between the medial aspect of the arm and lateral chest to reduce skin dose to the chest. Two calibrated MOSkin detectors were placed on the target and normal skin area for five treatment sessions for in vivo dose monitoring. The mean dose to the target area ranged between 19.9 and 21.1 Gy and was higher in comparison with the calculated dose due to contribution of backscattered dose from lead. The mean measured dose at normal skin chest area was 1.6 Gy (1.3–1.9 Gy), less than 2 Gy per fraction. Total dose in EQD2 received by chest skin was much lower than the recommended skin tolerance. The MOSkin detector presents a reliable real-time dose measurement. This study has confirmed the applicability of the MOSkin detector in monitoring skin dose during brachytherapy treatment due to its small sensitive volume and WED 0.07 mm.
引用
收藏
页码:1099 / 1107
页数:8
相关论文
共 50 条
  • [1] In vivo dosimetry using MOSkin detector during Cobalt-60 high-dose-rate (HDR) brachytherapy of skin cancer
    Jamalludin, Z.
    Jong, W. L.
    Ho, G. F.
    Rosenfeld, A. B.
    Ung, N. M.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (04) : 1099 - 1107
  • [2] Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
    Jamalludin, Zulaikha
    Jong, Wei Loong
    Malik, Rozita Abdul
    Rosenfeld, Anatoly
    Ung, Ngie Min
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 58 : 1 - 7
  • [3] Dosimetric impact of applicator displacement during high dose rate (HDR) Cobalt-60 brachytherapy for cervical cancer: A planning study
    Yong, J. S.
    Ung, N. M.
    Jamalludin, Z.
    Malik, R. A.
    Wong, J. H. D.
    Liew, Y. M.
    Ng, K. H.
    RADIATION PHYSICS AND CHEMISTRY, 2016, 119 : 264 - 271
  • [4] Comparison of planned and measured rectal dose in-vivo during high dose rate Cobalt-60 brachytherapy of cervical cancer
    Zaman, Z. K.
    Ung, N. M.
    Malik, R. A.
    Ho, G. F.
    Phua, V. C. E.
    Jamalludin, Z.
    Baharuldin, M. T. H.
    Ng, K. H.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (08): : 980 - 984
  • [5] In vivo dosimetry of skin dose during HDR breast brachytherapy using balloon catheter
    Kim, S.
    Oh, S.
    Gales, A.
    Vallow, L. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S524 - S525
  • [6] Comparing Iridium-192 with Cobalt-60 sources in high-dose-rate brachytherapy boost for localized prostate cancer
    Tamihardja, Joerg
    Weick, Stefan
    Lutyj, Paul
    Zimmermann, Marcus
    Bratengeier, Klaus
    Flentje, Michael
    Polat, Buelent
    ACTA ONCOLOGICA, 2022, 61 (06) : 714 - 719
  • [7] Commissioning and Implementation of a Cobalt-60 High Dose Rate Brachytherapy Source
    Dysart, Jonathan
    MEDICAL PHYSICS, 2016, 43 (08) : 4934 - 4934
  • [8] Outcomes for High-Dose-Rate (HDR) Brachytherapy in the Treatment of Cervical Cancer
    Chin, M.
    Mm, C.
    Sethi, R.
    Formenti, S. C.
    Jozsef, G.
    Blank, S. V.
    Pothuri, B.
    Boyd, L. R.
    Schiff, P. B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S406 - S406
  • [9] In vivo dosimetry with alanine/electron spin resonance dosimetry to evaluate the urethra dose during high-dose-rate brachytherapy
    Wagner, Daniela
    Herrnann, Markus
    Hille, Andrea
    BRACHYTHERAPY, 2017, 16 (04) : 815 - 821
  • [10] In vivo TLD dosimetry for a high dose rate (HDR) breast brachytherapy using MammoSite® applicator
    Selvaraj, RN
    Wu, A
    King, G
    Mogus, R
    Sonnik, D
    Fitian, ID
    Gerszten, K
    Heron, DE
    Saw, CB