Analysis of Data Measured with a Neutron Detector Based on a Boron-10 and Helium-3 Counter on a Source of Photoneutrons

被引:0
|
作者
Kuznetsov S.P. [1 ]
Meshkov I.V. [1 ]
Potashev S.I. [2 ]
Burmistrov Y.M. [2 ]
Karaevsky S.K. [2 ]
机构
[1] Lebedev Physical Institute, Russian Academy of Sciences, Moscow
[2] Institute for Nuclear Research, Russian Academy of Sciences, Moscow
关键词
D O I
10.3103/S1062873818060175
中图分类号
学科分类号
摘要
Data obtained on the IN-LUE photoneutron source at the Russian Academy of Sciences’ Institute for Nuclear Research are analyzed. The data are acquired using a position-sensitive neutron detector with an active 10B layer and a 3He SNM-18 counter. The efficiency of the 10B detector relative to the 3He counter ranges from 2 to 4% for different modes of accelerator operation. Waveforms of the 3He counter’s pulses are recorded using a built-in L-CARD L-783 multifunctional high speed signal processor with sampling frequencies of up to 3 MHz. The amplitude and time spectra of thermal neutrons are obtained from the waveforms using induced electromagnetic signals from the accelerator dump and neutron pulses. Using a cadmium mask with slits and a position-sensitive detector with a low (<10−7) sensitivity to electromagnetic interference and the gamma-ray background allows the spatial resolution of the10B detector to be determined and the contribution from neutrons with different energies to b estimated. © 2018, Allerton Press, Inc.
引用
收藏
页码:725 / 727
页数:2
相关论文
共 21 条