Optimization of multi-reflection time-of-flight mass analyzer operating in in-trap-lift mode

被引:7
|
作者
Huang W.-X. [1 ]
Tian Y.-L. [1 ,2 ]
Wang Y.-S. [1 ,2 ,3 ]
Wang J.-Y. [1 ]
Zhou X.-H. [1 ]
机构
[1] Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou
[2] School of Physical Sciences, University of Chinese Academy of Sciences, Beijing
[3] School of Nuclear Science and Technology, Lanzhou University, Lanzhou
基金
中国国家自然科学基金;
关键词
Exotic nuclei; Isobaric separation; Mass measurement; Multiple reflection; Time-of-flight mass spectrometer;
D O I
10.1007/s41605-017-0031-1
中图分类号
学科分类号
摘要
Purpose: We are building an MRTOF-MS (multi-reflection time-of-flight mass spectrometer) for isobaric separation for the Lanzhou Penning Trap. The potentials applied on the electrodes of our MRTOF mass analyzer operating in in-trap-lift mode have to be optimized to achieve a very high mass resolving power. Methods: Our method to design and optimize an MRTOF mass analyzer has been updated to introduce constraints on the potentials, and this method now can be used to optimize the parameters of MRTOF-MS both operating in mirror-switching mode and in in-trap-lift mode. By using this method, the optimal potential parameters of the electrodes have been obtained for our MRTOF mass analyzer operating in the in-trap-lift mode. Results and conclusion: With a beam size of 2.8 mm diameter and an initial average ion kinetic energy of 1500 eV, the maximal mass resolving power has been achieved to be 3.2 × 10 4 with a total TOF of 7.0 ms for an ion species of 40Ar1 +. It can reach up to 5.6 × 10 4 for a beam size of 0.3 mm diameter. The simulation shows that the inaccuracy of the potentials applied on the outermost mirror electrodes M1–M2 must be less than 50 ppm or preferably 20 ppm. © 2017, Institute of High Energy Physics, Chinese Academy of Sciences; China Nuclear Electronics and Nuclear Detection Society and Springer Nature Singapore Pte Ltd.
引用
收藏
相关论文
共 50 条
  • [21] A multi-reflection time-of-flight mass spectrometer for the offline ion source of the PUMA experiment
    Schlaich, M.
    Fischer, J.
    Fischer, P.
    Klink, C.
    Obertelli, A.
    Schmidt, A.
    Schweikhard, L.
    Wienholtz, F.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2024, 495
  • [22] Construction and simulation of a multi-reflection time-of-flight mass spectrometer at the University of Notre Dame
    Schultz, B. E.
    Kelly, J. M.
    Nicoloff, C.
    Long, J.
    Ryan, S.
    Brodeur, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2016, 376 : 251 - 255
  • [23] A multi-reflection time-of-flight mass spectrometer for mass measurements of short-lived nuclei
    Ishida, Y
    Wada, M
    Wollnik, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 241 (1-4): : 983 - 985
  • [24] Towards ultrahigh-resolution multi-reflection time-of-flight mass spectrometry at ISOLTRAP
    Wienholtz, F.
    Atanasov, D.
    Kreim, S.
    Manea, V.
    Rosenbusch, M.
    Schweikhard, L.
    Welker, A.
    Wolf, R. N.
    PHYSICA SCRIPTA, 2015, T166
  • [25] A multi-reflection time-of-flight mass separator for isobaric purification of radioactive ion beams
    Wolf, Robert N.
    Eritt, Markus
    Marx, Gerrit
    Schweikhard, Lutz
    HYPERFINE INTERACTIONS, 2011, 199 (1-3): : 115 - 122
  • [26] A multi-reflection time-of-flight mass separator for isobaric purification of radioactive ion beams
    Robert N. Wolf
    Markus Eritt
    Gerrit Marx
    Lutz Schweikhard
    Hyperfine Interactions, 2011, 199 : 115 - 122
  • [27] Concatenated multi-reflection time-of-flight spectra for wide-band mass spectrometry
    Fischer, Paul
    Koenig, Stephan
    Schweikhard, Lutz
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2021, 463
  • [28] Multi-reflection time-of-flight mass spectrograph for short-lived radioactive ions
    P. Schury
    K. Okada
    S. Shchepunov
    T. Sonoda
    A. Takamine
    M. Wada
    H. Wollnik
    Y. Yamazaki
    The European Physical Journal A, 2009, 42 : 343 - 349
  • [29] Multi-reflection time-of-flight mass spectrograph for short-lived radioactive ions
    Schury, P.
    Okada, K.
    Shchepunov, S.
    Sonoda, T.
    Takamine, A.
    Wada, M.
    Wollnik, H.
    Yamazaki, Y.
    EUROPEAN PHYSICAL JOURNAL A, 2009, 42 (03): : 343 - 349
  • [30] A Radio-Frequency Ion Trap System for the Multi-Reflection Time-of-Flight Mass Spectrometer at SHANS and Its Offline Commissioning
    Wang, Jun-Ying
    Huang, Wen-Xue
    Tian, Yu-Lin
    Wang, Yong-Sheng
    Wang, Yue
    Zhang, Wan-Li
    Huang, Yuan-Jun
    Gan, Zai-Guo
    Xu, Hu-Shan
    ATOMS, 2023, 11 (11)