Shape optimization of pulsatile ventricular assist devices using FSI to minimize thrombotic risk

被引:0
|
作者
C. C. Long
A. L. Marsden
Y. Bazilevs
机构
[1] Los Alamos National Laboratory,T
[2] University of California,3 Fluid Dynamics and Solid Mechanics
[3] San Diego,Department of Mechanical and Aerospace Engineering
[4] University of California,Department of Structural Engineering
[5] San Diego,undefined
来源
Computational Mechanics | 2014年 / 54卷
关键词
Pulsatile VAD; Residence time; Fluid–structure interaction; Isogeometric analysis; Optimization; SMF;
D O I
暂无
中图分类号
学科分类号
摘要
In this paper we perform shape optimization of a pediatric pulsatile ventricular assist device (PVAD). The device simulation is carried out using fluid–structure interaction (FSI) modeling techniques within a computational framework that combines FEM for fluid mechanics and isogeometric analysis for structural mechanics modeling. The PVAD FSI simulations are performed under realistic conditions (i.e., flow speeds, pressure levels, boundary conditions, etc.), and account for the interaction of air, blood, and a thin structural membrane separating the two fluid subdomains. The shape optimization study is designed to reduce thrombotic risk, a major clinical problem in PVADs. Thrombotic risk is quantified in terms of particle residence time in the device blood chamber. Methods to compute particle residence time in the context of moving spatial domains are presented in a companion paper published in the same issue (Comput Mech, doi:10.1007/s00466-013-0931-y, 2013). The surrogate management framework, a derivative-free pattern search optimization method that relies on surrogates for increased efficiency, is employed in this work. For the optimization study shown here, particle residence time is used to define a suitable cost or objective function, while four adjustable design optimization parameters are used to define the device geometry. The FSI-based optimization framework is implemented in a parallel computing environment, and deployed with minimal user intervention. Using five SEARCH/POLL steps the optimization scheme identifies a PVAD design with significantly better throughput efficiency than the original device.
引用
收藏
页码:921 / 932
页数:11
相关论文
共 50 条
  • [41] Effects of pulsatile- and continuous-flow left ventricular assist devices on left ventricular unloading
    Garcia, Santiago
    Kandar, Forum
    Boyle, Andrew
    Colvin-Adams, Monica
    Lliao, Kenneth
    Joyce, Lyle
    John, Ranjit
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (03): : 261 - 267
  • [42] The incidence of right ventricular dysfunction is similar in pulsatile and continuous-flow left ventricular assist devices
    Feller, Erika D.
    Sorensen, Erik N.
    Johnson, Frances L.
    Gottlieb, Stephen S.
    Haddad, Michel
    Pierson, Richard N.
    Griffith, Bartley P.
    [J]. JOURNAL OF CARDIAC FAILURE, 2006, 12 (06) : S49 - S49
  • [43] Total Artificial Heart Using Bilateral Paracorporeal Pulsatile Ventricular Assist Devices in an 8.2-kg Child
    Woods, Ronald K.
    Kindel, Steven J.
    Mitchell, Michael E.
    Hraska, Viktor
    Niebler, Robert A.
    [J]. ANNALS OF THORACIC SURGERY, 2018, 105 (06): : E255 - E257
  • [44] Risk factors for stroke on left ventricular assist devices
    Bishawi, Muath
    Joseph, Jeremy
    Patel, Chetan
    Schroder, Jacob
    Daneshmand, Mani
    Bowles, Dawn
    Rogers, Joseph
    Milano, Carmelo
    [J]. JOURNAL OF CARDIAC SURGERY, 2018, 33 (06) : 348 - 352
  • [45] Effect of pulsatile amd nonpulsatile left ventricular assist devices on plasma renin activity
    Welp, H.
    Etz, C.
    Tjan, T.
    Scheld, H. H.
    Schmid, C.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2005, 24 (02): : S58 - S58
  • [46] Stroke Risk and Mortality in Patients with Ventricular Assist Devices
    Cool, Josephine A.
    Parikh, Neal S.
    Kamel, Hooman
    Karas, Maria G.
    Boehme, Amelia K.
    [J]. STROKE, 2016, 47
  • [47] Gastrointestinal Bleeding Risk in Patients with Ventricular Assist Devices
    French, J. B.
    Pamboukian, S. V.
    Smallfield, G. B.
    Peter, S.
    Tallaj, J. A.
    Brown, R. N.
    Smallfield, M. C.
    Kirklin, J. K.
    George, J. F.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2012, 31 (04): : S47 - S48
  • [48] Stroke Risk and Mortality in Patients With Ventricular Assist Devices
    Parikh, Neal S.
    Cool, Josephine
    Karas, Maria G.
    Boehme, Amelia K.
    Kamel, Hooman
    [J]. STROKE, 2016, 47 (11) : 2702 - 2706
  • [49] Effect of pulsatile and nonpulsatile flow on cerebral perfusion in patients with left ventricular assist devices
    Cornwell, William K., III
    Tarumi, Takashi
    Aengevaeren, Vincent L.
    Ayers, Colby
    Divanji, Punag
    Fu, Qi
    Palmer, Dean
    Drazner, Mark H.
    Meyer, Dan M.
    Bethea, Brian T.
    Hastings, Jeffrey L.
    Fujimoto, Naoki
    Shibata, Shigeki
    Zhang, Rong
    Markham, David W.
    Levine, Benjamin D.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (12): : 1295 - 1303
  • [50] A physiological control system for ECG-synchronized pulsatile pediatric ventricular assist devices
    Cordeiro, Thiago D.
    Sousa, Daniel L.
    Cestari, Idagene A.
    Lima, Antonio M. N.
    [J]. BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2020, 57