Potentiometric Performance of Silver Ion-Selective Electrodes Based on Tridentate Schiff Base Derivatives

被引:0
|
作者
Masato Ueda
Takahiro Saito
Keiichi Fujimori
Takayo Moriuchi
Yasuhiko Shibutani
机构
[1] Osaka Institute of Technology,Department of Applied Chemistry, Faculty of Engineering
来源
Analytical Sciences | 2004年 / 20卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
To examine the directivity for improving the silver ion discrimination ability of a Schiff base, three kinds of tridentate ligands were synthesized and compared with the similar quadridentate ligand as the silver ionophore. Among the Schiff base derivatives tested, 3-(2-pyridylethylimino)-2-butanoneoxime, having one oxime and a pyridine substituent, was found to be the best ionophore for a silver-ion electrode. The electrode based on this derivative exhibited good silver-ion selectivity, -log KpAogt+,K+ = 3.8, comparable to that of a quadridentate Schiff base, N,N′-bis(2′-hydroxyimino-1′-phenylpropyleden)-1,3-propanediamine, reported previously, except for a pseudo Nernstian response (35.6 mV decade−1) with a wide silver-ion activity change in the activity change from 5.0 × 10−7 to 7.9 × 10−2 mol dm−3.
引用
收藏
页码:1649 / 1653
页数:4
相关论文
共 50 条
  • [32] POTENTIOMETRIC DETERMINATION OF GOLD IN ELECTROLYTES USING ION-SELECTIVE ELECTRODES
    BOROVSKII, ES
    RAKHMANKO, EM
    STAROBINETS, GL
    JOURNAL OF ANALYTICAL CHEMISTRY OF THE USSR, 1985, 40 (08): : 1127 - 1130
  • [33] POTENTIOMETRIC DETERMINATION OF HALOGENS IN ALUMINA PRODUCTION WITH ION-SELECTIVE ELECTRODES
    BLOKH, MS
    TASHUTA, GN
    KHALYAPINA, OB
    JOURNAL OF APPLIED CHEMISTRY OF THE USSR, 1992, 65 (10): : 1845 - 1849
  • [34] SINGLE-POINT POTENTIOMETRIC TITRATIONS WITH ION-SELECTIVE ELECTRODES
    HORVAI, G
    TOTH, K
    PUNGOR, E
    ANALYTICA CHIMICA ACTA, 1979, 107 (JUN) : 101 - 104
  • [35] GALVANIC CELL WITH 2 ION-SELECTIVE ELECTRODES AS A POTENTIOMETRIC SENSOR
    STEFANOVA, OK
    ALAGOVA, ZS
    GREKOVICH, AL
    ROZHDESTVENSKAYA, NV
    KHRIPUN, GA
    JOURNAL OF APPLIED CHEMISTRY OF THE USSR, 1991, 64 (11): : 2132 - 2135
  • [36] Selective fluoride recognition and potentiometric properties of ion-selective electrodes based on bis(halodiphenyistannyl)alkanes
    Perdikaki, K
    Tsagkatakis, I
    Chaniotakis, NA
    Altmann, R
    Jurkschat, K
    Reeske, G
    ANALYTICA CHIMICA ACTA, 2002, 467 (1-2) : 197 - 204
  • [37] Solid-Contact Potentiometric Anion Sensing Based on Classic Silver/Silver Insoluble Salts Electrodes without Ion-Selective Membrane
    Liao, Chunxian
    Zhong, Lijie
    Tang, Yitian
    Sun, Zhonghui
    Lin, Kanglong
    Xu, Longbin
    Lyu, Yan
    He, Dequan
    He, Ying
    Ma, Yingming
    Bao, Yu
    Gan, Shiyu
    Niu, Li
    MEMBRANES, 2021, 11 (12)
  • [38] Silver(I) ion-selective membrane based on Schiff base-p-tert-butylcalix[4]arene
    Mahajan, RK
    Kumar, M
    Sharma, V
    Kaur, I
    ANALYST, 2001, 126 (04) : 505 - 507
  • [39] Silver ion-selective electrodes based on π-coordinating ionophores without heteroatoms
    Bobacka, J
    Lahtinen, T
    Koskinen, H
    Rissanen, K
    Lewenstam, A
    Ivaska, A
    ELECTROANALYSIS, 2002, 14 (19-20) : 1353 - 1357
  • [40] SILVER BROMIDE BASED CHALCOGENIDE GLASSY CRYSTALLINE ION-SELECTIVE ELECTRODES
    VLASOV, YG
    MOSKVIN, LN
    BYCHKOV, EA
    GOLIKOV, DV
    ANALYST, 1989, 114 (02) : 185 - 190