Effect of table height displacement and patient center deviation on size-specific dose estimates calculated from computed tomography localizer radiographs

被引:0
|
作者
Kazuki Kuriyama
Kosuke Matsubara
Shu Hisahara
Yukie Nagata
Rumiko Nosaka
Rika Goto
Nobutaka Yanano
Kanichiro Shimizu
Tomokazu Shoji
机构
[1] The Jikei University Kashiwa Hospital,Department of Radiology
[2] Kanazawa University,Division of Health Sciences, Department of Quantum Medical Technology, Graduate Course of Medical Science and Technology, Graduate School of Medical Science
[3] Kanazawa University,Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences
[4] The Jikei University School of Medicine,Department of Radiology
[5] Kashiwa Hospital,Department of Radiology
[6] The Jikei University Hospital,undefined
关键词
Size-specific dose estimate; Miscentering; Table height; Computed tomography; Localizer radiograph;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of this study is to evaluate the effect of table height displacement and patient center deviation along the z\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$z$$\end{document}-axis on size-specific dose estimate (SSDE) calculations based on computed tomography (CT) localizer radiographs in pediatric and adult abdominal CT examinations. CT localizer radiographs and CT axial images were acquired with table heights of − 5.0, − 2.5, 0.0 (center), 2.5, and 5.0 cm using two acrylic self-made phantoms filled with water. Water-equivalent diameters (Dw\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$${D}_{\mathrm{w}}$$\end{document}) were calculated from the CT localizer radiographs and CT axial images. Relative errors of SSDEs from the CT localizer radiographs to SSDEs from the CT axial images were calculated to evaluate the effect of table height displacement. Furthermore, patient center deviations and indices of SSDE overestimation were measured from the clinical data of 110 abdominal CT examinations. The relative errors of SSDEs in phantoms equivalent to 1-year-old and 20-year-old Japanese reference persons ranged from − 2.45% (table height of 50 mm) to + 1.88% (− 50 mm) and from − 4.22% (50 mm) to + 3.79% (− 50 mm), respectively. The largest center deviation in all patients ranged from − 43.1 to 21.5 mm (median: − 14.4 mm). The indices of SSDE overestimation for all patients ranged from − 16.2 to 6.9 mm (median: − 2.2 mm). We found that the effects of table height displacement and patient center deviation along the z\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$z$$\end{document}-axis on SSDEs calculated from CT localizer radiographs in pediatric phantoms were smaller compared to adult phantoms. In order to correct these patient center deviations, it is necessary to apply an appropriate correction technique in each section along the z\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$z$$\end{document}-axis.
引用
收藏
页码:665 / 672
页数:7
相关论文
共 36 条
  • [31] From age-specific to size-specific dose protocol for paediatric head computed tomography: a simple practical strategy for necessity assessment and parameter setting
    Hu, Liwei
    Ouyang, Rongzhen
    Li, Jing
    Zhang, Guozhi
    Li, Jiong
    Guo, Chen
    Wang, Qian
    Zhong, Yumin
    RADIATION PROTECTION DOSIMETRY, 2024, 200 (20) : 1981 - 1989
  • [32] Local diagnostic reference level based on size-specific dose estimates: Assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children’s hospital
    Rumi Imai
    Osamu Miyazaki
    Tetsuya Horiuchi
    Hideo Kurosawa
    Shunsuke Nosaka
    Pediatric Radiology, 2015, 45 : 345 - 353
  • [33] Estimating Patient Dose from X-ray Tube Output Metrics: Automated Measurement of Patient Size from CT Images Enables Large-scale Size-specific Dose Estimates
    Ikuta, Ichiro
    Warden, Graham I.
    Andriole, Katherine P.
    Khorasani, Ramin
    Sodickson, Aaron
    RADIOLOGY, 2014, 270 (02) : 472 - 480
  • [34] Local diagnostic reference level based on size-specific dose estimates: Assessment of pediatric abdominal/pelvic computed tomography at a Japanese national children's hospital
    Imai, Rumi
    Miyazaki, Osamu
    Horiuchi, Tetsuya
    Kurosawa, Hideo
    Nosaka, Shunsuke
    PEDIATRIC RADIOLOGY, 2015, 45 (03) : 345 - 353
  • [35] An Age-Based Size-Specific Dose Estimate for Pediatric Computed Tomography Head Examinations Performed at Songklanagarind Hospital, Thailand, from 2017 to 2019
    Poosiri, Saowapark
    Chuboonlap, Kanokkwan
    Kaewlaied, Nuttita
    APPLIED SCIENCES-BASEL, 2024, 14 (17):
  • [36] Size-specific dose estimates of adult, chest computed tomography examinations: Comparison of Chinese and updated 2017 American College of Radiology diagnostic reference levels based on the water-equivalent diameter
    Hu, Xiaoyan
    Gou, Jie
    Lin, Wei
    Zou, Chunhua
    Li, Wenbo
    PLOS ONE, 2021, 16 (09):