Estimation of haemoglobin using non-invasive portable device with spectroscopic signal application

被引:0
|
作者
Amrutha, A. M. [1 ]
Sidenur, Bhagyalaxmi [1 ]
Balu, P. S. [2 ]
Savitha, S., V [3 ]
Nagendra Gowda, M. R. [1 ]
Rajagopal, Harshavardhan [4 ]
机构
[1] Basaveshwara Med Coll & Hosp, Dept Community Med, Chitradurga, India
[2] Basaveshwara Med Coll & Hosp, C DART, Chitradurga, India
[3] SJM Inst Nursing, Chitradurga, India
[4] Amer Hosp Dubai, Dubai, U Arab Emirates
关键词
Anaemia; Nursing students; Non-invasive; Spectroscopy; EzeCheck; Haemoglobinometer; PREVALENCE; ANEMIA;
D O I
10.1038/s41598-024-58990-z
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
An estimated 52% of non-pregnant women of reproductive age in India are estimated to be affected by anaemia, which is categorised as a chronic condition. In 2019-2021, the National Family Health Survey-5 (NFHS-5) which was undertaken revealed the following statistics about the prevalence of anaemia in the state of Karnataka. To estimate haemoglobin levels using non-invasive portable device among nursing students. A cross sectional study was done among students of Nursing college in central Karnataka for a period of 3 months. Total of 140 students were included in the study. EzeCheck haemoglobin estimation was done twice and was recorded in the same Google form. The frequency and percentage of variation of results between Haematology Analyzer and EzeCheck devices was presented with a range of difference such as 0, less than 1, 1.0 to 1.9, 2.0 to 2.9, 3.0 to 3.9, and 4.0 and above. The total prevalence of anaemia among nursing students was 57.8% and most of the students had moderate degree of anaemia (28.6%). Two readings of haemoglobin were taken and difference of readings were calculated and majority of the students had difference of < 0.5 gm/dl (61.4%) and only 1.4% of the students had difference of > 2 gm/dl. The mean difference of haemoglobin of two readings was 0.5 +/- 0.5 gm/dl. The technology employed in this study bridges the gap between patients and anaemia diagnosis by providing screening services. The device provides the diagnosis via a non-invasive, IoT-enabled service at a low cost.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] Smart Non-Invasive Hemoglobin Measurement Using Portable Embedded Technology
    Tatiparti, Padma
    Chintalapati, Usha Kumari
    [J]. INTERNATIONAL JOURNAL OF ONLINE AND BIOMEDICAL ENGINEERING, 2020, 16 (15) : 106 - 120
  • [32] Non-Invasive Estimation of Temperature by Using Wavelet Transform
    Tellez, A.
    Leija, L.
    Vera, A.
    [J]. PAHCE: 2009 PAN AMERICAN HEALTH CARE EXCHANGES, 2009, : 77 - 81
  • [33] Application of a portable XRF spectrometer for the non-invasive analysis of museum metal artefacts
    Karydas, A. G.
    [J]. ANNALI DI CHIMICA, 2007, 97 (07) : 419 - 432
  • [34] Non-invasive Blood Pressure Estimation Using Phonocardiogram
    Dastjerdi, Amirhossein Esmaili
    Kachuee, Mohammad
    Shabany, Mahdi
    [J]. 2017 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2017, : 80 - 83
  • [35] SPECTROSCOPIC SYSTEMS FOR NON-INVASIVE MEASUREMENTS
    CALLIS, JB
    [J]. CLINICAL CHEMISTRY, 1989, 35 (06) : 1059 - 1060
  • [36] Non-invasive Estimation of Cardiovascular Parameters using Ballistocardiography
    van Rooij, B. J. M.
    Tavakolian, K.
    Arzanpour, S.
    Blaber, A. P.
    Leguy, C. A. D.
    [J]. 2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 1247 - 1250
  • [37] Portable sensor for non-invasive prepreg assessment
    Prado, PJ
    Magnuson, E
    Menon, S
    [J]. BRIDGING THE CENTURIES WITH SAMPE'S MATERIALS AND PROCESSES TECHNOLOGY, VOL 45, BOOKS 1 AND 2, 2000, : 1305 - 1315
  • [38] Evaluation of accuracy of non-invasive haemoglobin estimation by pulse oximetry during major surgical haemorrhage
    Pradhan, S.
    Divatia, J.
    Verma, P.
    [J]. ANAESTHESIA, 2018, 73 : 14 - 14
  • [39] Objective assessment of dry mouth using a non-invasive device
    Ishimoto, Shin-Ichi
    Tsunoda, Koichi
    Akiya, Kumiko
    Fujimaki, Youko
    Okada, Kazunari
    Kinoshita, Makoto
    Yoshida, Tomohiko
    [J]. ACTA OTO-LARYNGOLOGICA, 2009, 129 (12) : 1527 - 1528
  • [40] Improving home oxigen therapy in patients with interstitial lung diseases (ILDs): Application of a portable non-invasive ventilation (NIV) device
    Cuerpo, Sandra
    Francesqui, Joel
    Hernandez, Fernanda
    Palomo, Maria
    Hernandez, Carmen
    Blanco, Isabel
    Embid, Cristina
    Sellares, Jacobo
    [J]. EUROPEAN RESPIRATORY JOURNAL, 2019, 54