Keratometry with five different techniques: a study of device repeatability and inter-device agreement

被引:0
|
作者
Shiva Mehravaran
Soheila Asgari
Sara Bigdeli
Ashkan Shahnazi
Hassan Hashemi
机构
[1] Noor Eye Hospital,Noor Ophthalmology Research Center
来源
关键词
Keratometry; Javal manual keratometer; Topcon automated kerato-refractometer; IOLMaster; EyeSys Corneal Analysis System; Pentacam Scheimpflug Topography System;
D O I
暂无
中图分类号
学科分类号
摘要
The purpose of this study was to determine and compare the repeatability of keratometry measurements with the Javal manual keratometer, Topcon automated kerato-refractometer, IOLMaster, EyeSys Corneal Analysis System, and Pentacam Scheimpflug Topography System, and determine the agreement of measurements of the latter four devices with the Javal keratometer as the gold standard. In this cross-sectional study, 21 people with no history of ocular disease or surgery were examined twice with each device. Minimum, maximum, and average keratometry readings were recorded in diopters (D) for each eye and used in the analyses. For statistical analysis, we determined correlation coefficients and used the Bland–Altman method and calculated the 95 % limits of agreement (LoA). All repeatability coefficients were satisfactorily high. Best repeatability for minimum keratometry and maximum keratometry readings was seen with IOLMaster (95 % LoA −0.23 to 0.19 D and −0.31 to 0.32 D, respectively). Agreement with Javal manual keratometry was best with Topcon and IOLMaster for minimum keratometry readings (95 % LoA −0.67 to 0.28 D and −0.57 to 0.38 D, respectively), and with Topcon for maximum keratometry readings (95 % LoA −0.85 to 0.63 D). In our series of normal eyes, keratometry readings with Topcon, IOLMaster, and Pentacam showed very good agreement with Javal measurements, and inter-device agreements were better than the repeatability of the manual gold standard. However, differences between Javal and EyeSys may not be clinically acceptable, and these two devices should not be used interchangeably. Further studies are needed to investigate these issues in non-virgin eyes.
引用
收藏
页码:869 / 875
页数:6
相关论文
共 50 条
  • [21] Inter-observer and Inter-device Agreement For The Assessment Of Anterior-Chamber Angle Photography with An Automated Gonioscope.
    Scotto, Riccardo
    Cutolo, Carlo Alberto
    Iester, Michele M.
    Traverso, Carlo Enrico
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (09)
  • [22] Inter-device Interaction Control Framework for the City Tour
    Chiu, Hung-Sheng
    Tsai, Tse-Ming
    Chou, Shih-Chun
    Wang, Hao-Cheng
    EDUCATION AND EDUCATION MANAGEMENT, 2012, 5 : 308 - 313
  • [23] Inter-device Variability of the Stratus Optical Coherence Tomography
    Barkana, Yaniv
    Burgansky-Eliash, Zvia
    Gerber, Yariv
    Melamed, Shlomo
    Neudorfer, Meira
    Avni, Isaac
    Bartov, Elisha
    Morad, Yair
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (02) : 260 - 266
  • [24] Inter-observer and inter-device agreement in choroidal and scleral thickness measurement between Spectralis and Atlantis OCT systems
    Chen, Fred Kuanfu
    Wong, Evan
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [25] A statistical model for removing inter-device differences in spectroscopy
    Wang, Lu
    Lee, Jong Soo
    Lane, Pierre
    Atkinson, E. Neely
    Zuluaga, Andres
    Follen, Michele
    MacAulay, Calum
    Cox, Dennis D.
    OPTICS EXPRESS, 2014, 22 (07): : 7617 - 7624
  • [26] AN INTER-DEVICE ACCURACY COMPARISON OF CONSUMER SLEEP TRACKERS
    Zavrel, Erik A.
    Krieger, Ana C.
    2019 DESIGN OF MEDICAL DEVICES CONFERENCE, 2019,
  • [27] Inter-Device Agreement between Fitbit Flex 1 and 2 for Assessing Sedentary Behavior and Physical Activity
    Kwon, Sunku
    Burns, Ryan D.
    Kim, Youngwon
    Bai, Yang
    Byun, Wonwoo
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2021, 18 (05) : 1 - 11
  • [28] Optical coherence tomography angiography for quantitative microvascular assessment in diabetic retinopathy: inter-device and intra-device agreement and correlation with clinical staging
    Emanuele Crincoli
    Donato Colantuono
    Zhanlin Zhao
    Eric H. Souied
    Alexandra Miere
    Acta Diabetologica, 2022, 59 : 1219 - 1227
  • [29] Optical coherence tomography angiography for quantitative microvascular assessment in diabetic retinopathy: inter-device and intra-device agreement and correlation with clinical staging
    Crincoli, Emanuele
    Colantuono, Donato
    Zhao, Zhanlin
    Souied, Eric H.
    Miere, Alexandra
    ACTA DIABETOLOGICA, 2022, 59 (09) : 1219 - 1227
  • [30] An Inter-Device Communication Protocol for Modular Smart-Objects
    Brama, Riccardo
    Tundo, Piergiuseppe
    Della Ducata, Armando
    Malvasi, Angelo
    2014 IEEE WORLD FORUM ON INTERNET OF THINGS (WF-IOT), 2014, : 422 - 427