Evaluation of MR-based PET Images with a Homogeneous μmap by Using Integrated PET/MR

被引:0
|
作者
Chan Rok Park
Jung Soo Kim
Beomyeol Baek
Seong Ho Kim
Hyungjin Yang
机构
[1] The Korea University,Department of Biomedical Science
[2] Seoul National University Hospital,Department of Nuclear Medicine
[3] Dongnam Health University,Department of Radiological Technology
[4] Kangbuk Samsung Hospital,Department of Nuclear Medicine
[5] Konkuk University Medical Center,Department of Radiology
[6] The Korea University,Department of Physics
来源
关键词
MR-based μmap; Nuclear Medicine; Integrated PET/MR; NEMA IEC body phantom;
D O I
暂无
中图分类号
学科分类号
摘要
This study intends to obtain magnetic resonance (MR)-based positron emission tomography (PET) images with a relatively uniform μmap by using a National Electrical Manufacturers Association (NEMA) International Electrotechnical Commission (IEC) body phantom and to evaluate the PET phantom images by applying various reconstruction parameters in accordance with the protocol of the NEMA. In order to obtain a uniform MR-based μmap, we set the initial voltage of the RF transmitter adjustment algorithm to 75 V, and we obtained the attenuation-corrected PET image by applying the two-point vibe Dixon MR sequence. The images reconstructed using the 3D ordered subset expectation maximization (OSEM) algorithm with the point spread function (PSF) showed the highest percentage of contrast recovery and the lowest percentage background variability compared to the other 3D reconstruction algorithms like the Filtered backprojection (FBP), 3D OSEM, and 3D OSEM with PSF algorithms. The 3D OSEM with PSF reconstruction algorithm is confirmed to be effective for maintaining the optimum image quality.
引用
收藏
页码:671 / 676
页数:5
相关论文
共 50 条
  • [31] Comparative evaluation of MR-based partial-volume correction schemes for PET
    Meltzer, CC
    Kinahan, PE
    Greer, PJ
    Nichols, TE
    Comtat, C
    Cantwell, MN
    Lin, MP
    Price, JC
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (12) : 2053 - 2065
  • [32] MR-BASED ATTENUATION MAP RE-ALIGNMENT AND MOTION CORRECTION IN SIMULTANEOUS BRAIN MR-PET IMAGING
    Sforazzini, Francesco
    Chen, Zhaolin
    Baran, Jakub
    Bradley, Jason
    Carey, Alexandra
    Shah, N. Jon
    Egan, Gary
    2017 IEEE 14TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI 2017), 2017, : 231 - 234
  • [33] Evaluation of atlas-based MR-attenuation correction for integrated PET-MR in human brain by using PET-CT and PET-MR patient data
    Sekine, Tetsuro
    Buck, Alfred
    Delso, Gaspar
    ter Voert, Edwin
    Huellner, Martin
    Veit-Haibach, Patrick
    Warnock, Geoffrey
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [34] MR- and PET-based respiratory gating for integrated PET/MR imaging
    Fuerst, Sebastian
    Grimm, Robert
    Hong, Inki
    Souvatzoglou, Michael
    Casey, Michael
    Schwaiger, Markus
    Nekolla, Stephan
    Ziegler, Sibylle
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [35] MR- and PET-based respiratory gating for integrated PET/MR imaging
    Fuerst, Sebastian
    Grimm, Robert
    Hong, Inki
    Souvatzoglou, Michael
    Casey, Michael
    Schwaiger, Markus
    Nekolla, Stephan
    Ziegler, Sibylle
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [36] Accurate hybrid template-based and MR-based attenuation correction using UTE images for simultaneous PET/MR brain imaging applications
    Baran, Jakub
    Chen, Zhaolin
    Sforazzini, Francesco
    Ferris, Nicholas
    Jamadar, Sharna
    Schmitt, Ben
    Faul, David
    Shah, Nadim Jon
    Cholewa, Marian
    Egan, Gary F.
    BMC MEDICAL IMAGING, 2018, 18
  • [37] Joint Reconstruction of Activity and Attenuation Using MR-Based Priors: Application to Clinical TOF PET/MR
    Ahn, Sangtae
    Cheng, Lishui
    Shanbhag, Dattesh
    Wiesinger, Florian
    Manjeshwar, Ravindra
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [38] Feasibility of using ZTE to generate an accurate MR-based attenuation map for head and neck PET/MR studies: Similarity analysis to CT-based attenuation map
    Lee, K.
    Zaharchuk, G.
    Gulaka, P. K.
    Levin, C. S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S445 - S446
  • [39] Truncation Completion of MR-based PET Attenuation Maps Using Time-of-flight Non-attenuation-corrected PET Images
    Qian, Rua
    Manjeshwar, Ravi M.
    Ambwani, Sonal
    Wollenweber, Scott D.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2773 - 2775
  • [40] INTEGRATED DISPLAY, MANIPULATION AND ANALYSIS OF MR AND PET IMAGES
    UDUPA, JK
    SAMARASEKERA, S
    ALAVI, A
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (05) : P124 - P124