A New Air Displacement Plethysmograph for the Measurement of Body Composition in Infants

被引:0
|
作者
Alessandro Urlando
Philip Dempster
Susan Aitkens
机构
[1] Life Measurement Incorporated,
来源
Pediatric Research | 2003年 / 53卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
This article introduces and evaluates the PEA POD Infant Body Composition System, an air displacement plethysmograph designed for the assessment of body composition in infants between birth and 6 mo of age. The performance of the PEA POD was evaluated by repeated testing of National Institute of Standards and Technology–traceable weights and volumes. Mass was measured in a single session. Volume was measured in four sessions during a 2-d period (five times/session for both). The mean values for repeated mass measurements were almost identical to the masses of traceable weights. The SD and CV for repeated volume measurements were 1.1-4.5 mL and 0.02-0.09%, respectively. Both the mean SD and CV were within very narrow ranges (1.4-3.1 mL for SD and 0.03-0.08% for CV) across all volume levels. Furthermore, mean CV values using results from the four sessions indicated excellent within- and between-day reliability. Regression analyses (by session or with all sessions combined) of the measured volume against actual volume gave very low standard error of the estimate (SEE) (0.853-1.957 mL) and very high R2 (1.000), with the intercept and slope not significantly different from 0 and 1, respectively. The mean percentage error in volume measurements was ≤0.05% at all volume levels. The study findings and the operational and physical characteristics of the system indicate that the PEA POD has the potential to provide clinicians and researchers with a diagnostic and research tool that is accurate, easily used by operators, and comfortable for subjects.
引用
收藏
页码:486 / 492
页数:6
相关论文
共 50 条
  • [1] A new air displacement plethysmograph for the measurement of body composition in infants
    Urlando, A
    Dempster, P
    Aitkens, S
    [J]. PEDIATRIC RESEARCH, 2003, 53 (03) : 486 - 492
  • [2] Validation of a new pediatric air displacement plethysmograph for assessing body composition in infants
    Ma, GS
    Yao, MJ
    Urlando, A
    Wong, W
    Nommsen-Rivers, L
    Dewey, K
    [J]. OBESITY RESEARCH, 2003, 11 : A134 - A134
  • [3] Preliminary evaluation of a new pediatric air displacement plethysmograph for body composition assessment in infants
    M. Yao
    L. Nommsen-Rivers
    K. Dewey
    A. Urlando
    [J]. Acta Diabetologica, 2003, 40 : s55 - s58
  • [4] Preliminary evaluation of a new pediatric air displacement plethysmograph for body composition assessment in infants
    Yao, M
    Nommsen-Rivers, L
    Dewey, K
    Urlando, A
    [J]. ACTA DIABETOLOGICA, 2003, 40 (Suppl 1) : S55 - S58
  • [5] Validation of a new pediatric air-displacement plethysmograph for assessing body composition in infants
    Ma, GS
    Yao, MJ
    Liu, Y
    Lin, AW
    Zou, H
    Urlando, A
    Wong, WW
    Nommsen-Rivers, L
    Dewey, KG
    [J]. AMERICAN JOURNAL OF CLINICAL NUTRITION, 2004, 79 (04): : 653 - 660
  • [6] A new air displacement plethysmograph for the determination of Japanese body composition
    Miyatake, N
    Nonaka, K
    Fujii, M
    [J]. DIABETES OBESITY & METABOLISM, 1999, 1 (06): : 347 - 351
  • [7] EVALUATION OF A NEW AIR DISPLACEMENT PLETHYSMOGRAPH FOR MEASURING HUMAN-BODY COMPOSITION
    MCCRORY, M
    GOMEZ, T
    BERNAUER, E
    MOLE, P
    [J]. FASEB JOURNAL, 1995, 9 (04): : A980 - A980
  • [8] EVALUATION OF A NEW AIR DISPLACEMENT PLETHYSMOGRAPH FOR MEASURING HUMAN-BODY COMPOSITION
    MCCRORY, MA
    GOMEZ, TD
    BERNAUER, EM
    MOLE, PA
    [J]. MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1995, 27 (12): : 1686 - 1691
  • [9] Body Composition Analysis in Bariatric Surgery: Use of Air Displacement Plethysmograph
    Huettner, Franziska
    Rammos, Charalambos K.
    Dynda, Danuta I.
    Lange, Melinda L.
    Marshall, J. Stephen
    Rossi, Thomas R.
    DeBord, James R.
    [J]. AMERICAN SURGEON, 2012, 78 (06) : 698 - 701
  • [10] MEASUREMENT OF BODY VOLUME IN INFANTS BY AIR DISPLACEMENT
    PEARSE, R
    SCOPES, J
    TAYLOR, A
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1976, 256 (01): : P5 - P5