Implantable CMOS neuro-stimulus chip for visual prosthesis

被引:0
|
作者
Xu Zhang
WeiHua Pei
BeiJu Huang
ShuJing Wang
Ning Guan
Kai Guo
Yu Wang
Qiang Gui
Jin Chen
Kai Wang
HuiJuan Wu
XiaoXin Li
Kai Li
HongDa Chen
机构
[1] Chinese Academy of Sciences,State Key Laboratory on Integrated Optoelectronics, Institute of Semiconductors
[2] People’s Hospital of Peking University,Department of Ophthalmology
[3] Peking University,Department of Epidemiology and Biostatistics, Health Science Center
来源
关键词
biomedical microelectronics; visual prosthesis; functional electrical stimulation; acute animal experiment; electrical evoked potential (EEP) test;
D O I
暂无
中图分类号
学科分类号
摘要
A prototype chip with 2×2 pixels for implanting in blind patients affected by outer retinal degeneration is presented in this paper. This visual prosthesis chip imitates the degenerated photoreceptor cells, senses the incident light and stimulates the remaining healthy layers of retina or optic nerve. Each pixel integrates photodiode and stimulus pulse generator, converting the illumination on the eyes into 3-bit resolution bi-phasic current pulses. On-chip charge cancellation modules are used to discharge each electrode site for tissue safety. The prototype chip is designed and fabricated in 0.18-μm N-well CMOS (complementary metal oxide semiconductor) 1P6M Mix-signal process, with a ±2.5 V dual voltage supply. The functionality of the fabricated chip is demonstrated on anesthetized rabbits. Neural responses in visual cortex are successfully evoked by the neuro-stimulus chip through an on-board trigger interface and flexible electrode.
引用
收藏
页码:898 / 908
页数:10
相关论文
共 50 条
  • [41] Flexible Multi-Electrode Array with Integrated Bendable CMOS-Chip for Implantable Systems
    Winkin, N.
    Mokwa, W.
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 3882 - 3885
  • [42] Pillar-shaped stimulus electrode array for high-efficiency stimulation of fully implantable epiretinal prosthesis
    Lee, Kang-Wook
    Watanabe, Yoshinobu
    Kigure, Chikashi
    Fukushima, Takafumi
    Koyanagi, Mitsumasa
    Tanaka, Tetsu
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2012, 22 (10)
  • [43] Fabrication and In vivo Evaluation of Poly(3,4-ethylenedioxythiophene) Stimulus Electrodes for Fully Implantable Retinal Prosthesis
    Kigure, Chikashi
    Naganuma, Hideki
    Sasaki, Yuichiro
    Kino, Hisashi
    Tomita, Hiroshi
    Tanaka, Tetsu
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2013, 52 (04)
  • [44] A 0.8-V 4096-Pixel CMOS Sense-and-Stimulus Imager for Retinal Prosthesis
    Lee, Chih-Lin
    Hsieh, Chih-Cheng
    IEEE TRANSACTIONS ON ELECTRON DEVICES, 2013, 60 (03) : 1162 - 1168
  • [45] A new CMOS multimode digital pixel sensor dedicated to an implantable visual cortical stimulator
    M. Sawan
    A. Trépanier
    J.-L. Trépanier
    Y. Audet
    R. Ghannoum
    Analog Integrated Circuits and Signal Processing, 2006, 49 : 187 - 197
  • [46] A new CMOS multimode digital pixel sensor dedicated to an implantable visual cortical stimulator
    Sawan, M.
    Trepanier, A.
    Trepanier, J. -L.
    Audet, Y.
    Ghannoum, R.
    ANALOG INTEGRATED CIRCUITS AND SIGNAL PROCESSING, 2006, 49 (02) : 187 - 197
  • [47] Cortical visual neuro-prosthesis for the blind:: Retina-like software/hardware preprocessor
    Pelayo, FJ
    Martinez, A
    Romero, S
    Morillas, CA
    Ros, E
    Fernández, E
    1ST INTERNATIONAL IEEE EMBS CONFERENCE ON NEURAL ENGINEERING 2003, CONFERENCE PROCEEDINGS, 2003, : 150 - 153
  • [48] CMOS LSI-based multi-chip flexible retinal prosthesis device for subretinal implantation
    Tokuda, T.
    Sawamura, S.
    Terasawa, Y.
    Tano, Y.
    Ohta, J.
    PROCEEDINGS OF THE IEEE 2008 CUSTOM INTEGRATED CIRCUITS CONFERENCE, 2008, : 325 - +
  • [49] Self Vth-compensating CMOS on-chip rectifier for inductively powered implantable medical devices
    Potocny, Miroslav
    Stopjakova, Viera
    Kovac, Martin
    2018 IEEE 21ST INTERNATIONAL SYMPOSIUM ON DESIGN AND DIAGNOSTICS OF ELECTRONIC CIRCUITS AND SYSTEMS (DDECS), 2018, : 158 - 161
  • [50] A THIN FILM FLEXIBLE ANTENNA WITH CMOS RECTIFIER CHIP FOR RF-POWERED IMPLANTABLE NEURAL INTERFACES
    Okabe, K.
    Akita, I.
    Yamagiwa, S.
    Kawano, T.
    Ishida, M.
    2015 TRANSDUCERS - 2015 18TH INTERNATIONAL CONFERENCE ON SOLID-STATE SENSORS, ACTUATORS AND MICROSYSTEMS (TRANSDUCERS), 2015, : 1751 - 1754