Procedures for trace cobalt determinations by adsorptive stripping voltammetry at in situ and ex situ plated bismuth film electrodes are presented. These exploit the enhancement of the cobalt peak obtained by using the Co(II)–dimethylglyoxime–cetyltrimethylammonium bromide–piperazine-N,N′-bis(2-ethanesulfonic acid) system. The calibration graph for an accumulation time of 120 s was linear from 2 × 10–10 to 2 × 10–8 mol L−1. The relative standard deviation from five determinations of cobalt at a concentration of 5 × 10–9 mol L−1 was 5.2%. The detection limit for an accumulation time of 300 s was 1.8 × 10–11 mol L−1. The proposed procedure was applied to cobalt determination in certified reference materials and in tap and river water samples.
引用
收藏
页码:1678 / 1682
页数:4
相关论文
共 50 条
  • [41] Determination of trace bismuth by adsorptive stripping voltammetry with pyrogallol red modified carbon paste electrode
    Guo, HS
    Li, YH
    [J]. CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 28 (12) : 1527 - 1530
  • [42] Determination of Se(IV) by adsorptive cathodic stripping voltammetry at a Bi/Hg film electrode
    Sharifian, Parisa
    Aliakbar, Alireza
    [J]. ANALYTICAL METHODS, 2015, 7 (05) : 2121 - 2128
  • [43] Sensitive and fast determination of papaverine by adsorptive stripping voltammetry on renewable mercury film electrode
    Piech, Robert
    Paczosa-Bator, Beata
    [J]. CENTRAL EUROPEAN JOURNAL OF CHEMISTRY, 2013, 11 (05): : 736 - 741
  • [44] Determination of Diazepam, Temazepam and Oxazepam at the Lead Film Electrode by Adsorptive Cathodic Stripping Voltammetry
    Tyszczuk, Katarzyna
    [J]. ELECTROANALYSIS, 2010, 22 (17-18) : 1975 - 1984
  • [45] Trace determination of chromium by square-wave adsorptive stripping voltammetry on bismuth film electrodes
    Chatzitheodorou, E
    Economou, A
    Voulgaropoulos, A
    [J]. ELECTROANALYSIS, 2004, 16 (21) : 1745 - 1754
  • [46] Determination of Trace Antimony by Square-Wave Adsorptive Cathodic Stripping Voltammetry at an Ex Situ Prepared Bismuth Film Electrode
    Su, Jiali
    Zhong, Shangwei
    Li, Xiangjun
    Zou, Hong
    [J]. JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2014, 161 (09) : H512 - H516
  • [47] Disposable electrochemical flow cells for catalytic adsorptive stripping voltammetry (CAdSV) at a bismuth film electrode (BiFE)
    Nahid Gharib Naseri
    Sara J. Baldock
    Anastasios Economou
    Nicholas J. Goddard
    Peter R. Fielden
    [J]. Analytical and Bioanalytical Chemistry, 2008, 391 : 1283 - 1292
  • [48] Disposable electrochemical flow cells for catalytic adsorptive stripping voltammetry (CAdSV) at a bismuth film electrode (BiFE)
    Naseri, Nahid Gharib
    Baldock, Sara J.
    Economou, Anastasios
    Goddard, Nicholas J.
    Fielden, Peter R.
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2008, 391 (04) : 1283 - 1292
  • [49] Adsorptive stripping voltammetry of nickel with 1-nitroso-2-napthol using a bismuth film electrode
    Segura, Rodrigo
    Pradena, Milenco
    Pinto, Deny
    Godoy, Fernando
    Nagles, Edgar
    Arancibia, Veronica
    [J]. TALANTA, 2011, 85 (05) : 2316 - 2319
  • [50] Determination of cobalt in seawater by catalytic adsorptive cathodic stripping voltammetry
    Vega, M
    van den Berg, CMG
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (05) : 874 - 881