Device for measuring the forces exerted on the shaft of an endoscope during colonoscopy

被引:0
|
作者
C. A. Mosse
T. N. Mills
G. D. Bell
C. P. Swain
机构
[1] University College London,Department of Medical Physics
[2] Sunderland Royal Hospital,Department of Gastroenterology
[3] The Royal London Hospital,Department of Gastroenterology
关键词
Colonoscopy; Endoscopy; Force measurement;
D O I
暂无
中图分类号
学科分类号
摘要
Colonoscopy involves advancing a flexible endoscope into and along the entire length of the colon. The procedure can be painful and carries the risk of perforating the organ, yet very little is known of the forces involved. A device to measure the forces exerted on the endoscope during colonoscopy is described. The device features a handle designed on the shape of a hinged split cylinder that locks around the endoscope, gripping it tightly. The handle has two parts, an inner part that grips the endoscope, and an outer part that is gripped by the endoscopist. The two parts are jointed together by members that transmit the forces through to the endoscope. One of the members incorporates strain gauges that measure the torque applied to the endoscope, as well as the push and pull forces. The handle can easily be unlocked and moved along the endoscope as the colonoscopy proceeds. The device is used to measure the forces applied to the endoscope during 11 routine colonoscopies, and summary results are presented. These are believed to be the first accurate measurements of the forces exerted during colonoscopy.
引用
收藏
页码:186 / 190
页数:4
相关论文
共 50 条
  • [41] THE FORCES EXERTED ON THE DUODENUM AND SPHINCTER OF ODDI DURING ERCP, AND SPHINCTEROTOMY
    RATANI, RS
    SWAIN, CP
    GASTROINTESTINAL ENDOSCOPY, 1993, 39 (02) : 334 - 334
  • [42] Forces exerted by a conventional dental explorer during clinical examination
    Wagner, J
    Thomas, G
    Stanford, C
    CARIES RESEARCH, 2003, 37 (05) : 365 - 368
  • [43] CONTROL OF FORCES EXERTED DURING SURGICAL-CORRECTION OF SCOLIOSIS
    ARCAN, M
    ROBIN, G
    SIMKIN, A
    ISRAEL JOURNAL OF TECHNOLOGY, 1975, 13 (03): : 225 - 228
  • [44] Traction forces exerted by multicellular clusters during stretch.
    Casares Garcia, L.
    Navajas, D.
    Trepat, X.
    MOLECULAR BIOLOGY OF THE CELL, 2012, 23
  • [45] MEASUREMENT OF FORCES EXERTED ON STARTING BLOCKS DURING SPRINT EVENTS
    ROSTAD, B
    MCCLEMENTS, J
    GANDER, RE
    SANDERSON, L
    CANADIAN MEDICAL AND BIOLOGICAL ENGINEERING SOCIETY CONFERENCE : PROCEEDINGS - 1989, 1989, : 151 - 152
  • [46] A comparison of the forces exerted by 20 laryngoscope blades during laryngoscopy
    Rassam, S
    Hall, JE
    Mecklenburgh, J
    Wilkes, AR
    BRITISH JOURNAL OF ANAESTHESIA, 2004, 93 (04) : 613P - 613P
  • [47] FORCES EXERTED DURING SWING-THROUGH CRUTCH GAIT
    BURDETT, RG
    REISMAN, M
    SIMON, SR
    PHYSICAL THERAPY, 1983, 63 (05): : 772 - 772
  • [48] PNEUMATIC DEVICE FOR THE AUTOMATED ADVANCEMENT OF THE FIBEROPTIC ENDOSCOPE FOR TOTAL COLONOSCOPY - A PRELIMINARY-REPORT
    SUGARBAKER, PH
    PENLAND, WZ
    LYDDY, J
    GASTROINTESTINAL ENDOSCOPY, 1985, 31 (03) : 210 - 213
  • [49] Contactless Measuring Device for Flexible Shaft Coupling Twist Angle
    Urbansky, Matej
    Grega, Robert
    CURRENT METHODS OF CONSTRUCTION DESIGN, 2020, : 193 - 200
  • [50] Model-based estimation of muscle forces exerted during movements
    Erdemir, Ahmet
    McLean, Scott
    Herzog, Walter
    van den Bogert, Antonle J.
    CLINICAL BIOMECHANICS, 2007, 22 (02) : 131 - 154