A New Calibration Apparatus of Port Quick Leak Detector for Spacecraft

被引:0
|
作者
Lan Zhao
Yongjun Cheng
Wenjun Sun
Yongjun Wang
Lian Chen
Xuemin Sheng
Ruifang Zhang
Meiru Guo
Yanwu Li
机构
[1] Lanzhou Institute of Physics,Science and Technology on Vacuum Technology and Physics Laboratory
来源
MAPAN | 2018年 / 33卷
关键词
Calibration apparatus; Port quick leak detector; Direct comparison calibration method; Pressure-rising method;
D O I
暂无
中图分类号
学科分类号
摘要
The leak rate of spacecraft port must meet certain standards during the opening and closing processes. The port quick leak detector can be used to on-line measure leak rates of spacecraft port. In this work, a new calibration apparatus is designed and investigated. It can work at different temperatures with different gases in a wide leak rate range. This apparatus consists of the standard leak rate system, calibration system, temperature control system, gas supply and pumping system, which is designed on the basis of direct comparison calibration method and pressure-rising method with a constant volume, and the leak rates are provided by the standard leak rate system. The calibration range of the calibration apparatus is from 2 × 10–6 to 2 × 10–2 Pa m3/s. Its combined standard uncertainty is about 4.7%, and its calibration temperature is from − 50 to 20 °C. Within the above leak rate and temperature ranges, the correction factor of the port quick leak detector is about 0.88–1.09.
引用
收藏
页码:91 / 97
页数:6
相关论文
共 50 条
  • [1] A New Calibration Apparatus of Port Quick Leak Detector for Spacecraft
    Zhao, L.
    Cheng, Y.
    Sun, W.
    Wang, Y.
    Chen, L.
    Sheng, X.
    Zhang, R.
    Guo, M.
    Li, Y.
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2018, 33 (02): : 91 - 97
  • [2] Study on Calibration Technology of Port Quick Leak Detector for Spacecraft
    Zhao, Lan
    Cheng, Yongjun
    Sun, Wenjun
    Wang, Yongjun
    Li, Yanwu
    2017 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR AEROSPACE (METROAEROSPACE), 2017, : 546 - 549
  • [3] Reliability test and evaluating methods of port quick leak detector for manned spacecraft
    Cui, Z., 1600, Chinese Academy of Space Technology (32):
  • [4] DETECTOR CALIBRATION WITH AN ASSOCIATED PARTICLE APPARATUS
    MEIER, MM
    CARLSON, AD
    LAMAZE, GP
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1975, 20 (02): : 137 - 137
  • [5] Leak rate calibration with gas flow standard apparatus
    Zhang, DX
    Liu, SL
    Li, DT
    Yang, M
    Liu, QA
    Li, WK
    VACUUM, 1996, 47 (6-8) : 515 - 517
  • [6] A NEW GROSS LEAK CALIBRATION FOR LEAK DETECTORS
    SCUOTTO, F
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1986, 4 (05): : 2287 - 2292
  • [7] 85KR SCINTILLATION DETECTOR FOR MEASURING SPACECRAFT LEAK RATES
    RUSCH, RJ
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1968, 11 (02): : 467 - &
  • [8] Leak-detector and its calibration technology for docking system
    Li, Detian
    Yang, Xiaolin
    Cheng, Yunjun
    Sheng, Xuemin
    2016 IEEE METROLOGY FOR AEROSPACE (METROAEROSPACE), 2016, : 398 - 402
  • [9] ION GAUGE CALIBRATION AT LOW PRESSURES USING LEAK DETECTOR
    BERGSNOVHANSON, B
    ENDOW, N
    PASTERNAK, RA
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY, 1964, 1 (01): : 7 - &
  • [10] REFERENCE CUMULATIVE APPARATUS FOR THE CALIBRATION OF ARTIFICIAL LEAKS (FOR VACUUM LEAK DETECTORS)
    KUZMIN, VV
    SOVIET JOURNAL OF NONDESTRUCTIVE TESTING-USSR, 1989, 25 (10): : 738 - 742