Blood flow measurement in reconstructed anterior cruciate ligaments using laser Doppler flowmetry

被引:10
|
作者
Yamagishi T. [1 ]
Fujii K. [1 ]
Roppongi S. [1 ]
Hatsuumi H. [1 ]
机构
[1] Department of Orthopaedic Surgery, Jikei University, School of Medicine, 3-25-8 Nishishinbashi, Minato-ku, Tokyo
关键词
Anterior cruciate ligament reconstruction; Blood cell flux; Laser Doppler flowmetry; Magnetic resonance imaging evaluation; Second-look arthroscopy;
D O I
10.1007/s001670050092
中图分类号
学科分类号
摘要
Laser Doppler flowmetry (LDF) was used to measure blood flow in anterior cruciate ligaments (ACL) reconstructed using a bone-patellar tendon-bone autograft. Thirty-five patients (17 men, 18 women) undergoing second-look arthroscopy for a reconstructed ACL were selected at random for postoperative participation in the study at 6, 12, and 18 months. Eight patients with an intact ACL had their blood flow measured arthroscopically for a control. Under arthroscopic visualization, a 3 mm probe was placed through a trocar sleeve into the ACL. Functional flow was evaluated using LDF, for which the output signal, the blood cell flux (BCF), is expressed in terms of volts. These patients also underwent magnetic resonance imaging (MRI); MRI, IKDC final evaluation, and second-look arthroscopic findings for each patient were compared with the others. In normal ACL, the BCF value ranged from 120 to 130 mV, and synovial blood flow, 90-132 mV. Significantly high BCF values for the reconstructed ACL were noted at 6 and 12 months, but there was no change in synovial BCF postoperatively. There was a gradual return to near normal BCF values for the reconstructed ACL 18 months after surgery. Significantly abnormal BCF values (580 ± 20 mV) were seen in cases with a severely abnormal IKDC final evaluation; there were few differences in BCF values between normal, nearly normal, and abnormal. LDF is easy to use and appears to be a reproducible technique for evaluating blood flow in the reconstructed ACL, offering distinct advantages for evaluating its maturation. Six months after surgery, there was a decrease in MRI with a high signal intensity, whereas the BCF values in the reconstructed ACL generally needed 18 months to return to near normal; the return was gradual. We believe the blood supply of the reconstructed ACL may originate from the synovium of posterior joint capsule within 6 months after surgery.
引用
收藏
页码:160 / 164
页数:4
相关论文
共 50 条
  • [31] ASSESSMENT OF GASTRIC BLOOD-FLOW WITH LASER DOPPLER FLOWMETRY
    AHN, H
    IVARSSON, LE
    JOHANSSON, K
    LINDHAGEN, J
    LUNDGREN, O
    SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1988, 23 (10) : 1203 - 1210
  • [32] TENDON BLOOD-FLOW ASSESSED BY LASER DOPPLER FLOWMETRY
    ASTROM, M
    SVENSSON, H
    SCANDINAVIAN JOURNAL OF PLASTIC AND RECONSTRUCTIVE SURGERY AND HAND SURGERY, 1991, 25 (03): : 213 - 215
  • [33] Measurement of median nerve blood flow during carpal tunnel release with laser Doppler flowmetry
    Soejima, O
    Iida, H
    Naito, M
    MINIMALLY INVASIVE NEUROSURGERY, 2001, 44 (04) : 202 - 204
  • [34] Laser Doppler flowmetry and optic nerve head blood flow
    Petrig, BL
    Riva, CE
    Hayreh, SS
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1999, 127 (04) : 413 - 425
  • [35] CHOROIDAL BLOOD-FLOW BY LASER-DOPPLER FLOWMETRY
    RIVA, CE
    PETRIG, BL
    OPTICAL ENGINEERING, 1995, 34 (03) : 746 - 752
  • [36] VALIDATION OF LASER-DOPPLER FLOWMETRY IN MEASUREMENT OF SPINAL-CORD BLOOD-FLOW
    LINDSBERG, PJ
    ONEILL, JT
    PAAKKARI, IA
    HALLENBECK, JM
    FEUERSTEIN, G
    AMERICAN JOURNAL OF PHYSIOLOGY, 1989, 257 (02): : H674 - H680
  • [37] THE CHARACTERISTICS OF LASER-DOPPLER FLOWMETRY FOR THE MEASUREMENT OF REGIONAL CEREBRAL BLOOD-FLOW - COMMENT
    BAKER, CJ
    SOLOMON, RA
    NEUROSURGERY, 1995, 36 (02) : 364 - 364
  • [38] Laser Doppler flowmetry is valid for measurement of cerebral blood flow autoregulation lower limit in rats
    Tonnesen, J
    Pryds, A
    Larsen, EH
    Paulson, OB
    Hauerberg, J
    Knudsen, GM
    EXPERIMENTAL PHYSIOLOGY, 2005, 90 (03) : 349 - 355
  • [39] CONTINUOUS MEASUREMENT OF GASTRIC BLOOD-FLOW BY LASER-DOPPLER FLOWMETRY DURING GASTROSCOPY
    CHRISTOFORIDIS, EC
    HOVENDAL, C
    BJERRING, P
    KRUSE, A
    SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1989, 24 (01) : 16 - 20
  • [40] Assessment of Retinal/Choroidal Blood Flow Using Laser Doppler Flowmetry (LDF) in Rats
    Hetu, S.
    Cordahi, G.
    Couture, R.
    Vaucher, E.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)