Detection of bubble nucleation event in superheated drop detector by the pressure sensor

被引:0
|
作者
MALA DAS
NILANJAN BISWAS
机构
[1] Saha Institute of Nuclear Physics,Astroparticle Physics and Cosmology Division
来源
Pramana | 2017年 / 88卷
关键词
Bubble nucleation; pressure sensor; superheated liquid drop.; 07.07.Df; 07.64. ; z; 29.40. ; n;
D O I
暂无
中图分类号
学科分类号
摘要
Superheated drop detector consisting of drops of superheated liquid suspended in polymer or gel matrix is of great demand, mainly because of its insensitivity to ß-particles and ?-rays and also because of the low cost. The bubble nucleation event is detected by measuring the acoustic shock wave released during the nucleation process. The present work demonstrates the detection of bubble nucleation events by using the pressure sensor. The associated circuits for the measurement are described in this article. The detection of events is verified by measuring the events with the acoustic sensor. The measurement was done using drops of various sizes to study the effect of the size of the drop on the pressure recovery time. Probability of detection of events has increased for larger size of the superheated drops and lesser volume of air in contact with the gel matrix. The exponential decay fitting to the pressure sensor signals shows the dead time for pressure recovery of such a drop detector to be a few microseconds.
引用
收藏
相关论文
共 50 条
  • [1] Detection of bubble nucleation event in superheated drop detector by the pressure sensor
    Das, Mala
    Biswas, Nilanjan
    PRAMANA-JOURNAL OF PHYSICS, 2017, 88 (01):
  • [2] Observation of acoustically silent bubble-nucleation events in superheated drop detector
    Chatterjee, B. K.
    Sarkar, R.
    Mondal, P. K.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2025, 1073
  • [3] A new optical method for the detection of bubble nucleation in superheated droplet detector
    Sarkar, Rupa
    Mondal, Prasanna Kumar
    Datta, Manas
    Chatterjee, Barun Kumar
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2017, 88 (06):
  • [4] Advances in superheated drop (bubble) detector techniques
    dErrico, F
    Alberts, WG
    Matzke, M
    RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) : 103 - 108
  • [5] Pressure dependence of neutron detection sensitivity in a superheated drop detector
    Tsuda, S
    Sawamura, T
    Homma, A
    Narita, M
    RADIATION PROTECTION DOSIMETRY, 2000, 87 (02) : 87 - 92
  • [6] SUPERHEATED DROP NUCLEATION FOR NEUTRON DETECTION
    APFEL, RE
    CHU, BT
    MENGEL, J
    APPLIED SCIENTIFIC RESEARCH, 1982, 38 : 117 - 122
  • [7] The nucleation parameter for heavy-ion induced bubble nucleation in superheated emulsion detector
    Seth, S.
    Das, M.
    Bhattacharya, S.
    Bhattacharjee, P.
    Saha, S.
    JOURNAL OF INSTRUMENTATION, 2013, 8
  • [8] Study of the nucleation in superheated emulsion detector at high pressure
    Sarkar, R.
    Datta, M.
    Chatterjee, B. K.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 682 : 31 - 35
  • [9] SUPERHEATED DROP DETECTOR
    APFEL, RE
    NUCLEAR INSTRUMENTS & METHODS, 1979, 162 (1-3): : 603 - 608
  • [10] An active device for volumetric measurement of drop nucleation in a superheated emulsion detector
    Mondal, P. K.
    Chatterjee, B. K.