Comparative evaluation of Transient Evoked Oto-acoustic Emissions and Brainstem Evoked Response Audiometry as screening modality for hearing impairment in neonates

被引:7
|
作者
Dhawan R. [1 ]
Mathur N.N. [1 ,2 ]
机构
[1] Department of ENT and Head Neck Surgery, Lady Hardinge Medical College and Associated, SK Hospital and KS Children's Hospital
[2] Delhi 110092
关键词
BERA; Hearing impairment; Screening; TEOAE;
D O I
10.1007/s12070-007-0004-7
中图分类号
学科分类号
摘要
Objective: This study was designed to evaluate Transient Evoked Oto-acoustic Emission (TEOAE) as screening modality for hearing impairment in neonates. Brainstem Evoked Response Auditometry (BERA) was used as gold standard diagnostic tool in this study. The factors affecting the specificity of TEOAE were also studied. Methods: The study group of 200 randomly selected neonates was subjected to TEOAE and BERA (400 ears). Oto-endoscopy was done in all TEOAE failures and a repeat test was done after suction cleaning of blocked external auditory canal (EAC). Results: Otoscopic evaluation of all 52 TEOAE failures was done. EAC obstruction was noticed in 31 ears and 4 ears showed collapsible EAC. TEOAE was repeated after suction cleaning of the obstructed EAC and using long probe tips for collapsible EAC. This improved the Pass rate of TEOAE from 87% to 92%. EAC obstruction and collapsible EAC were the two factors identified in this study that significantly affected the specificity of TEOAE as a screening test. Pass rate of TEOAE in <48 hrs age group was found to be 55.5%, which was nearly half of over-all pass rate. This was because of high prevalence of obstructed EAC in this age group. TEOAE was found to be a rapid screening tool as average time taken for BERA was 35 min/neonate and for TEOAE was 17.4 min/neonate. Acceptability of TEOAE was found to be higher as compared to BERA. Conclusions: TEOAE is a simple and rapid test with relatively higher acceptability. But, the low sensitivity and specificity are the main shortcomings that take away from TEOAE, the status of independent screening modality for hearing impairment in neonates. TEOAE cannot completely replace BERA as screening modality for hearing impairment in neonates, however can complement it. © 2007 Association of Otolaryngologists of India.
引用
收藏
页码:15 / 18
页数:3
相关论文
共 50 条
  • [31] Evaluation of hearing loss in at risk neonates by evoked otoacoustic emissions
    Ashraf Husain
    [J]. Indian Journal of Otolaryngology and Head and Neck Surgery, 1997, 49 (Suppl 1): : 32 - 35
  • [32] Transient-evoked and 2F1-F2 distortion product oto-acoustic emissions in dogs:: Preliminary findings
    Sockalingam, R
    Filippich, L
    Sommerlad, S
    Murdoch, B
    Charles, B
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 1998, 3 (06) : 373 - 385
  • [33] USING TRANSIENT EVOKED OTOACOUSTIC EMISSIONS FOR NEONATAL HEARING SCREENING
    MAXON, AB
    WHITE, KR
    VOHR, BR
    BEHRENS, TR
    [J]. BRITISH JOURNAL OF AUDIOLOGY, 1993, 27 (02): : 149 - 153
  • [34] Universal Hearing Screening in Newborns Using Otoacoustic Emissions and Brainstem Evoked Response in Eastern Uttar Pradesh
    Kumar A.
    Gupta S.C.
    Sinha V.R.
    [J]. Indian Journal of Otolaryngology and Head & Neck Surgery, 2017, 69 (3) : 296 - 299
  • [35] BRAIN-STEM EVOKED-RESPONSE AUDIOMETRY IN NEWBORN HEARING SCREENING
    SCHULMANGALAMBOS, C
    GALAMBOS, R
    [J]. ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1979, 105 (02) : 86 - 90
  • [36] Optimizing hearing screening in newborn infants with transient evoked otoacoustic emissions
    WelzlMuller, K
    Boheim, K
    Stephan, K
    Schlogel, H
    Stadlmann, A
    Nekahm, D
    [J]. HNO, 1997, 45 (04) : 227 - 232
  • [37] Neonatal hearing screening with transient evoked otoacoustic emissions:: A learning curve
    De Ceulaer, G
    Daemers, K
    Van Driessche, K
    Mariën, S
    Somers, T
    Offeciers, FE
    Govaerts, PJ
    [J]. AUDIOLOGY, 1999, 38 (06): : 296 - 302
  • [38] Universal newborn hearing screening; automated transient evoked otoacoustic emissions
    Arslan, S.
    Isik, A. U.
    Imamoglu, M.
    Topbas, M.
    Aslan, Y.
    Ural, A.
    [J]. B-ENT, 2013, 9 (02) : 123 - 131
  • [39] The results of newborn hearing screening by means of transient evoked otoacoustic emissions
    Wroblewska-Seniuk, K
    Chojnacka, K
    Pucher, B
    Szczapa, J
    Gadzinowski, J
    Grzegorowski, M
    [J]. INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2005, 69 (10) : 1351 - 1357
  • [40] 4-aminopyridine in scala media reversibly alters the cochlear potentials and suppresses electrically evoked oto-acoustic emissions
    Kirk, DL
    Yates, GK
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 1998, 3 (01) : 21 - 39