Comparison of a hydrophilic and a hydrophobic apodized diffractive multifocal intraocular lens

被引:0
|
作者
Jan Willem van der Linden
Ivanka J. van der Meulen
Maarten P. Mourits
Ruth Lapid-Gortzak
机构
[1] Academic Medical Center,Department of Ophthalmology
[2] University of Amsterdam,undefined
[3] Retina Total Eye Care,undefined
来源
关键词
Anterior Chamber Depth; Corrected Distance Visual Acuity; Posterior Capsular Opacification; Uncorrected Distance Visual Acuity; Diffractive Lens;
D O I
暂无
中图分类号
学科分类号
摘要
To compare outcomes between a new design apodized diffractive hydrophilic multifocal intraocular lens (IOL) (Seelens MF; study group), and a well-known apodized diffractive hydrophobic multifocal IOL (SN6AD1; control group). A comparative case series comparing refractive and visual outcomes at distance and near. Patient satisfaction with a validated questionnaire, dysphotopsia and straylight measurement scores were recorded at 3 months post-operatively. The study group comprised 48 eyes and the control group 37 eyes. At 3 months post-operatively the mean uncorrected distance visual acuity (UDVA) was not statistically significant different between the study group and the control group (0.02 ± 0.07 logMAR [SD] vs 0.04 ± 0.09 logMAR). Corrected distance visual acuity (CDVA) was statistically significantly better with the study lens (−0.04 ± 0.05 logMAR vs −0.01 ± 0.04 logMAR (p < 0.019). There was no clinical or statistical significant difference at the 40 cm distance (0.09 ± 0.12 logMAR vs 0.08 ± 0.09 logMAR). The study group had statistically significant better uncorrected near acuity at 50 and 60 cm distances (p < 0.03 and p < 0.007, respectively). In terms of satisfaction the lenses performed equally. Halos were seen less often with the study lens. Straylight, as a parameter for visual quality, was significantly less with the study lens. Conclusion: The Seelens MF performs equally as well as the well-known SN6AD1 for UCDA and CDVA. The Seelens MF performs better at intermediate distance, and seems to allow for better depth of focus, and increased visual quality. More study is needed to corroborate the last finding.
引用
收藏
页码:493 / 500
页数:7
相关论文
共 50 条
  • [31] European multicenter study of the AcrySof ReSTOR apodized diffractive intraocular lens
    Kohnen, T
    Allen, D
    Boureau, C
    Dublineau, P
    Hartmann, C
    Mehdorn, E
    Rozot, P
    Tassinari, G
    OPHTHALMOLOGY, 2006, 113 (04) : 578 - 584
  • [32] Clinical outcomes of a new diffractive multifocal intraocular lens
    Toygar, Baha
    Kiziloglu, Ozge Yabas
    Toygar, Okan
    Hacimustafaoglu, Ali Murat
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2017, 10 (12) : 1844 - 1850
  • [33] Visual outcomes after progressive apodized diffractive intraocular lens implantation
    Garcia-Bella, Javier
    Ventura-Abreu, Nestor
    Morales-Fernandez, Laura
    Talavero-Gonzalez, Paula
    Carballo-Alvarez, Jesus
    Carlos Sanz-Fernandez, Juan
    Vazquez-Molini, Jose M.
    Martinez-de-la-Case, Jose M.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2018, 28 (03) : 282 - 286
  • [34] Near add power of the diffractive multifocal intraocular lens
    Tchah, H
    Choi, WJ
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (09): : 1232 - 1235
  • [35] Medium-Term Visual Outcomes of Apodized Diffractive Multifocal Intraocular Lens with+3.00 D Addition Power
    Guo, Xiaohong
    Sun, Yi
    Zhang, Bowen
    Zheng, Danying
    JOURNAL OF OPHTHALMOLOGY, 2014, 2014
  • [36] Diffractive multifocal intraocular lens implantation in patients with monofocal intraocular lens in the contralateral eye
    Jae Yong Kim
    Yunhan Lee
    Hun Jae Won
    Hyerin Jeong
    Jin Hyoung Park
    Myoung Joon Kim
    Hungwon Tchah
    International Journal of Ophthalmology, 2020, (05) : 737 - 743
  • [37] Diffractive multifocal intraocular lens implantation in patients with monofocal intraocular lens in the contralateral eye
    Kim, Jae Yong
    Lee, Yunhan
    Won, Hun Jae
    Jeong, Hyerin
    Park, Jin Hyoung
    Kim, Myoung Joon
    Tchah, Hungwon
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2020, 13 (05) : 737 - 743
  • [38] Near and intermediate visual and reading performance of patients with a multifocal apodized diffractive intraocular lens using an electronic reading desk
    Attia, Mary S. A.
    Khoramnia, Ramin
    Auffarth, Gerd U.
    Kirchner, Merietta
    Holzer, Mike P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (04): : 582 - 590
  • [39] Optical quality evaluation and vision simulation of a binocular diffractive multifocal intraocular lens system vs. a conventional diffractive multifocal intraocular lens
    Naujokaitis, Tadas
    Labuz, Grzegorz
    Auffarth, Gerd U.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [40] Comparative Visual Outcome Analysis of a Diffractive Multifocal Intraocular Lens and a New Diffractive Multifocal Lens with Extended Depth of Focus
    Moshirfar, Majid
    Stapley, Seth R.
    Corbin, Wyatt M.
    Bundogji, Nour
    Conley, Matthew
    Darquea, Ines M.
    Ronquillo, Yasmyne C.
    Hoopes, Phillip C.
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (24)