Determination of trace Pb(II), Cd(II) and Zn(II) using differential pulse stripping voltammetry without Hg modification

被引:0
|
作者
Qi Lin
HongMei Lin
YuanHui Zhang
MingCong Rong
HuiXian Ke
XinHua Tang
Xi Chen
机构
[1] SOA,Key Lab of Global Change and Marine
[2] SOA,Atmospheric Chemistry
[3] Xiamen University,Third Institute of Oceanography
[4] Xiamen Standard Scientific Instrument Co. Ltd.,Department of Chemistry, College of Chemistry and Chemical Engineering
来源
Science China Chemistry | 2013年 / 56卷
关键词
bismuth film electrode; differential pulse stripping voltammetry; tea leaves; tea infusion;
D O I
暂无
中图分类号
学科分类号
摘要
In this work, we reported a simultaneous determination approach for Pb(II), Cd(II) and Zn(II) at μg L−1 concentration levels using differential pulse stripping voltammetry on a bismuth film electrode (BiFE). The BiFE could be prepared in situ when the sample solution contained a suitable amount of Bi(NO)3, and its analytical performance was evaluated for the simultaneous determination of Pb(II), Cd(II) and Zn(II) in solutions. The determination limits were found to be 0.19 μg L−1 for Zn(II), and 0.28 μg L−1 for Pb(II) and Cd(II), with a preconcentration time of 300 s. The BiFE approach was successfully applied to determine Pb(II), Cd(II) and Zn(II) in tea leaf and infusion samples, and the results were in agreement with those obtained using an atomic absorption spectrometry approach. Without Hg usage, the in situ preparation for BiFE supplied a green and acceptability sensitive method for the determination of the heavy metal ions.
引用
收藏
页码:1749 / 1756
页数:7
相关论文
共 50 条