Quality assurance for diffusion tensor imaging using an ACR phantom: Comparative analysis with 6, 15, and 32 directions at 1.5T and 3.0T MRI systems

被引:0
|
作者
Jung-Hoon Lee
Sang-Young Kim
Do-Wan Lee
Jin-Young Jung
Kyu-Ho Song
Bo-Young Choe
机构
[1] College of Medicine,Department of Biomedical Engineering, Research Institute of Biomedical Engineering
[2] The Catholic University of Korea,Department of Radiology
[3] Kyunghee Medical Center,undefined
来源
关键词
Magnetic resonance imaging; Diffusion tensor imaging; ACR MRI phantom;
D O I
暂无
中图分类号
学科分类号
摘要
Although diffusion tensor imaging (DTI) has been widely used for the quantitative analyses of the integrity of white matter in the brain in clinical and research fields, quality assurance (QA) for DTI has not been fully established. Thus, we suggest a QA guideline for DTI using the American College of Radiology (ACR) Magnetic resonance imaging (MRI) head phantom. In this study, the geometric accuracy, slice-position accuracy, image intensity uniformity, percent signalghosting, low-contrast object detectability, image distortion, fractional anisotropy (FA), and apparent diffusion coefficient (ADC) were measured and evaluated in 1.5T and 3.0T MRI scanners equipped with an 8-channel SENSE head coil. The standard axial spin echo (SE) T1-weighted MR images and DTI with 6, 15 and 32 directions were obtained. Concerning geometric accuracy, image twisting in the three directions was observed due to the inhomogeneity of echo planar imaging (EPI). Image intensity uniformity was significantly lower for DTI than for the standard SE T1-weighted MR images. Percent signal ghosting was higher for images from 3.0T MRI than for images from 1.5T MRI. Low-contrast object detectability was visually identified and measured at a low contrasttonoise ratio (CNR) and a low signaltonoise ratio (SNR). Image distortion changed remarkably to the phaseencoding direction. The present study using the ACR MRI phantom suggests a QA method for DTI with high reproducibility and easy accessibility.
引用
收藏
页码:103 / 110
页数:7
相关论文
共 38 条
  • [21] Evaluation of a Polyethylene Glycol Phantom for Measuring Apparent Diffusion Coefficients Using Three 3.0 T MRI Systems
    Eisuke Sato
    Kei Fukuzawa
    Hiroyuki Takashima
    Yuya Yamatani
    Yasuo Takatsu
    Junichi Hata
    Keigo Hikishima
    Kenta Miwa
    Applied Magnetic Resonance, 2021, 52 : 619 - 631
  • [22] Evaluation of a Polyethylene Glycol Phantom for Measuring Apparent Diffusion Coefficients Using Three 3.0 T MRI Systems
    Sato, Eisuke
    Fukuzawa, Kei
    Takashima, Hiroyuki
    Yamatani, Yuya
    Takatsu, Yasuo
    Hata, Junichi
    Hikishima, Keigo
    Miwa, Kenta
    APPLIED MAGNETIC RESONANCE, 2021, 52 (05) : 619 - 631
  • [23] Machine learning classificatory as a tool in the diagnosis of amyotrophic lateral sclerosis using diffusion tensor imaging parameters collected with 1.5T MRI scanner: A case study
    Jamrozy, Milosz
    Maj, Edyta
    Bielecki, Maksymilian
    Bartoszek, Marta
    Golebiowski, Marek
    Kuzma-Kozakiewicz, Magdalena
    ELECTRONIC JOURNAL OF GENERAL MEDICINE, 2023, 20 (06):
  • [24] Correlation of Patellofemoral Chondromalacia and Body Mass Index (BMI) in Relation to Sex and Age Analysis of 1.5T and 3.0T Magnetic Resonance (MR) Images Using the Outerbridge Scale
    Sieron, Dominik
    Jablonska, Izabela
    Niemiec, Pawel
    Lukoszek, Dawid
    Szyluk, Karol
    Platzek, Ivan
    Meusburger, Hugo
    Delimpasis, Georgios
    Christe, Andreas
    MEDICAL SCIENCE MONITOR, 2022, 28
  • [25] Analysis of variability of fractional anisotropy values at 3T using a novel diffusion tensor imaging phantom
    Proyenzale, James M.
    Taylor, Brian A.
    Wilde, Elisabeth A.
    Boss, Michael
    Schneider, Walter
    NEURORADIOLOGY JOURNAL, 2018, 31 (06): : 581 - 586
  • [26] Serial assessments of delayed encephalopathy after carbon monoxide poisoning using magnetic resonance spectroscopy and diffusion tensor imaging on 3.0T system
    Terajima, Kenshi
    Igarashi, Hironaka
    Hirose, Masaki
    Matsuzawa, Hitoshi
    Nishizawa, Masatoyo
    Nakada, Tsutomu
    EUROPEAN NEUROLOGY, 2008, 59 (1-2) : 55 - 61
  • [27] Reconstruction of white matter fibre tracts using diffusion kurtosis tensor imaging at 1.5T: Pre-surgical planning in patients with gliomas
    Leote, Joao
    Nunes, Rita G.
    Cerqueira, Luis
    Loucao, Ricardo
    Ferreira, Hugo A.
    EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2018, 5 : 20 - 23
  • [28] Direct comparison of the quality of 32 echo MRI T2 decays used to measure myelin water in vivo at 3T and 1.5T using the decay roughness measure
    Cover, K. S.
    de Graaf, W.
    Soriano, A. Lopez
    Kuijer, J. P. A.
    Steenwijk, M. D.
    Geurts, J. J. G.
    van Dijk, B. W.
    Vrenken, H.
    Barkhof, F.
    MULTIPLE SCLEROSIS JOURNAL, 2011, 17 : S371 - S372
  • [30] 1H magnetic resonance spectroscopy and diffusion weighted imaging findings of medulloblastoma in 3.0T MRI A retrospective analysis of 17 cases
    Wu, Guangyao
    Pang, Haopeng
    Ghimire, Prasanna
    Liu, Guobing
    NEURAL REGENERATION RESEARCH, 2012, 7 (32) : 2554 - 2559