Development of a wide-range and fast-response digitizing pulse signal acquisition and processing system for neutron flux monitoring on EAST

被引:0
|
作者
Li Yang
Hong-Rui Cao
Jin-Long Zhao
Zi-Han Zhang
Qiang Li
Guo-Bin Wu
Yong-Qiang Zhang
Guo-Qiang Zhong
Li-Qun Hu
Zi-Jun Zhang
机构
[1] Chinese Academy of Sciences,Institute of Plasma Physics, Hefei Institutes of Physical Science
[2] University of Science and Technology of China,undefined
[3] Hefei University of Technology,undefined
来源
关键词
EAST; Neutron flux monitoring; High time resolution; Wide range; Pulse signal acquisition and processing system;
D O I
暂无
中图分类号
学科分类号
摘要
The neutron count rate fluctuation reaches six orders of magnitude between the ohmic plasma scenario and high power of auxiliary heating on an experimental advanced superconducting tokamak (EAST). The measurement result of neutron flux monitoring (NFM) is a significant feedback parameter related to the acquisition of radiation protection-related information and rapid fluctuations in neutron emission induced by plasma magnetohydrodynamic activity. Therefore, a wide range and high time resolution are required for the NFM system on EAST. To satisfy these requirements, a digital pulse signal acquisition and processing system with a wide dynamic range and fast response time was developed. The present study was conducted using a field-programmable gate array (FPGA) and peripheral component interconnect extension for instrument express (PXIe) platform. The digital dual measurement modes, which are composed of the pulse-counting mode and AC coupled square integral’s Campbelling mode, were designed to expand the measurement range of the signal acquisition and processing system. The time resolution of the signal acquisition and processing system was improved from 10 to 1 ms owing to utilizing high-speed analog-to-digital converters (ADCs), a high-speed PXIe communication with a direct memory access (DMA) mode, and online data preprocessing technology of FPGA. The signal acquisition and processing system was tested experimentally in the EAST radiation field. The test results showed that the time resolution of NFM was improved to 1 ms, and the dynamic range of the neutron counts rate was expanded to more than 106\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$10^{6}$$\end{document} counts per second. The Campbelling mode was calibrated using a multipoint average linear fitting method; subsequently, the fitting coefficient reached 0.9911. Therefore, the newly developed pulse signal acquisition and processing system ensures that the NFM system meets the requirements of high-parameter experiments conducted on EAST more effectively.
引用
收藏
相关论文
共 24 条
  • [21] High-precision and wide-range real-time neutron flux monitor system through multipoint linear calibration
    Ren-Jie Zhu
    Xiang Zhou
    Zi-Hao Liu
    Wen-Di Wang
    Xiao-Li Mou
    Teng-Fei Fang
    Qing-Li Ma
    Xiu-Feng Xu
    Guo-Liang Yuan
    Li Zhao
    Ling-Feng Wei
    Ze-Jie Yin
    Nuclear Science and Techniques, 2020, 31
  • [22] High-precision and wide-range real-time neutron flux monitor system through multipoint linear calibration
    Zhu, Ren-Jie
    Zhou, Xiang
    Liu, Zi-Hao
    Wang, Wen-Di
    Mou, Xiao-Li
    Fang, Teng-Fei
    Ma, Qing-Li
    Xu, Xiu-Feng
    Yuan, Guo-Liang
    Zhao, Li
    Wei, Ling-Feng
    Yin, Ze-Jie
    NUCLEAR SCIENCE AND TECHNIQUES, 2020, 31 (09)
  • [23] A Wide-Range Low-Jitter PLL Based on Fast-Response VCO and Simplified Straightforward Methodology of Loop Stabilization in Integer-N PLLs
    Kazeminia, Sarang
    Hadidi, Khayrollah
    Khoei, Abdollah
    JOURNAL OF CIRCUITS SYSTEMS AND COMPUTERS, 2015, 24 (07)
  • [24] Development of fast measurement system of neutron emission profile using a digital signal processing technique in JT-60U
    Ishikawa, M.
    Itoga, T.
    Okuji, T.
    Nakhostin, M.
    Shinohara, K.
    Baba, M.
    Nishitani, T.
    BURNING PLASMA DIAGNOSTICS, 2008, 988 : 295 - +