Preparation of a Reference Material for the Determination of Hexavalent Chromium in Tap Water

被引:0
|
作者
Tetsuo Inui
Hamana Shirota
Shoji Sakao
机构
[1] Kanagawa Institute of Industrial Science and Technology,Analytical Chemistry Group, Chemical Engineering Department
来源
Analytical Sciences | 2019年 / 35卷
关键词
Hexavalent chromium; tap water; reference material; diphenylcarbazide absorptiometry; atomic absorption spectrometry;
D O I
暂无
中图分类号
学科分类号
摘要
We developed a reference material (RM) for the determination of hexavalent chromium (Cr(VI)) in tap water. The tap water RM was prepared by adding a Cr(VI) standard solution to the raw material without acidification, i.e., under the original pH conditions of 7.6, because the decrease in the concentration of Cr(VI) was observed when the tap water had been adjusted to pH 1 with HNO3. The prepared tap-water RM (2 L) was packed in 10 fluororesin (PFA) bottles with an inside plug (200 mL each). Each PFA bottle (Cr(VI)-containing tap water) was sealed in a reclosable poly bag and then stored at 5°C in a refrigerator. The tap water RM had a Cr(VI) concentration of 51 μg L−1. The concentration of Cr(VI) was determined by diphenylcarbazide absorptiometry using a 100-mm quartz cell. The detection limit of Cr(VI) in the sample solution corresponding to three-times the standard deviation (n = 5) of blank values was 0.51 μg L−1. The homogeneity of Cr(VI) in the tap water RM was evaluated by an analysis of the variance after the Cochran test. There was no significant difference between the within-bottle and between-bottle variances of the analytical results, indicating that the tap water RM was sufficiently homogeneous. The stability of Cr(VI) in the tap water RM was investigated by monitoring the Cr(VI) concentration over a period of 6 months. The slope of the regression line of the Cr(VI) concentration versus the storage time did not significantly differ from zero, indicating that the tap water RM was stable for 6 months. The concentrations (50–51 μg L−1) of Cr(VI) in the tap water RM were in good agreement with the total chromium concentrations (50–51 μg L−1) obtained by atomic absorption spectrometry.
引用
收藏
页码:1375 / 1379
页数:4
相关论文
共 50 条
  • [41] Determination of water in NIST reference material for mineral oils
    Cedergren, A
    Nordmark, U
    ANALYTICAL CHEMISTRY, 2000, 72 (14) : 3392 - 3395
  • [42] Certified reference material usable in controlling the environment pollution with hexavalent chromium, manganese, chloride and nitrogen/nitrites dioxide
    Debiasi, R
    Nicolau, M
    Buzica, D
    Mitran, E
    REVISTA DE CHIMIE, 2001, 52 (05): : 240 - 244
  • [43] Preparation of chitosan-stabilized Fe0 nanoparticles for removal of hexavalent chromium in water
    Geng, Bing
    Jin, Zhaohui
    Li, Tielong
    Qi, Xinhua
    SCIENCE OF THE TOTAL ENVIRONMENT, 2009, 407 (18) : 4994 - 5000
  • [44] Determination of Chromium Speciation in Tap Water Using Diffusive Gradients in Thin Film Technique
    Guo, Lianwen
    Chen, Hong
    Zhang, Yangyang
    Bo, Le
    Li, Jianrong
    CHEMISTRY LETTERS, 2014, 43 (06) : 849 - 850
  • [45] Preparation of hierarchically nanofibrous membrane and its high adaptability in hexavalent chromium removal from water
    Xu, Guo-Rong
    Wang, Jiao-Na
    Li, Cong-Ju
    CHEMICAL ENGINEERING JOURNAL, 2012, 198 : 310 - 317
  • [46] Adsorptive cathodic stripping voltammetric determination of hexavalent chromium
    Ghandour, MA
    ElShatoury, SA
    Aly, AMM
    Ahmed, SM
    ANALYTICAL LETTERS, 1996, 29 (08) : 1431 - 1445
  • [47] STUDY ON METHOD VALIDATION FOR THE DETERMINATION OF HEXAVALENT CHROMIUM IN CEMENT
    Toader, Margareta
    Ungureanu, Simona
    Martin, Marina
    Cionoiu, Carmen
    PROCEEDINGS OF THE 3RD INTERNATIONAL PROFICIENCY TESTING CONFERENCE, PTCONF 2011, 2011, : 142 - 147
  • [48] Determination of hexavalent chromium in aqueous and alkaline soil extracts
    Voyslavov, Tsvetomil
    Tsakovski, Stefan
    Ganeva, Sonya
    TURKISH JOURNAL OF CHEMISTRY, 2016, 40 (06) : 944 - 952
  • [49] A centrifugal microanalysis system for the determination of nitrite and hexavalent chromium
    LaCroix-Fralish, Angela
    Clare, Jennifer
    Skinner, Cameron D.
    Salin, Eric D.
    TALANTA, 2009, 80 (02) : 670 - 675
  • [50] TITRIMETRIC DETERMINATION OF BORON IN PRESENCE OF COPPER AND HEXAVALENT CHROMIUM
    SCHULZ, WO
    CELIK, G
    HOLZ ALS ROH-UND WERKSTOFF, 1975, 33 (02) : 74 - 78