Effect of Gaussian Smoothing Filter Size for CT-Based Attenuation Correction on Quantitative Assessment of Bone SPECT/CT: A Phantom Study

被引:0
|
作者
Yoya Tomita
Yasutaka Ichikawa
Kengo Hashizume
Hajime Sakuma
机构
[1] Mie University Hospital,Central Division of Radiology
[2] Mie University Hospital,Department of Radiology
关键词
Gaussian filter; CT-based attenuation correction; Attenuation coefficient; Bone quantitative single photon emission computed tomography/computed tomography (SPECT/CT);
D O I
暂无
中图分类号
学科分类号
摘要
This study aims to determine the effect of Gaussian filter size for CT-based attenuation correction (CTAC) on the quantitative assessment of bone SPECT. An experiment was performed using a cylindrical phantom containing six rods, of which one was filled with water and five were filled with various concentrations of K2HPO4 solution (120–960 mg/cm3) to simulate different bone densities. 99mTc-solution of 207 kBq/ml was also included within the rods. SPECT data were acquired at 120 views for 30 s/view. CT for attenuation correction were obtained at 120 kVp and 100 mA. Sixteen different CTAC maps processed with different Gaussian filter sizes (ranging from 0 to 30 mm in 2 mm increments) were generated. SPECT images were reconstructed for each of the 16 CTAC maps. Attenuation coefficients and radioactivity concentrations in the rods were compared with those in the water-filled rod without K2HPO4 solution as a reference. Gaussian filter sizes below 14–16 mm resulted in an overestimation of radioactivity concentrations for rods with high concentrations of K2HPO4 (≥ 666 mg/cm3). The overestimation of radioactivity concentration measurement was 3.8% and 5.5% for 666 mg/cm3 and 960 mg/cm3 K2HPO4 solutions, respectively. The difference in radioactivity concentration between the water rod and the K2HPO4 rods was minimal at 18–22 mm. The use of Gaussian filter sizes smaller than 14–16 mm caused an overestimation of radioactivity concentration in regions of high CT values. Setting the Gaussian filter size to 18–22 mm enables radioactivity concentration to be measured with the least influence on bone density.
引用
收藏
页码:2313 / 2321
页数:8
相关论文
共 50 条
  • [41] The effect of metal artefact reduction on CT-based attenuation correction for PET imaging in the vicinity of metallic hip implants: a phantom study
    Harnish, Roy
    Prevrhal, Sven
    Alavi, Abass
    Zaidi, Habib
    Lang, Thomas F.
    ANNALS OF NUCLEAR MEDICINE, 2014, 28 (06) : 540 - 550
  • [42] Comparative Assessment of Different Energy Mapping Methods in CT-Based Attenuation Correction in PET/CT Systems Using Whole Body XCAT Phantom
    Soleymani, Elham
    Soleymani, Ali
    Ahmadi, Shahla
    Nordin, Md Jan
    4TH INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING AND INFORMATICS (ICEEI 2013), 2013, 11 : 962 - 973
  • [43] CT-Based Attenuation Correction in Brain SPECT/CT Can Improve the Lesion Detectability of Voxel-Based Statistical Analyses
    Kato, Hiroki
    Shimosegawa, Eku
    Fujino, Koichi
    Hatazawa, Jun
    PLOS ONE, 2016, 11 (07):
  • [44] The evaluation of image correction methods for SPECT/CT using CT-based attenuation correction and scatter correction in various radioisotopes with different energy levels
    Shin, Byung Ho
    Kim, Seung Jeong
    Yun, Seok Hwan
    Woo, Jae Ryong
    Oh, So Won
    Kim, Yu Kyeong
    JOURNAL OF NUCLEAR MEDICINE, 2013, 54
  • [45] Intravenous contrast-enhanced CT can be used for CT-based attenuation correction in clinical 111In-octreotide SPECT/CT
    Klausen T.L.
    Mortensen J.
    de Nijs R.
    Andersen F.L.
    Højgaard L.
    Beyer T.
    Holm S.
    EJNMMI Physics, 2 (1) : 1 - 13
  • [46] Effect of CT based attenuation correction on uptake ratios in skeletal SPECT
    Schulz, V.
    Nickel, I.
    Noemayr, A.
    Vija, A. H.
    Hocke, C.
    Hornegger, J.
    Bautz, W.
    Roemer, W.
    Kuwert, T.
    NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2007, 46 (01): : 36 - 42
  • [47] Dose reduction of CT-based attenuation correction in myocardial perfusion CZT-SPECT using iterative CT reconstruction
    van Dalen, J.
    van Osch, J.
    Mouden, M.
    Timmer, J.
    Ottervanger, J.
    Reiffers, S.
    Knollema, S.
    Jager, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S222 - S222
  • [48] Usefulness of the Chang attenuation correction method using CT-based μ map in SPECT-MPI
    Nakamura, Yuya
    Tomiguchi, Seiji
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [49] Impact of reconstruction algorithms and CT-based attenuation correction in I-123-Ioflupane SPECT
    Richetta, E.
    Balma, M.
    Guida, A.
    Codegone, A.
    Manfredi, M.
    Cutaia, C.
    Valzano, S.
    Radici, L.
    Pellerito, R.
    Stasi, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S620 - S620
  • [50] Usefulness of the Chang attenuation correction method using CT-based μ map in SPECT-MPI
    Nakamura, Yuya
    Tomiguchi, Seiji
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55