Importance of re-calibration time on pulse contour analysis agreement with thermodilution measurements of cardiac output: a retrospective analysis of intensive care unit patients

被引:0
|
作者
Christopher G. Scully
Shanti Gomatam
Shawn Forrest
David G. Strauss
机构
[1] U.S. Food and Drug Administration,Office of Science and Engineering Laboratories, Division of Biomedical Physics, Center for Devices and Radiological Health
[2] U.S. Food and Drug Administration,Office of Surveillance and Biometrics, Division of Biostatistics, Center for Devices and Radiological Health
[3] U.S. Food and Drug Administration,Office of Device Evaluation, Division of Cardiovascular Devices, Center for Devices and Radiological Health
关键词
Cardiac output; Pulse contour analysis; Method comparison; Trend monitoring;
D O I
暂无
中图分类号
学科分类号
摘要
We assessed the effect of re-calibration time on cardiac output estimation and trending performance in a retrospective analysis of an intensive care unit patient population using error grid analyses. Paired thermodilution and arterial blood pressure waveform measurements (N = 2141) from 222 patient records were extracted from the Multiparameter Intelligent Monitoring in Intensive Care II database. Pulse contour analysis was performed by implementing a previously reported algorithm at calibration times of 1, 2, 8 and 24 h. Cardiac output estimation agreement was assessed using Bland–Altman and error grid analyses. Trending was assessed by concordance and a 4-Quadrant error grid analysis. Error between pulse contour and thermodilution increased with longer calibration times. Limits of agreement were −1.85 to 1.66 L/min for 1 h maximum calibration time compared to −2.70 to 2.41 L/min for 24 h. Error grid analysis resulted in 74.2 % of points bounded by 20 % error limits of thermodilution measurements for 1 h calibration time compared to 65 % for 24 h. 4-Quadrant error grid analysis showed <75 % of changes in pulse contour estimates to be within ±80 % of the change in the thermodilution measurement at any calibration time. Shorter calibration times improved the agreement of cardiac output pulse contour estimates with thermodilution. Use of minimally invasive pulse contour methods in intensive care monitoring could benefit from prospective studies evaluating calibration protocols. The applied pulse contour analysis method and thermodilution showed poor agreement to monitor changes in cardiac output.
引用
收藏
页码:577 / 586
页数:9
相关论文
共 50 条
  • [21] ASSESSMENT OF THE AGREEMENT BETWEEN PULSE CONTOUR AND TRANSPULMONARY THERMODILUTION FOR CARDIAC OUTPUT MEASUREMENT IN ICU PATIENTS: THE IMPACT OF THERAPEUTIC INTERVENTIONS
    Zoric, L.
    Candela, D.
    Mattatia, L.
    Louart, G.
    Muller, L.
    Lefrant, J. Y.
    INTENSIVE CARE MEDICINE, 2010, 36 : S98 - S98
  • [22] Cardiac output measured by lithium dilution and transpulmonary thermodilution in patients in a paediatric intensive care unit
    R. A. Linton
    M. M. Jonas
    S. M. Tibby
    I. A. Murdoch
    T. K. O'Brien
    N. W. F. Linton
    D. M. Band
    Intensive Care Medicine, 2000, 26 : 1507 - 1511
  • [23] Validation of cardiac output measurements utilizing arterial pulse contour analysis by Vigileo™ in cardiac surgery patients
    Kumaran, Rani
    Hwang, James
    Mansuroglu, Naim
    Oribabor, Charles
    Skubas, Nikolaos
    Osario, James
    Tortolani, Anthony
    Lee, Leonard
    Kurugundla, Navatha
    Devakonda, Arun
    Bernstein, Larry
    Sung, Arthur
    Raoof, Suhail
    CRITICAL CARE MEDICINE, 2007, 35 (12) : A77 - A77
  • [24] Cardiac output measured by lithium dilution and transpulmonary thermodilution in patients in a paediatric intensive care unit
    Linton, RA
    Jonas, MM
    Tibby, SM
    Murdoch, IA
    O'Brien, TK
    Linton, NWF
    Band, DM
    INTENSIVE CARE MEDICINE, 2000, 26 (10) : 1507 - 1511
  • [25] COMPARISON BETWEEN ULTRASONIC AND THERMODILUTION CARDIAC-OUTPUT MEASUREMENTS IN INTENSIVE-CARE PATIENTS
    VANDENBOGAERDE, JF
    SCHELDEWAERT, RG
    RIJCKAERT, DL
    CLEMENT, DL
    COLARDYN, FA
    CRITICAL CARE MEDICINE, 1986, 14 (04) : 294 - 297
  • [26] CORRELATION OF CARDIAC OUTPUT MEASURED BY TRANSPULMONARY THERMODILUTION AND PULSE CONTOUR ANALYSIS IN A EXPERIMENTAL ARDS MODEL.
    Diaz, F.
    Carvajal, C.
    Salomon, T.
    Torres, M.
    Erranz, B.
    Cruces, P.
    Donoso, A.
    CRITICAL CARE MEDICINE, 2008, 36 (12) : A56 - A56
  • [27] Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output
    Bein, B
    Worthmann, F
    Tonner, PH
    Paris, A
    Steinfath, M
    Hedderich, J
    Scholz, J
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2004, 18 (02) : 185 - 189
  • [28] Effects of changes in vascular tone on the agreement between pulse contour and transpulmonary thermodilution cardiac output measurements within an up to 6-hour calibration-free period
    Hamzaoui, Olfa
    Monnet, Xavier
    Richard, Christian
    Osman, David
    Chemla, Denis
    Teboul, Jean-Louis
    CRITICAL CARE MEDICINE, 2008, 36 (02) : 434 - 440
  • [29] Echocardiography and pulse contour analysis to assess cardiac output in trauma patients
    Franchi, F.
    Falciani, E.
    Donadello, K.
    Zaca, V.
    Silvestri, R.
    Taccone, F. S.
    Cubattoli, L.
    Mongelli, P.
    Giomarelli, R.
    Scolletta, S.
    MINERVA ANESTESIOLOGICA, 2013, 79 (02) : 137 - 146
  • [30] Measurement of cardiac output by pulse contour analysis (Vigileo®) in critically ill patients with prone positioning-A comparison with transpulmonary thermodilution
    Sakka, Samir G.
    Bruecken, Ulrike
    Gloeckner, Ulrike
    Knuettgen, Dirk
    Wappler, Frank
    SHOCK, 2008, 29 : 65 - 65