Performance comparison of high-speed photoacoustic microscopy: opto-ultrasound combiner versus ring-shaped ultrasound transducer

被引:0
|
作者
Hyojin Kim
Jin Young Kim
Seonghee Cho
Joongho Ahn
Yeonggeun Kim
Hyungham Kim
Chulhong Kim
机构
[1] Pohang University of Science and Technology (POSTECH),Departments of Electrical Engineering, Convergence IT Engineering, Mechanical Engineering, and Interdisciplinary Bioscience and Bioengineering, Medical Device Innovation Center
来源
关键词
Performance comparison; Photoacoustic microscopy; Opto-ultrasound combiner; Ring-shaped ultrasound transducer; Sensitivity; Signal-to-noise ratio;
D O I
暂无
中图分类号
学科分类号
摘要
Photoacoustic microscopy (PAM) embedded with a 532 nm pulse laser is widely used to visualize the microvascular structures in both small animals and humans in vivo. An opto-ultrasound combiner (OUC) is often utilized in high-speed PAM to confocally align the optical and acoustic beams to improve the system's sensitivity. However, acoustic impedance mismatch in the OUC results in little improvement in the sensitivity. Alternatively, a ring-shaped ultrasound transducer (RUT) can also accomplish the confocal configuration. Here, we compare the performance of OUC and RUT modules through ultrasound pulse-echo tests and PA imaging experiments. The signal-to-noise ratios (SNRs) of the RUT-based system were 15 dB, 12 dB, and 7 dB higher when compared to the OUC-based system for ultrasound pulse-echo test, PA phantom imaging test, and PA in-vivo imaging test, respectively. In addition, the RUT-based system could image the microvascular structures of small parts of a mouse body in a few seconds with minimal loss in SNR. Thus, with increased sensitivity, improved image details, and fast image acquisition, we believe the RUT-based systems could play a significant role in the design of future fast-PAM systems.
引用
收藏
页码:147 / 153
页数:6
相关论文
共 22 条
  • [1] Performance comparison of high-speed photoacoustic microscopy: opto-ultrasound combiner versus ring-shaped ultrasound transducer
    Kim, Hyojin
    Kim, Jin Young
    Cho, Seonghee
    Ahn, Joongho
    Kim, Yeonggeun
    Kim, Hyungham
    Kim, Chulhong
    [J]. BIOMEDICAL ENGINEERING LETTERS, 2022, 12 (02) : 147 - 153
  • [2] Development of photoacoustic imaging system of finger vasculature using ring-shaped ultrasound transducer
    Nishiyama, Misaki
    Namita, Takeshi
    Kondo, Kengo
    Yamakawa, Makoto
    Shiina, Tsuyoshi
    [J]. PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2018, 2018, 10494
  • [3] A High-Speed Photoacoustic Tomography System based on a Commercial Ultrasound and a Custom Transducer Array
    Wang, Xueding
    Cannata, Jonathan
    DeBusschere, Derek
    Hu, Changhong
    Fowlkes, J. Brian
    Carson, Paul
    [J]. PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2010, 2010, 7564
  • [4] High-speed wide-field photoacoustic microscopy using a cylindrically focused transparent high-frequency ultrasound transducer
    Chen, Maomao
    Jiang, Laiming
    Cook, Clare
    Zeng, Yushun
    Vu, Tri
    Chen, Ruimin
    Lu, Gengxi
    Yang, Wei
    Hoffmann, Ulrike
    Zhou, Qifa
    Yao, Junjie
    [J]. PHOTOACOUSTICS, 2022, 28
  • [5] Photoacoustic Tomography with a Ring Ultrasound Transducer: A Comparison of Different Illumination Strategies
    Alijabbari, Naser
    Alshahrani, Suhail S.
    Pattyn, Alexander
    Mehrmohammadi, Mohammad
    [J]. APPLIED SCIENCES-BASEL, 2019, 9 (15):
  • [6] High-Speed Integrated Endoscopic Photoacoustic and Ultrasound Imaging System
    Li, Yan
    Zhu, Zhikai
    Jing, Joseph C.
    Chen, Jason J.
    Heidari, Andrew Emon
    He, Youmin
    Zhu, Jiang
    Ma, Teng
    Yu, Mingyue
    Zhou, Qifa
    Chen, Zhongping
    [J]. IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2019, 25 (01) : 1 - 5
  • [7] Highly precise acoustic calibration method of ring-shaped ultrasound transducer array for plane-wave-based ultrasound tomography
    Terada, Takahide
    Yamanaka, Kazuhiro
    Suzuki, Atsuro
    Tsubota, Yushi
    Wu, Wenjing
    Kawabata, Ken-ichi
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2017, 56 (07)
  • [8] 3D Structure of Ring-shaped Microtubule Swarms Revealed by High-speed Atomic Force Microscopy
    Rashid, Mst. Rubaya
    Ganser, Christian
    Akter, Mousumi
    Nasrin, Syeda Rubaiya
    Kabir, Arif Md. Rashedul
    Sada, Kazuki
    Uchihashi, Takayuki
    Kakugo, Akira
    [J]. CHEMISTRY LETTERS, 2023, 52 (02) : 100 - 104
  • [9] Design and Simulation of a High-Frequency Ring-Shaped Linear Array for Capsule Ultrasound Endoscopy
    Lay, Holly S.
    Seetohul, Vipin
    Cox, Ben
    Demore, Christine E. M.
    Cochran, Sandy
    [J]. 2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 683 - 686
  • [10] Experimental Evaluation of a High-Speed Photoacoustic Tomography System based on a Commercial Ultrasound Unit
    Wang, X.
    Fowkes, J. B.
    Carson, P. L.
    Mo, L.