Choosing the amplitude interval for a proportional counter at a low signal level

被引:0
|
作者
P. S. Antsiferov
机构
[1] Russian Academy of Sciences,Institute of Spectroscopy
来源
Technical Physics | 1998年 / 43卷
关键词
Iron; Distribution Function; Background Level; Pulse Amplitude; Amplitude Distribution;
D O I
暂无
中图分类号
学科分类号
摘要
A procedure is described for choosing the optimum pulse amplitude interval for a proportional counter when the useful signal level is comparable with the natural background level. It is shown that the optimum interval for counter pulses having a Gaussian amplitude distribution function corresponds to 1.4σ. The application of the procedure is demonstrated in a test example with processing of the iron Kα x-ray spectrum.
引用
收藏
页码:596 / 597
页数:1
相关论文
共 50 条
  • [1] Choosing the amplitude interval for a proportional counter at a low signal level
    Antsiferov, PS
    TECHNICAL PHYSICS, 1998, 43 (05) : 596 - 597
  • [2] LOW LEVEL PROPORTIONAL COUNTER FOR TRITIUM
    CHARALAMBUS, S
    GOEBEL, K
    NUCLEAR INSTRUMENTS & METHODS, 1963, 25 (01): : 109 - 117
  • [3] A PROPORTIONAL COUNTER FOR LOW LEVEL COUNTING WITH HIGH EFFICIENCY
    OLSSON, I
    DEVRIES, H
    ARKIV FOR FYSIK, 1960, 16 (06): : 516 - 516
  • [4] A PROPORTIONAL COUNTER FOR LOW LEVEL COUNTING WITH HIGH EFFICIENCY
    DEVRIES, H
    STUIVER, M
    OLSSON, I
    NUCLEAR INSTRUMENTS & METHODS, 1959, 5 (02): : 111 - 114
  • [5] PROPORTIONAL COUNTER FOR LOW-LEVEL PU-141 MEASUREMENT
    WINKLER, R
    ROSNER, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 274 (1-2): : 359 - 361
  • [6] AN INTERNAL GAS PROPORTIONAL COUNTER FOR MEASURING LOW LEVEL ENVIRONMENTAL RADIONUCLIDES
    SCHELL, WR
    ADVANCES IN CHEMISTRY SERIES, 1970, (93): : 173 - &
  • [7] A PROPORTIONAL COUNTER FOR LOW-LEVEL MEASUREMENT OF PLUTONIUM-239 IN LUNGS
    TAYLOR, BT
    HEALTH PHYSICS, 1969, 17 (01): : 59 - +
  • [8] Proportional counter amplification at low voltages
    Rose, ME
    Ramsey, WE
    PHYSICAL REVIEW, 1942, 61 (7/8): : 504 - 509
  • [9] Controlling low-rate signal path microdischarge for an ultra-low-background proportional counter
    E. K. Mace
    C. E. Aalseth
    R. M. Bonicalzi
    A. R. Day
    E. W. Hoppe
    M. E. Keillor
    A. W. Myers
    C. T. Overman
    A. Seifert
    Journal of Radioanalytical and Nuclear Chemistry, 2013, 296 : 753 - 758
  • [10] Controlling low-rate signal path microdischarge for an ultra-low-background proportional counter
    Mace, E. K.
    Aalseth, C. E.
    Bonicalzi, R. M.
    Day, A. R.
    Hoppe, E. W.
    Keillor, M. E.
    Myers, A. W.
    Overman, C. T.
    Seifert, A.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2013, 296 (02) : 753 - 758