Unicortical PEEK inset locking fixation for metacarpal fractures: a biomechanical study

被引:1
|
作者
Mudrick C.A. [1 ]
Owen J.R. [2 ]
Wayne J.S. [2 ]
Isaacs J.E. [1 ]
机构
[1] Orthopaedic Research Laboratory, Departments of Orthopaedic Surgery and Biomedical Engineering, Virginia Commonwealth University, P.O. Box 980153, Richmond, 23298-0153, VA
[2] Orthopaedic Research Laboratory, Departments of Orthopaedic Surgery and Biomedical Engineering, Virginia Commonwealth University, P.O. Box 980694, Richmond, 23298-0153, VA
关键词
Bending load; Failure; Internal fixation; polyetheretherketone (PEEK); Porcine metacarpal;
D O I
10.1007/s00590-013-1322-y
中图分类号
学科分类号
摘要
Purpose: There are numerous constructs employed in the treatment of metacarpal fractures with varying degrees of success. While plate fixation commonly involves dorsal application of a bicortical non-locking plate, there has been recent exploration of other fixation options including unicortical locked plating. The purpose of this study was to evaluate the biomechanical integrity of a polyetheretherketone (PEEK) inset locking plate and, in doing so, compare it to standard plate fixation (utilizing a clinically proven bicortical non-locking titanium plate) in a simulated porcine metacarpal fracture model.; Methods: Reproducible mid-shaft fractures were created in porcine second metacarpals. The fractured specimens were reduced and plated with either a bicortical non-locking plate or a unicortical locking plate with a PEEK locking design. Constructs were then loaded to failure in the same fashion as performed to create the fracture. Peak load was measured as the apex on the load-to-failure deflection curve. Stiffness was calculated as the linear slope on the load-to-failure deflection curve. Data were analyzed via Student’s t test.; Results: Unicortical locking constructs failed at 344 ± 119 N, while bicortical non-locking constructs were found to fail at 277 ± 101 N (p = 0.19). The unicortical locking constructs demonstrated a stiffness of 80 ± 36 N/mm compared with the bicortical non-locking constructs (69 ± 36 N/mm) although again the difference was not found to be statistically different (p = 0.49).; Conclusion: Based on this study, a locked plating construct using a polymer mechanism provides an interesting new locking fixation method for small bone fractures and with our limited number of specimens tested, provided at least a similar strength and rigidity profile in comparison with bicortical fixation in the treatment of metacarpal fractures. © 2013, Springer-Verlag France.
引用
收藏
页码:1415 / 1420
页数:5
相关论文
共 50 条
  • [1] Unicortical fixation of metacarpal fractures: is it strong enough?
    Dickson, J. K.
    Bhat, W.
    Gujral, S.
    Paget, J.
    O'Neill, J.
    Lee, S. J.
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2016, 41 (04) : 367 - 372
  • [2] Biomechanical evaluation of locking plate radial shaft fixation: Unicortical locking fixation versus mixed bicortical and unicortical fixation in a sawbone model
    Roberts, Jason W.
    Grindel, Steven I.
    Rebholz, Brandon
    Wang, Mei
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2007, 32A (07): : 971 - 975
  • [3] Biomechanical study of the fixation ability of the dorsal and volar locking plate for transverse metacarpal neck fractures
    Ho, Tsung-Yu
    Hsu, Cheng-En
    Wang, Wei-Chih
    Ting, Yen-Nien
    Tsai, Ming-Tzu
    Chiu, Yung-Cheng
    Hsu, Jui-Ting
    MEDICINE, 2023, 102 (36) : E34981
  • [4] Bicortical versus unicortical fixation of plated clavicular fractures: A biomechanical study
    Zaidenberg, Ezequiel E.
    Voor, Michael
    Pereira, Enrique
    Rossi, Luciano A.
    Zaidenberg, Carlos R.
    SHOULDER & ELBOW, 2021, 13 (04) : 426 - 432
  • [5] Unicortical and bicortical plating in the fixation of comminuted fractures of the clavicle: a biomechanical study
    Looft, John M.
    Correa, Lincoln
    Patel, Minoo
    Rawlings, Mathew
    Ackland, David C.
    ANZ JOURNAL OF SURGERY, 2017, 87 (11) : 915 - 920
  • [6] Biomechanical analysis of locking plates for fixation of metacarpal shaft fractures: A finite element analysis
    Zhang, Jian-Feng
    Li, Yong
    Huo, Yong-Xin
    Yan, Ming
    Liang, Shou-Lei
    Wang, Lei
    Wang, Bao-Cang
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2022, 108 (08)
  • [7] A Biomechanical Comparison of Locked Unicortical with Non-Locking Bicortical Screws for Metacarpal Shaft Fractures in a Randomized Human Cadaveric Model
    Minogue, Febe
    Reilly, Keryn
    Walker, Cameron
    Malak, Sharif
    Mayo, Andrew
    Heiss-Dunlop, Wolfgang
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2021, 26 (01): : 10 - 16
  • [8] Biomechanical study on fixation methods for horizontal oblique metacarpal shaft fractures
    Yung-Cheng Chiu
    Cheng-En Hsu
    Tsung-Yu Ho
    Yen-Nien Ting
    Ming-Tzu Tsai
    Jui-Ting Hsu
    Journal of Orthopaedic Surgery and Research, 17
  • [9] Biomechanical study on fixation methods for horizontal oblique metacarpal shaft fractures
    Chiu, Yung-Cheng
    Hsu, Cheng-En
    Ho, Tsung-Yu
    Ting, Yen-Nien
    Tsai, Ming-Tzu
    Hsu, Jui-Ting
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [10] A Biomechanical Comparison of Fixation Techniques in Metacarpal Shaft Fractures
    Albanese, Kevin M.
    Schreck, Michael J.
    Werner, Frederick W.
    Esper, Garrett W.
    Ordway, Nathaniel R.
    JOURNAL OF WRIST SURGERY, 2023, 12 (01) : 46 - 51