Technical and clinical validation of three commercial real-time PCR kits for the diagnosis of neuroborreliosis in cerebrospinal fluid on three different real-time PCR platforms

被引:0
|
作者
L. Maes
T. Carolus
V. De Preter
S. Ignoul
R. Cartuyvels
L. Braeken
P.-J. D’Huys
V. Saegeman
B. Kabamba
M. Raymaekers
机构
[1] Research group Lab4U,Translational Research Center for Gastrointestinal Disorders (TARGID)
[2] Group of Management and Technology,Clinical Biology, Laboratory for Molecular Diagnostics, Campus Salvator
[3] UC Leuven-Limburg,Research group Lab4U, Faculty of Industrial Engineering
[4] University Hospital Gasthuisberg,Department of Laboratory Medicine
[5] KULeuven,Department of Microbiology and Immunology
[6] Jessa Hospital,Clinical Microbiology, Laboratory of Microbiology
[7] KU Leuven - Campus Diepenbeek,undefined
[8] University Hospitals Leuven,undefined
[9] KU Leuven,undefined
[10] University Clinic St-Luc UCL,undefined
关键词
Sensu Stricto; Borrelia Burgdorferi; Lyme Borreliosis; Lyme Neuroborreliosis; ospA Gene;
D O I
暂无
中图分类号
学科分类号
摘要
This study reports the evaluation of the technical and clinical validation of the O-DiaBorburg kit (DIA), Borrelia burgdorferi PCR kit, ISEX (GENE), and Borrelia burgdorferi sensu lato Real-TM (SAC) for the diagnosis of neuroborreliosis in cerebrospinal fluid based on both Borrelia DNA and CSF samples from patients with clinical suspicion of neuroborreliosis. This validation study was done by analysing the kits on the Rotorgene Q (RGQ), CFX96, and LightCycler480 (LC480). For all kits, the linear range was larger on RGQ than on CFX96 and LC480. A good reproducibility was obtained for all assays on all instruments. Storage at −20 °C resulted in a decreased reproducibility for SAC. Results of the limit of detection (LOD95) experiments indicated a better sensitivity than described in the kit insert for all kits on all PCR platforms. No cross-reactivity was found for genetically related organisms nor for other pathogens which may be present in CSF. All species of the Borrelia burgdorferi sensu lato complex were detected with the GENE and SAC kits. The DIA kit failed to detect B. lusitaniae. The results seemed to indicate a better overall performance for the GENE kit on RGQ. However, its diagnostic value could not be confirmed in the clinical validation study, wherein none of the 103 CSF samples from clinical neuroborreliosis cases showed a positive real-time PCR result with the GENE kit analysed on RGQ.
引用
收藏
页码:273 / 279
页数:6
相关论文
共 50 条
  • [1] Technical and clinical validation of three commercial real-time PCR kits for the diagnosis of neuroborreliosis in cerebrospinal fluid on three different real-time PCR platforms
    Maes, L.
    Carolus, T.
    De Preter, V.
    Ignoul, S.
    Cartuyvels, R.
    Braeken, L.
    D'Huys, P. -J.
    Saegeman, V.
    Kabamba, B.
    Raymaekers, M.
    [J]. EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2017, 36 (02) : 273 - 279
  • [2] Evaluation of three commercial real-time PCR kits for dengue diagnosis
    Najioullah, F.
    Angla-Gre, M.
    Viron, F.
    Cesaire, R.
    [J]. INTERNATIONAL JOURNAL OF INFECTIOUS DISEASES, 2014, 21 : 373 - 373
  • [3] Agreement between two real-time commercial PCR kits and an in-house real-time PCR for diagnosis of mucormycosis
    Rafanomezantsoa, Lovanirina Clemencia
    Sabourin, Estelle
    Guennouni Sebbouh, Nadia
    Sitterle, Emilie
    Ben Halima, Nada
    Raveloarisaona, Yannick Sonjah
    Quesne, Gilles
    Dannaoui, Eric
    Bougnoux, Marie-Elisabeth
    [J]. MICROBIOLOGY SPECTRUM, 2024, 12 (08):
  • [4] Three years experience of real-time PCR for the diagnosis of Q fever
    Ughetto, E.
    Gouriet, F.
    Raoult, D.
    Rolain, J. -M.
    [J]. CLINICAL MICROBIOLOGY AND INFECTION, 2009, 15 : 200 - 201
  • [5] Assessment of the real-time PCR and different digital PCR platforms for DNA quantification
    Pavsic, Jernej
    Zel, Jana
    Milavec, Mojca
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2016, 408 (01) : 107 - 121
  • [6] Assessment of the real-time PCR and different digital PCR platforms for DNA quantification
    Jernej Pavšič
    Jana Žel
    Mojca Milavec
    [J]. Analytical and Bioanalytical Chemistry, 2016, 408 : 107 - 121
  • [7] Comparison of Diagnostic Performance of Three Real-Time PCR Kits for Detecting Mycobacterium Species
    Cho, Sun Young
    Kim, Min Jin
    Suh, Jin-Tae
    Lee, Hee Joo
    [J]. YONSEI MEDICAL JOURNAL, 2011, 52 (02) : 301 - 306
  • [8] Comparison and Sensitivity Evaluation of Three Different Commercial Real-Time Quantitative PCR Kits for SARS-CoV-2 Detection
    Banko, Ana
    Petrovic, Gordana
    Miljanovic, Danijela
    Loncar, Ana
    Vukcevic, Marija
    Despot, Dragana
    Cirkovic, Andja
    [J]. VIRUSES-BASEL, 2021, 13 (07):
  • [9] The evolution of real-time PCR machines to real-time PCR chips
    Lee, Dasheng
    Chen, Pei-Jer
    Lee, Gwo-Bin
    [J]. BIOSENSORS & BIOELECTRONICS, 2010, 25 (07): : 1820 - 1824
  • [10] Clinical experience with real-time PCR in the diagnosis of malaria
    Sloma, Cari R.
    Rose, Heather
    Sloane, Lynne
    Rosenblatt, John E.
    [J]. AMERICAN JOURNAL OF TROPICAL MEDICINE AND HYGIENE, 2006, 75 (05): : 52 - 52