Model predictive control of relative blood volume and heart rate during hemodialysis

被引:0
|
作者
Faizan Javed
Andrey V. Savkin
Gregory S. H. Chan
Paul M. Middleton
Philip Malouf
Elizabeth Steel
James Mackie
Teddy M. Cheng
机构
[1] The University of New South Wales,School of Electrical Engineering and Telecommunications
[2] Prince of Wales Hospital Sydney,undefined
关键词
Model predictive control; Hemodialysis; System identification;
D O I
暂无
中图分类号
学科分类号
摘要
To maintain the hemodynamic stability of patient undergoing hemodialysis, this article proposes a novel model-based control methodology to regulate the changes in relative blood volume (RBV) and percentage change in heart rate (∆HR(%)) during hemodialysis by adjusting the ultrafiltration rate (UFR). The control algorithm uses model predictive control (MPC) to account for system variability and to explicitly handle the constraints on UFR. Linear state-space system with time-varying parameters is introduced to model the RBV and ∆HR. MPC was used to track the change in RBV and ∆HR to pre-defined reference trajectories. At each sampling instant, the system parameters are updated to get the best fitting into the parameterized model. Simulation results demonstrate that the system is able to regulate RBV and ∆HR to the reference by adjusting UFR while keeping it within practically realizable bounds. The results show that adjusting UFR may improve the stability of patient during dialysis when compared to conventional hemodialysis with constant UFR.
引用
收藏
页码:389 / 397
页数:8
相关论文
共 50 条
  • [1] Model predictive control of relative blood volume and heart rate during hemodialysis
    Javed, Faizan
    Savkin, Andrey V.
    Chan, Gregory S. H.
    Middleton, Paul M.
    Malouf, Philip
    Steel, Elizabeth
    Mackie, James
    Cheng, Teddy M.
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2010, 48 (04) : 389 - 397
  • [2] Modeling and control of the heart rate and blood volume responses to hemodialysis
    Javed, Faizan
    Savkin, Andrey V.
    Chan, Gregory S. H.
    Middleton, Paul M.
    Malouf, Philip
    Steel, Elizabeth
    Mackie, James
    Cheng, Teddy M.
    [J]. 2009 IEEE INTERNATIONAL CONFERENCE ON CONTROL AND AUTOMATION, VOLS 1-3, 2009, : 625 - +
  • [3] Relative blood volume changes underestimate total blood volume changes during hemodialysis
    Dasselaar, Judith J.
    Lub-de Hooge, Marjolijn N.
    Pruim, Jan
    Nijnuis, Hugo
    Wiersum, Anneke
    de Jong, Paul E.
    Huisman, Roel M.
    Franssen, Casper F. M.
    [J]. CLINICAL JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2007, 2 (04): : 669 - 674
  • [4] NOVEL ULTRAFILTRATION RATE FEEDBACK CONTROLLER FOR ATTAINMENT OF RELATIVE BLOOD VOLUME TARGETS DURING HEMODIALYSIS
    Rivera-Fuentes, Lemuel
    Tapia, Leticia
    Thijssenl, Stephan
    Rogg, Sabrina
    Kotanko, Peter
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2020, 35 : 1406 - 1406
  • [5] A NEW ULTRAFILTRATION RATE FEEDBACK CONTROLLER FOR ATTAINMENT OF RELATIVE BLOOD VOLUME TARGETS DURING HEMODIALYSIS
    Wang, Lin-Chun
    Thijssen, Stephan
    Rivera, Lemuel
    Silva, Leticia M. Tapia
    Ye, Xiaoling
    Casper, Sabrina
    Fuertinger, Doris
    Fuertinger, Stefan
    Kotanko, Peter
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2023, 38 : I147 - I148
  • [6] Relative blood volume changes underestimate absolute blood volume changes during hemodialysis
    Dasselaar, Judith
    Hooge, Marjolijn Lub-de
    Pruim, Jan
    Nijnuis, Hugo
    Wiersum, Anneke
    de Jong, Paul
    Huisman, Roel
    Franssen, Casper
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2007, 22 : 215 - 216
  • [7] Scattering effect for relative blood volume monitoring during hemodialysis
    Zhou, Lianqiao
    Li, Qinlan
    Wei, Shuang
    Wang, Shugang
    Cui, Jingqiang
    Wang, Guosheng
    Liu, Chong
    Su, Yewang
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2024, 377
  • [8] Effect of high and low ultrafiltration volume during hemodialysis on relative blood volume
    Dasselaar, JJ
    de Jong, PE
    Huisman, RM
    Franssen, CFM
    [J]. ASAIO JOURNAL, 2006, 52 (02) : 169 - 173
  • [9] Model-Free Predictive Control of Human Heart Rate and Blood Pressure
    Sarabadani, A.
    Bernasconi, S.
    Klamroth-Marganska, V.
    Nussbaumer, S.
    Riener, R.
    [J]. 2013 IEEE 13TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING (BIBE), 2013,
  • [10] Effects of relative blood volume and ultrafiltration volume on pulse wave characteristics during hemodialysis
    Krenn, Simon
    Hecking, Manfred
    Mussnig, Sebastian
    Wassertheurer, Siegfried
    Mayer, Christopher
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2024, 39 : I2754 - I2756