Construction and performance test of prototype parallel plate avalanche counters at RAON

被引:0
|
作者
Charles Akers
Kwang Bok Lee
Young Jin Kim
Eun Hee Kim
Young Kwan Kwon
Hyo Sang Lee
Jun Young Moon
Jin Hyung Park
Min Sang Ryu
Taeksu Shin
Satou Yoshiteru
机构
[1] Institute for Basic Science,Rare Isotope Science Project
来源
Journal of the Korean Physical Society | 2017年 / 70卷
关键词
PPAC; RISP; RAON; Position sensitive; delay-line readout;
D O I
暂无
中图分类号
学科分类号
摘要
Two prototype position sensitive parallel plate avalanche counters (PPACs) have been fabricated at the Rare Isotope Science Project (RISP). Both detectors have active areas of 10 × 10 cm2 and utilize a stripped electrode structure, together with a delay line readout, to extract position information. Testing was conducted with an 241Am α source, and 2-dimensional position information was obtained. A position resolution of 0.55 mm full width at half maximum (FWHM), together with a position linearity of ±55 μm over a 67-mm detection range, were measured. A time resolution of 450 ps (FWHM) was also measured.
引用
收藏
页码:682 / 686
页数:4
相关论文
共 50 条
  • [1] Construction and performance test of prototype parallel plate avalanche counters at RAON
    Akers, Charles
    Lee, Kwang Bok
    Kim, Young Jin
    Kim, Eun Hee
    Kwon, Young Kwan
    Lee, Hyo Sang
    Moon, Jun Young
    Park, Jin Hyung
    Ryu, Min Sang
    Shin, Taeksu
    Yoshiteru, Satou
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 70 (07) : 682 - 686
  • [2] Parallel plate avalanche counters
    Martinez-Davalos, A
    Alfaro-Molina, R
    INSTRUMENTATION IN ELEMENTARY PARTICLE PHYSICS: THE VII ICFA SCHOOL, 1998, (422): : 324 - 332
  • [3] Fabrication and performance evaluation of parallel plate avalanche counters for RISP
    Lee, DongYeup
    Akers, Charles
    Kwag, Minsik
    Lee, Kwang Bok
    Park, HyangKyu
    Tshoo, Kyoungho
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 1014
  • [4] DEVELOPMENT OF PARALLEL-PLATE AVALANCHE COUNTERS FOR DETECTION OF FISSION FRAGMENTS
    HEMPEL, G
    HOPKINS, F
    SCHATZ, G
    NUCLEAR INSTRUMENTS & METHODS, 1975, 131 (03): : 445 - 450
  • [5] DEVELOPMENT OF PARALLEL-PLATE AVALANCHE COUNTERS FOR DETECTION OF FISSION FRAGMENTS
    SCHATZ, G
    HEMPEL, G
    HOPKINS, F
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1975, 20 (01): : 40 - 40
  • [6] CONSTRUCTION AND OPERATION OF THIN PARALLEL PLATE PROPORTIONAL COUNTERS
    HILBERG, RH
    DRAPER, JE
    ALSTON, WJ
    NUCLEAR INSTRUMENTS & METHODS, 1967, 50 (01): : 121 - +
  • [7] BRAGG CURVE SPECTROSCOPY OF FISSION FRAGMENTS BY USING PARALLEL PLATE AVALANCHE COUNTERS
    NEUBERT, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 237 (03): : 535 - 542
  • [8] APPLICATION OF PARALLEL PLATE AVALANCHE COUNTERS FOR PROTON-INDUCED FISSION STUDIES
    KOTOV, AA
    NEUBERT, W
    ANDRONENKO, LN
    GORSHKOV, BL
    KOVSHEVNY, GG
    VAISHNENE, LA
    YAZIKOV, MI
    NUCLEAR INSTRUMENTS & METHODS, 1980, 178 (01): : 55 - 60
  • [9] SIMPLE AND ACCURATE METHOD FOR BIDIMENSIONAL POSITION READOUT OF PARALLEL PLATE AVALANCHE COUNTERS
    BRESKIN, A
    ZWANG, N
    NUCLEAR INSTRUMENTS & METHODS, 1977, 146 (02): : 461 - 463
  • [10] A DETECTOR SYSTEM FOR ANGULAR-CORRELATION MEASUREMENTS BY USING PARALLEL PLATE AVALANCHE COUNTERS
    NEUBERT, W
    KOTOV, AA
    ANDRONENKO, LN
    ILJIN, AI
    KOVSHEVNY, GG
    VAISHNENE, LA
    VOLKOV, SS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 204 (2-3): : 453 - 461