Pivot wear of a centrifugal blood pump developed for circulatory assist

被引:0
|
作者
Takashi Yamane
Katsunobu Nonaka
Hisashi Miyoshi
Osamu Maruyama
Masahiro Nishida
Ryo Kosaka
Yoshiyuki Sankai
Tatsuo Tsutsui
机构
[1] National Institute of Advanced Industrial Science and Technology,School of Engineering
[2] University of Tsukuba,School of Clinical Medicine
[3] University of Tsukuba,undefined
来源
Journal of Artificial Organs | 2008年 / 11卷
关键词
Artificial heart; Centrifugal pump; Pivot bearing; Durability; Hemocompatibility;
D O I
暂无
中图分类号
学科分类号
摘要
We are developing a monopivot centrifugal pump for circulatory assist for a period of more than 2 weeks. The impeller is supported by a pivot bearing at one end and by a passive magnetic bearing at the other. The pivot undergoes concentrated exposure to the phenomena of wear, hemolysis, and thrombus formation. The pivot durability, especially regarding the combination of male/female pivot radii, was examined through rotating wear tests and animal tests. As a result, combinations of similar radii for the male/female pivots were found to provide improved pump durability. In the extreme case, the no-gap combination would result in no thrombus formation.
引用
收藏
页码:232 / 237
页数:5
相关论文
共 50 条
  • [31] Centrifugal Blood Pump for Temporary Ventricular Assist Devices With Low Priming and Ceramic Bearings
    Leme, Juliana
    da Silva, Cibele
    Fonseca, Jeison
    da Silva, Bruno Utiyama
    Uebelhart, Beatriz
    Biscegli, Jose F.
    Andrade, Aron
    ARTIFICIAL ORGANS, 2013, 37 (11) : 942 - 945
  • [32] Development of the Baylor Gyro permanently implantable centrifugal blood pump as a biventricular assist device
    Nonaka, K
    Linneweber, J
    Ichikawa, S
    Yoshikawa, T
    Kawahito, S
    Mikami, M
    Motomura, T
    Ishitoya, H
    Nishimura, I
    Oestmann, D
    Glueck, J
    Schima, H
    Wolner, E
    Shinohara, T
    Nosé, Y
    ARTIFICIAL ORGANS, 2001, 25 (09) : 675 - 682
  • [33] CIRCULATORY SUPPORT WITH A CENTRIFUGAL PUMP AS A BRIDGE TO CARDIAC TRANSPLANTATION
    BOLMAN, RM
    COX, JL
    MARSHALL, W
    KOUCHOUKOS, N
    SPRAY, TL
    CANCE, C
    GENTON, RE
    SAFFITZ, J
    ANNALS OF THORACIC SURGERY, 1989, 47 (01): : 108 - 112
  • [34] A STRAIGHT PATH CENTRIFUGAL BLOOD PUMP CONCEPT IN THE CAPIOX CENTRIFUGAL PUMP
    KIJIMA, T
    OSHIYAMA, H
    HORIUCHI, K
    NOGAWA, A
    HAMASAKI, H
    AMANO, N
    NOJIRI, C
    FUKASAWA, H
    AKUTSU, T
    ARTIFICIAL ORGANS, 1993, 17 (07) : 593 - 598
  • [35] Centrifugal blood pump scaling
    Lorenz, M.
    Smith, W.
    ASAIO Journal, 2000, 46 (02)
  • [37] Wear analysis of centrifugal slurry pump impellers
    Khalid, Y. A.
    Sapuan, S. M.
    INDUSTRIAL LUBRICATION AND TRIBOLOGY, 2007, 59 (01) : 18 - 28
  • [38] CENTRIFUGAL WATERFLOOD PUMP WEAR RING METALLURGY
    BALDWIN, CJ
    SABATHIE.CD
    MECHANICAL ENGINEERING, 1974, 96 (10) : 61 - 61
  • [39] Interaction of an Idealized Cavopulmonary Circulation With Mechanical Circulatory Assist Using an Intravascular Rotary Blood Pump
    Bhavsar, Sonya S.
    Moskowitz, William B.
    Throckmorton, Amy L.
    ARTIFICIAL ORGANS, 2010, 34 (10) : 816 - 827
  • [40] Evaluation of hemolysis in a pulsatile assist device for centrifugal pump
    Department of Cardiovascular Surgery, Fukushima Medical College, Senko Medical Inc., Fukushima, Japan
    不详
    ARTIF. ORGANS, 7 (700-703):