Impact of recording length and other arrhythmias on atrial fibrillation detection from wrist photoplethysmogram using smartwatches

被引:0
|
作者
Min-Tsun Liao
Chih-Chieh Yu
Lian-Yu Lin
Ke-Han Pan
Tsung-Hsien Tsai
Yu-Chun Wu
Yen-Bin Liu
机构
[1] National Taiwan University Hospital Hsinchu Branch,Division of Cardiology, Department of Internal Medicine
[2] National Taiwan University,Department of Medicine, College of Medicine
[3] National Taiwan University Hospital,Division of Cardiology, Department of Internal Medicine
[4] Acer Inc.,Value Lab
来源
Scientific Reports | / 12卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
This study aimed to evaluate whether quantitative analysis of wrist photoplethysmography (PPG) could detect atrial fibrillation (AF). Continuous electrocardiograms recorded using an electrophysiology recording system and PPG obtained using a wrist-worn smartwatch were simultaneously collected from patients undergoing catheter ablation or electrical cardioversion. PPG features were extracted from 10, 25, 40, and 80 heartbeats of the split segments. Machine learning with a support vector machine and random forest approach were used to detect AF. A total of 116 patients were evaluated. We annotated > 117 h of PPG. A total of 6475 and 3957 segments of 25-beat pulse-to-pulse intervals (PPIs) were annotated as AF and sinus rhythm, respectively. The accuracy of the 25 PPIs yielded a test area under the receiver operating characteristic curve (AUC) of 0.9676, which was significantly better than the AUC for the 10 PPIs (0.9453; P < .001). PPGs obtained from another 38 patients with frequent premature ventricular/atrial complexes (PVCs/PACs) were used to evaluate the impact of other arrhythmias on diagnostic accuracy. The new AF detection algorithm achieved an AUC of 0.9680. The appropriate data length of PPG for optimizing the PPG analytics program was 25 heartbeats. Algorithm modification using a machine learning approach shows robustness to PVCs/PACs.
引用
收藏
相关论文
共 50 条
  • [21] Atrial Fibrillation Detection Using a Novel Cardiac Ambulatory Monitor Based on Photo-Plethysmography at the Wrist
    Bonomi, Alberto G.
    Schipper, Fons
    Eerikainen, Linda M.
    Margarito, Jenny
    van Dinther, Ralph
    Muesch, Guido
    de Morree, Helma M.
    Aarts, Ronald M.
    Babaeizadeh, Saeed
    McManus, David D.
    Dekker, Lukas R. C.
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2018, 7 (15):
  • [22] DIAGNOSTIC ACCURACY OF SMARTWATCHES AND WEARABLE DEVICES USING PHOTOPLETHYSMOGRAPHY AND ELECTROCARDIOGRAPHY FOR ATRIAL FIBRILLATION DETECTION: A SYSTEMATIC REVIEW AND META-ANALYSIS
    Alkatiri, Zuhair
    Nurisma, Najiba Gina
    Nurisma, Nabila
    Ramadhan, Muhammad Bimo
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2025, 85 (12) : 2842 - 2842
  • [23] Pseudo-V1 ECG lead recording with quality smartwatches and medical interpretation may be necessary for timely and accurate remote detection of atrial flutter/fibrillation
    Brisinda, D.
    Fenici, P.
    Fenici, R.
    EUROPEAN HEART JOURNAL, 2022, 43 : 413 - 413
  • [24] Automatic Detection of Atrial Fibrillation and Other Arrhythmias in ECG Recordings Acquired by a Smartphone Device (vol 7, 199, 2018)
    Billeci, Lucia
    Costi, Magda
    Lombardi, David
    Chiarugi, Franco
    Varanini, Maurizio
    ELECTRONICS, 2018, 7 (10)
  • [25] The Accuracy of Atrial Fibrillation Detection from Wrist Photoplethysmography. A Study on Post-Operative Patients
    Tarniceriu, Adrian
    Harju, Jarkko
    Yousefi, Zeinab Rezaei
    Vehkaoja, Antti
    Parak, Jakub
    Yli-Hankala, Arvi
    Korhonen, Ilkka
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 4844 - 4847
  • [26] Detection of atrial fibrillation and other abnormal rhythms from ECG using a multi-layer classifier architecture
    Mukherjee, Ayan
    Choudhury, Anirban Dutta
    Datta, Shreyasi
    Puri, Chetanya
    Banerjee, Rohan
    Singh, Rituraj
    Ukil, Arijit
    Bandyopadhyay, Soma
    Pal, Arpan
    Khandelwal, Sundeep
    PHYSIOLOGICAL MEASUREMENT, 2019, 40 (05)
  • [27] Arrhythmias Other Than Atrial Fibrillation in Those With an Irregular Pulse Detected With a Smartwatch Findings From the Apple Heart Study
    Perino, Alexander C.
    Gummidipundi, Santosh E.
    Lee, Justin
    Hedlin, Haley
    Garcia, Ariadna
    Ferris, Todd
    Balasubramanian, Vidhya
    Gardner, Rebecca M.
    Cheung, Lauren
    Hung, Grace
    Granger, Christopher B.
    Kowey, Peter
    Rumsfeld, John S.
    Russo, Andrea M.
    True Hills, Mellanie
    Talati, Nisha
    Nag, Divya
    Tsay, David
    Desai, Sumbul
    Desai, Manisha
    Mahaffey, Kenneth W.
    Turakhia, Mintu P.
    Perez, Marco V.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2021, 14 (10): : E010063
  • [28] Atrial Fibrillation Classification from a Short Single Lead ECG Recording Using Hierarchical Classifier
    Coppola, Erin E.
    Gyawali, Prashnna K.
    Vanjara, Nihar
    Giaime, Daniel
    Wang, Linwei
    2017 COMPUTING IN CARDIOLOGY (CINC), 2017, 44
  • [29] Computing the cycle length of atrial fibrillation from the ECG using spectral and spatial analyses
    Hoppe, BL
    Bhargava, V
    Hassankhani, A
    Feld, GK
    Narayan, SM
    CIRCULATION, 2004, 110 (17) : 578 - 578
  • [30] Impact of long-term ECG recording on the detection of paroxysmal atrial fibrillation in patients after an acute ischemic stroke
    Schuchert, A
    Behrens, G
    Meinertz, T
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (07): : 1082 - 1084