Usefulness of an automated urinary flow cytometer in mass screening for nephritis

被引:0
|
作者
Kazunari Kaneko
Mutsumi Murakami
Kazumi Shiraishi
Masaru Matsumoto
Kuniaki Yamauchi
Teruo Kitagawa
Yuichiro Yamashiro
机构
[1] Juntendo University Urayasu Hospital,Department of Pediatrics
[2] Nippon Medical School Hospital,Department of Pediatrics
[3] Tokyo Health Service Association,Department of Pediatrics
[4] Juntendo University School of Medicine,undefined
来源
Pediatric Nephrology | 2004年 / 19卷
关键词
Urinary screening; Automated urinary flow cytometer; Glomerular hematuria;
D O I
暂无
中图分类号
学科分类号
摘要
We aimed to assess the utility of a newly developed automated urinary flow cytometer (UFCM) in differentiating the origin of hematuria in the mass screening system for renal diseases in school children. In total, 4,620 children aged 6–14 years with abnormal urinary findings by the screening program in Tokyo were enrolled. Tentative diagnoses were made based on the criteria for this screening program: nephritis in 11, suspected nephritis in 104, hematuria in 771, minimal hematuria in 1,506, proteinuria in 477, urinary tract infection in 83, and healthy in 1,668. Glomerular hematuria, assessed by UFCM, was found in 81.8% of nephritis, 58.7% of suspected nephritis, 59.7% of hematuria, 57.4% of minimal hematuria, 13.4% of proteinuria, and 21.5% of the healthy group. The presence of glomerular hematuria assessed by UFCM had a sensitivity of 61.0%, specificity of 78.5%, positive predictive value of 16.3%, and negative predictive value of 96.7% for the diagnosis of nephritis and suspected nephritis. Thus, our results imply that the absence of glomerular hematuria as assessed by UFCM is highly predictive of the absence of nephritis. If “healthy” children with non-glomerular hematuria by UFCM could be excluded, the total annual cost saving would be approximately 147.4 million JPY, which is equivalent to 5.2% of the current system. In conclusion, a simple, automated and rapid test for glomerular hematuria using UFCM is useful for mass screening. It can precisely predict children with only transient hematuria who do not have any serious illness, and therefore can exclude further unnecessary examinations, resulting in the improvement of cost-effectiveness of the current program.
引用
收藏
页码:499 / 502
页数:3
相关论文
共 50 条
  • [1] Usefulness of an automated urinary flow cytometer in mass screening for nephritis
    Kaneko, K
    Murakami, M
    Shiraishi, K
    Matsumoto, M
    Yamauchi, K
    Kitagawa, T
    Yamashiro, Y
    PEDIATRIC NEPHROLOGY, 2004, 19 (05) : 499 - 502
  • [2] The validity of the criteria to differentiate the origin of hematuria in the automated urinary flow cytometer
    Hyodo, T
    Kumano, K
    Haga, M
    Sakai, T
    Fukuda, M
    Isami, Y
    Okada, T
    NEPHRON, 1997, 75 (01): : 112 - 112
  • [3] Analysis of urinary red blood cells of healthy individuals by an automated urinary flow cytometer
    Hyodo, T
    Kumano, K
    Haga, M
    Sakai, T
    Fukuda, M
    Isami, Y
    Okada, T
    NEPHRON, 1997, 75 (04): : 451 - 457
  • [4] Asymptomatic hyperleukocyturia in hemodialysis patients analyzed by the automated urinary flow cytometer
    Hyodo, T
    Yoshida, K
    Sakai, T
    Baba, S
    THERAPEUTIC APHERESIS AND DIALYSIS, 2005, 9 (05) : 402 - 406
  • [5] AUTOMATED FLOW CYTOMETER AND DIPSTICKS FOR PREDICTING URINARY TRACT INFECTION IN FEBRILE CHILDREN
    Duong, Hong Phuoc
    Wissing, Karl Martin
    Tram, Nathalie
    Mascart, Georges
    Lepage, Philippe
    Hall, Michelle
    Ismaili, Khalid
    PEDIATRIC NEPHROLOGY, 2015, 30 (09) : 1554 - 1554
  • [6] Evaluation of the Sysmex UF-1000i flow cytometer for screening of urinary tract infection
    de Frutos-Serna, Monica
    Luz Asensio-Calle, Maria
    Maria Haro-Perez, Ana
    Maria Blazquez-de Castro, Ana
    Nieves Gutierrez-Zufiaurre, Maria
    Iglesias-Garcia, Jesus
    ENFERMEDADES INFECCIOSAS Y MICROBIOLOGIA CLINICA, 2014, 32 (03): : 147 - 151
  • [7] Screening for Urinary Tract Infection with the Sysmex UF-1000i Urine Flow Cytometer
    Broeren, Maarten A. C.
    Bahceci, Semiha
    Vader, Huib L.
    Arents, Niek L. A.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2011, 49 (03) : 1025 - 1029
  • [8] Evaluation of the SediMax automated microscopy sediment analyzer and the Sysmex OF-1000i flow cytometer as screening tools to rule out negative urinary tract infections
    Inigo, Melania
    Coello, Andreu
    Fernandez-Rivas, Gema
    Carrasco, Maria
    Marco, Clara
    Fernandez, Anabel
    Casamajor, Teresa
    Ausina, Vicente
    CLINICA CHIMICA ACTA, 2016, 456 : 31 - 35
  • [9] Microfluidic Flow Cytometer for Multiparameter Screening of Fluorophore Photophysics
    Dean, Kevin M.
    Davis, Lloyd M.
    Lubbeck, Jennifer L.
    Manna, Premashis
    Palmer, Amy E.
    Jimenez, Ralph
    2014 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2014,
  • [10] Differential diagnosis between glomerular and nonglomerular hematuria by automated urinary flow cytometer - Kitasato University Kidney Center criteria
    Hyodo, T
    Kumano, K
    Sakai, T
    NEPHRON, 1999, 82 (04): : 312 - 323