Nanomolar detection limit for determination of norepinephrine in the presence of acetaminophen and tryptophan using carbon nanotube-based electrochemical sensor

被引:0
|
作者
Mohammad Mazloum-Ardakani
Mohammad Ali Sheikh-Mohseni
Bibi-Fatemeh Mirjalili
机构
[1] Yazd University,Department of Chemistry, Faculty of Science
来源
Ionics | 2014年 / 20卷
关键词
Nanomolar detection limit; Carbon nanotube; Electrochemical sensor; Norepinephrine; Tryptophan;
D O I
暂无
中图分类号
学科分类号
摘要
A carbon paste electrode modified by carbon nanotubes and a synthesized hydroquinone derivative (abbreviated as DHB) was fabricated. It was used as an electrochemical sensor for simultaneous determination of norepinephrine (NE), acetaminophen (AC), and tryptophan (Trp). Oxidation potential of NE decreased about 220 mV at the modified electrode in comparison with unmodified electrode because of electrocatalysis of oxidation of NE via EĆ mechanism at the modified electrode. Differential pulse voltammetry was used for obtaining the calibration plot of NE and two linear range of 0.2–20.0 μM and 20.0–1,500.0 μM and an interesting detection limit (3σ) of 40.0 nM were obtained for NE. Also, simultaneous determination of NE, AC, and Trp was described by the proposed sensor and linear range of 20.0–800.0 μM was found for AC and Trp. Finally, the electrochemical sensor was used for the determination of NE, AC, and Trp in mixture.
引用
收藏
页码:431 / 437
页数:6
相关论文
共 50 条