High Precision Temperature Measurement for Microfluidic Chip Applications

被引:0
|
作者
Xiong Y. [1 ]
Wu X. [1 ]
Zeng Z. [1 ]
Huang S. [1 ]
Chen T. [1 ]
机构
[1] National Key Laboratory of Science and Technology on Micro/Nano Fabrication, Department of Micro/Nano Electronics, Shanghai Jiao Tong University, Shanghai
关键词
A; fluctuation error; microfluidic chip; proportion method; temperature measurement; TP; 274; two-layer filtering algorithm;
D O I
10.1007/s12204-021-2370-9
中图分类号
学科分类号
摘要
Biochemical reaction in microfluidic chip is sensitive to temperature. Temperature precise control in a small size device requires the temperature measurement with high measurement precision. Traditional temperature measurement method usually measures the voltage drop of the thermistor, which is excited by a constant current source. This method requires the constant current source with high precision and stability. The output of the constant current source is influenced by environmental factors, resulting in a larger measurement error. To solve this problem, a proportion method, a two-layer filtering algorithm, and a power management technique were applied to improve the temperature measurement precision. The proportion method can reduce the low frequency fluctuation error. The two-layer filtering algorithm can reduce the high frequency fluctuation error furtherly. The power management technique used can improve the system stability. Through testing the temperature measurement system built, the experimental results show that the fluctuation error can be significantly decreased from 0.5 ◦C to 0.2◦C. © 2021, Shanghai Jiao Tong University and Springer-Verlag GmbH Germany, part of Springer Nature.
引用
收藏
页码:699 / 705
页数:6
相关论文
共 50 条
  • [31] High precision Pt-resistance temperature measurement system
    Hu, P.-C. (hupc@hit.edu.cn), 1600, Chinese Academy of Sciences (22):
  • [32] High-precision temperature sensor based on weak measurement
    Li, Yanjia
    Li, Hongjing
    Huang, Jingzheng
    Fang, Cheng
    Liu, Miaomiao
    Huang, Chaozheng
    Zeng, Guihua
    OPTICS EXPRESS, 2019, 27 (15): : 21455 - 21462
  • [33] Design of High Precision Temperature Measurement System based on Labview
    Zhu, Weimin
    Liu, Jin
    Yang, Haima
    Yan, Chaochao
    INTERNATIONAL JOURNAL OF ADVANCED COMPUTER SCIENCE AND APPLICATIONS, 2015, 6 (06) : 153 - 155
  • [34] Integratable magnetic shape memory micropump for high-pressure, precision microfluidic applications
    Saren, A.
    Smith, A. R.
    Ullakko, K.
    MICROFLUIDICS AND NANOFLUIDICS, 2018, 22 (04)
  • [35] Integratable magnetic shape memory micropump for high-pressure, precision microfluidic applications
    A. Saren
    A. R. Smith
    K. Ullakko
    Microfluidics and Nanofluidics, 2018, 22
  • [36] High temperature potential of flip chip assemblies for automotive applications
    Braun, T
    Becker, KF
    Sommer, JP
    Löher, T
    Schottenloher, K
    Kohl, R
    Pufall, R
    Bader, V
    Koch, M
    Aschenbrenner, R
    Reichl, H
    55TH ELECTRONIC COMPONENTS & TECHNOLOGY CONFERENCE, VOLS 1 AND 2, 2005 PROCEEDINGS, 2005, : 376 - 383
  • [37] Monitoring Cell Temperature Fluctuation in Microenvironment Chip With a High-precision Microchip
    Zhao Xue-Fei
    Gao Wan-Lei
    Yin Jia-Wen
    Guan Yi-Hua
    Jin Qing-Hui
    PROGRESS IN BIOCHEMISTRY AND BIOPHYSICS, 2021, 48 (08) : 959 - 965
  • [38] Miniature interferometer for refractive index measurement in microfluidic chip
    Chen, Minghui
    Geiser, Martial
    Truffer, Frederic
    Song, Chengli
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS V, 2012, 8553
  • [39] Measurement of density and chemical concentration using a microfluidic chip
    Sparks, D
    Smith, R
    Straayer, M
    Cripe, J
    Schneider, R
    Chimbayo, A
    Anasari, S
    Najafi, N
    LAB ON A CHIP, 2003, 3 (01) : 19 - 21
  • [40] Simultaneous measurement of surface and bilayer tension in a microfluidic chip
    Khangholi, Navid
    Seemann, Ralf
    Fleury, Jean-Baptiste
    BIOMICROFLUIDICS, 2020, 14 (02)