Application of wavelength-scanned wavelength-modulation spectroscopy H2O absorption measurements in an engineering-scale high-pressure coal gasifier

被引:0
|
作者
Kai Sun
Ritobrata Sur
Jay B. Jeffries
Ronald K. Hanson
Tommy Clark
Justin Anthony
Scott Machovec
John Northington
机构
[1] Stanford University,High Temperature Gasdynamics Laboratory
[2] Southern Company Services,National Carbon Capture Center
来源
Applied Physics B | 2014年 / 117卷
关键词
Mole Fraction; Gasification; Syngas; Water Vapor Content; Syngas Product;
D O I
暂无
中图分类号
学科分类号
摘要
A real-time, in situ water vapor (H2O) sensor using a tunable diode laser near 1,352 nm was developed to continuously monitor water vapor in the synthesis gas of an engineering-scale high-pressure coal gasifier. Wavelength-scanned wavelength-modulation spectroscopy with second harmonic detection (WMS-2f) was used to determine the absorption magnitude. The 1f-normalized, WMS-2f signal (WMS-2f/1f) was insensitive to non-absorption transmission losses including beam steering and light scattering by the particulate in the synthesis gas. A fitting strategy was used to simultaneously determine the water vapor mole fraction and the collisional-broadening width of the transition from the scanned 1f-normalized WMS-2f waveform at pressures up to 15 atm, which can be used for large absorbance values. This strategy is analogous to the fitting strategy for wavelength-scanned direct absorption measurements. In a test campaign at the US National Carbon Capture Center, the sensor demonstrated a water vapor detection limit of ~800 ppm (25 Hz bandwidth) at conditions with more than 99.99 % non-absorption transmission losses. Successful unattended monitoring was demonstrated over a 435 h period. Strong correlations between the sensor measurements and transient gasifier operation conditions were observed, demonstrating the capability of laser absorption to monitor the gasification process.
引用
收藏
页码:411 / 421
页数:10
相关论文
共 47 条