Validation of a noninvasive blood pressure monitoring device in normotensive and hypertensive pediatric intensive care patients

被引:3
|
作者
Wankum P.C. [1 ,4 ]
Thurman T.L. [2 ]
Holt S.J. [2 ]
Hall R.A. [3 ]
Simpson P.M. [3 ]
Heulitt M.J. [2 ]
机构
[1] Section of Critical Care Medicine, Department of Child Health, University of Missouri-Columbia, Columbia, MO
[2] Section of Critical Care Medicine, Univ. of Arkansas for Med. Sciences, Little Rock, AR
[3] Section of Biostatistics, Department of Pediatrics, Univ. of Arkansas for Med. Sciences, Little Rock, AR
[4] Critical Care Medicine, University Hospital, Columbia, MO 65212, One Hospital Drive
关键词
Blood pressure; Blood pressure monitors; Monitoring; Oscillometry; Peripheral; Physiologic; Software validation;
D O I
10.1007/s10877-005-1421-3
中图分类号
学科分类号
摘要
Objective: To evaluate the performance and to define limitations of a noninvasive blood pressure monitoring device in the critically ill pediatric population. Method. Patients were included in the study if they were admitted to the Pediatric Intensive Care Unit, were between the ages of 1 month and 18 years with wrist circumferences of ≥ 10 cm, and had an indwelling arterial line. Patients were excluded if their systolic blood pressure differed by ≥ 7.5% between their upper extremities. The measurements were collected simultaneously with those from an arterial line by a computer interfaced with the noninvasive blood pressure monitoring system and the patient's monitor. Heart rates were calculated from the recorded pulse waveforms of the arterial lines. Comparison analyses were performed via bias and precision plots of the blood pressure and heart rate data in addition to calculation of Pearson's correlation coefficients and concordance correlation coefficients. As a nonparametric method of comparison, the proportion of measurements that differed by greater than 10% was calculated. Results. Blood pressures and heart rates of 20 patients between the ages of 12 months and 17 years were monitored by a noninvasive blood pressure monitor for 30 min per patient. This data collection resulted in 2015 data points for each blood pressure and heart rate for comparison of methods. Concordance correlation coefficients were the following: systolic blood pressure, 0.93; diastolic blood pressure, 0.93; mean blood pressure, 0.94; and heart rate, 0.85. Conclusions. The noninvasive blood pressure monitor is capable of producing an accurate blood pressure measurement every 12-15 heartbeats in addition to providing a pulse waveform and digital display of the heart rate. Our study showed good agreement between the methods in the normotensive and hypertensive critically ill pediatric population with a wrist circumference limitation defined at ≥ 11 cm. © Springer Science + Business Media, Inc. 2005.
引用
收藏
页码:253 / 263
页数:10
相关论文
共 50 条
  • [31] Noninvasive Blood Pressure Monitoring in Patients on Chronic Hemodialysis
    Lee, Frank
    Truong, Tai
    Li, Whitney
    Cabush, Abigail
    Chaudhari, Akhil
    Kumar, Parth
    Tantisattamo, Ekamol
    Lau Wei Ling
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2024, 35 (10):
  • [32] Ambulatory Blood Pressure Monitoring in Normotensive Patients with Subclinical Hypothyroidism
    Ferreira, Marcia Martins
    dos Santos Teixeira, Patricia de Fatima
    Mansur, Vera Aleta R.
    Reuters, Vaneska Spinelli
    Almeida, Cloyra Paiva
    Vaisman, Mario
    ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2010, 94 (06) : 806 - 812
  • [33] A standard, noninvasive monitoring of hematocrit algorithm improves blood pressure control in pediatric hemodialysis patients
    Patel, Hiren P.
    Goldstein, Stuart L.
    Mahan, John D.
    Smith, Beth
    Fried, Cheryl B.
    Currier, Helen
    Flynn, Joseph T.
    CLINICAL JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2007, 2 (02): : 252 - 257
  • [34] Validation of Uright model TD 3127AT wireless-portable ambulatory blood pressure monitoring device with timer trigger modification, standard cuff size, in normotensive and mild hypertensive patients of Thailand registry (Thai valid ambulatory blood pressure monitoring)
    Laosanguanek, Wanchat
    Kaewkanha, Ponthakorn
    Khowinthaseth, Sirhavich
    Sirisuth, Sirawee
    Phanthong, Tanattida
    Chokesuwattanaskul, Ronpichai
    Sangwatanaroj, Somkiat
    BLOOD PRESSURE MONITORING, 2022, 27 (06) : 397 - 401
  • [35] USE OF NONINVASIVE AMBULATORY BLOOD-PRESSURE MONITORING TO SCREEN FOR HIGH-RISK HYPERTENSIVE PATIENTS
    CHRISTEN, Y
    GANSLMAYER, M
    WAEBER, B
    BURNIER, M
    NUSSBERGER, J
    BRUNNER, HR
    JOURNAL OF HYPERTENSION, 1990, 8 : S119 - S124
  • [36] Development of a simulator for the validation of noninvasive blood pressure-monitoring devices
    Doh, Il
    Lim, Hyun Kyoon
    Ahn, Bongyoung
    BLOOD PRESSURE MONITORING, 2016, 21 (03) : 189 - 191
  • [37] AMBULATORY BLOOD-PRESSURE MONITORING IN AGE-MATCHED NORMOTENSIVE AND HYPERTENSIVE MEN
    DRAYER, JIM
    WEBER, MA
    CHEST, 1984, 86 (02) : 305 - 305
  • [38] Ambulatory blood pressure monitoring and echocardiography - Noninvasive techniques for evaluation of the hypertensive patient
    Phillips, RA
    Diamond, JA
    PROGRESS IN CARDIOVASCULAR DISEASES, 1999, 41 (06) : 397 - 440
  • [39] Radial artery applanation tonometry for continuous noninvasive arterial blood pressure monitoring in the cardiac intensive care unit
    Nicolas Langwieser
    Luisa Prechtl
    Agnes S. Meidert
    Alexander Hapfelmeier
    Christian Bradaric
    Tareq Ibrahim
    Karl-Ludwig Laugwitz
    Roland M. Schmid
    Julia Y. Wagner
    Bernd Saugel
    Clinical Research in Cardiology, 2015, 104 : 518 - 524
  • [40] Radial artery applanation tonometry for continuous noninvasive arterial blood pressure monitoring in the cardiac intensive care unit
    Langwieser, Nicolas
    Prechtl, Luisa
    Meidert, Agnes S.
    Hapfelmeier, Alexander
    Bradaric, Christian
    Ibrahim, Tareq
    Laugwitz, Karl-Ludwig
    Schmid, Roland M.
    Wagner, Julia Y.
    Saugel, Bernd
    CLINICAL RESEARCH IN CARDIOLOGY, 2015, 104 (06) : 518 - 524