A critical evaluation of the graphite furnace conditions for the direct determination of chromium in urine

被引:0
|
作者
S. P. Quináia
J. A. Nóbrega
机构
[1] Grupo de Química Analítica,
[2] Departamento de Química,undefined
[3] Centro de Ciências Exatas e de Tecnologia,undefined
[4] Universidade Federal de São Carlos,undefined
[5] Caixa Postal 676,undefined
[6] 13565-905 - São Carlos - SP,undefined
[7] Brasil,undefined
关键词
Pyrolysis; Absorption Rate; Tube Wall; Direct Determination; Graphite Furnace;
D O I
暂无
中图分类号
学科分类号
摘要
A critical study of the conditions for the direct determination of Cr in urine using GFAAS was carried out. A chemometric approach showed that the most important parameters that influence the efficiency of atomic Cr cloud formation were the atomization from the tube wall and the proper implementation of a pyrolysis step. The established procedure does not require any sample pre-treatment thus minimizing the risk of contamination. The Cr content in urine was determined using a calibration curve prepared with Cr spiked urine, easily correcting all potential matrix interferences. The heating program proposed took 68 s for a 30 μL urine sample. The procedure was applied for the determination of Cr in urine to investigate the absorption rate of Cr picolinate. The limit of detection and the characteristic mass for a 30 μL urine sample were 0.18 μg/L and 5.4 pg, respectively.
引用
收藏
页码:333 / 337
页数:4
相关论文