Quantification of dopamine transporters in the mouse brain using ultra-high resolution single-photon emission tomography

被引:0
|
作者
Paul D. Acton
Seok-Rye Choi
Karl Plössl
Hank F. Kung
机构
[1] Department of Radiology,
[2] University of Pennsylvania,undefined
[3] 3700 Market Street Room 305,undefined
[4] Philadelphia,undefined
[5] PA 19104,undefined
[6] Department of Pharmacology,undefined
[7] University of Pennsylvania,undefined
[8] Philadelphia,undefined
关键词
Small animal imaging Pinhole SPET Dopamine transporter Quantification Reference tissue model;
D O I
暂无
中图分类号
学科分类号
摘要
Functional imaging of small animals, such as mice and rats, using ultra-high resolution positron emission tomography (PET) and single-photon emission tomography (SPET), is becoming a valuable tool for studying animal models of human disease. While several studies have shown the utility of PET imaging in small animals, few have used SPET in real research applications. In this study we aimed to demonstrate the feasibility of using ultra-high resolution SPET in quantitative studies of dopamine transporters (DAT) in the mouse brain. Four healthy ICR male mice were injected with (mean±SD) 704±154 MBq [99mTc]TRODAT-1, and scanned using an ultra-high resolution SPET system equipped with pinhole collimators (spatial resolution 0.83 mm at 3 cm radius of rotation). Each mouse had two studies, to provide an indication of test-retest reliability. Reference tissue kinetic modeling analysis of the time-activity data in the striatum and cerebellum was used to quantitate the availability of DAT. A simple equilibrium ratio of striatum to cerebellum provided another measure of DAT binding. The SPET imaging results were compared against ex vivo biodistribution data from the striatum and cerebellum. The mean distribution volume ratio (DVR) from the reference tissue kinetic model was 2.17±0.34, with a test-retest reliability of 2.63%±1.67%. The ratio technique gave similar results (DVR=2.03±0.38, test-retest reliability=6.64%±3.86%), and the ex vivo analysis gave DVR=2.32±0.20. Correlations between the kinetic model and the ratio technique (R2=0.86, P<0.001) and the ex vivo data (R2=0.92, P=0.04) were both excellent. This study demonstrated clearly that ultra-high resolution SPET of small animals is capable of accurate, repeatable, and quantitative measures of DAT binding, and should open up the possibility of further studies of cerebral binding sites in mice using pinhole SPET.
引用
收藏
页码:691 / 698
页数:7
相关论文
共 50 条
  • [21] An ultra-high gain single-photon transistor in the microwave regime
    Wang, Zhiling
    Bao, Zenghui
    Li, Yan
    Wu, Yukai
    Cai, Weizhou
    Wang, Weiting
    Han, Xiyue
    Wang, Jiahui
    Song, Yipu
    Sun, Luyan
    Zhang, Hongyi
    Duan, Luming
    NATURE COMMUNICATIONS, 2022, 13 (01)
  • [22] An ultra-high gain single-photon transistor in the microwave regime
    Zhiling Wang
    Zenghui Bao
    Yan Li
    Yukai Wu
    Weizhou Cai
    Weiting Wang
    Xiyue Han
    Jiahui Wang
    Yipu Song
    Luyan Sun
    Hongyi Zhang
    Luming Duan
    Nature Communications, 13
  • [23] Single-photon emission tomography imaging of monoamine transporters in impulsive violent behaviour
    Jari Tiihonen
    Jyrki T. Kuikka
    Kim A. Bergström
    Jari Karhu
    Heimo Viinamäki
    Johannes Lehtonen
    Tero Hallikainen
    Jiwei Yang
    Panu Hakola
    European Journal of Nuclear Medicine, 1997, 24 : 1253 - 1260
  • [24] Radiopharmaceuticals for single-photon emission computed tomography brain imaging
    Kung, HF
    Kung, MP
    Choi, SR
    SEMINARS IN NUCLEAR MEDICINE, 2003, 33 (01) : 2 - 13
  • [25] PET and Single-Photon Emission Computed Tomography in Brain Concussion
    Raji, Cyrus A.
    Henderson, Theodore A.
    NEUROIMAGING CLINICS OF NORTH AMERICA, 2018, 28 (01) : 67 - +
  • [26] Dual-isotope single-photon emission computed tomography for dopamine and serotonin transporters in normal and parkinsonian monkey brains
    Li, I-Hsun
    Huang, Wen-Sheng
    Yeh, Chin-Bin
    Liao, Mei-Hsiu
    Chen, Chia-Chieh
    Shen, Lie-Hang
    Liu, Jiang-Chuan
    Ma, Kuo-Hsing
    NUCLEAR MEDICINE AND BIOLOGY, 2009, 36 (06) : 605 - 611
  • [27] Single-photon emission tomography imaging of serotonin transporters in the nonhuman primate brain with [123I]ODAM
    Paul D. Acton
    Mu Mu
    Karl Plössl
    Catherine Hou
    Michael Siciliano
    Zhi-Ping Zhuang
    Shunichi Oya
    Seok-Rye Choi
    Hank F. Kung
    European Journal of Nuclear Medicine, 1999, 26 : 1359 - 1362
  • [28] Single-photon emission tomography imaging of serotonin transporters in the nonhuman primate brain with [123I]ODAM
    Acton, PD
    Mu, M
    Plössl, K
    Hou, C
    Siciliano, M
    Zhuang, ZP
    Oya, S
    Choi, SR
    Kung, HF
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (10) : 1359 - 1362
  • [29] SENSITIVITY AND RESOLUTION MEASUREMENTS OF 3 SINGLE-PHOTON EMISSION TOMOGRAPHY SYSTEMS
    FLOWER, MA
    KEYES, WI
    ROWE, RW
    BRITISH JOURNAL OF RADIOLOGY, 1980, 53 (634): : 1027 - 1027
  • [30] Positron emission tomography and single-photon emission computed tomography brain imaging in the evaluation of dementia
    Van Heertum, RL
    Tikofsky, RS
    SEMINARS IN NUCLEAR MEDICINE, 2003, 33 (01) : 77 - 85