Electrochemical etching technique for neutron dosimetry

被引:0
|
作者
Ravi Chand Singh
Manwinder Singh
H. S. Virk
机构
[1] Guru Nanak Dev University,Department of Physics
[2] Irrigation & Power Research Institute,Chemistry Division
来源
Indian Journal of Physics | 2009年 / 83卷
关键词
Electrochemical etching; neutron dosimetry; CR-39; CTA; 87.53.Pb; Qc;
D O I
暂无
中图分类号
学科分类号
摘要
In the present work we have employed allyl diglycol carbonate (CR-39) and cellulose triacetate (CTA) plastic for detection of neutron recoil tracks without radiator. For CR-39, the results reveal that registration efficiency is a function of duration of chemical pre-etching and the best results are obtained with chemical pre-etching of 3 hours. It was also investigated that the ac field strength of 28 kV/cm having 2.5 kHz frequency was optimum for revelation of tracks. Interestingly the sensitivity is fluence dependent and it was constant up to a fluence of about 108 n/cm2. The sensitivity abruptly decreased with increased fluence. At optimum experimental conditions the minimum detection limit for CR-39 was found to be 0.47 mSv. For CTA, the tracks have been revealed by electrochemical etching (ECE) only and the minimum detection limit was found to be 0.85 mSv at optimum experimental parameters.
引用
收藏
页码:827 / 832
页数:5