Dose verifications by use of liquid ionization chamber of an electronic portal imaging device (EPID).

被引:0
|
作者
Tateoka K. [1 ]
Oouchi A. [1 ]
Nakata K. [1 ]
Hareyama M. [1 ]
机构
[1] Radiation Oncology, Imaging and Diagnosis, Division of Organ Function and Therapeutics, Molecular and Organ Regulation, Sapporo Medical University Graduate School of Medicine, Sapporo
关键词
EPID; Photon dosimetry;
D O I
10.1007/s12194-008-0024-3
中图分类号
学科分类号
摘要
In this study, we examined the ability of an L-EPID to verify rectangular and irregular fields and to measure the transmitted exit doses. With respect to the beam profile of rectangular and irregular fields and the doses transmitted through an inhomogeneous phantom, the L-EPID dose obtained from the L-EPID measurement was compared with the conventional dose measured by use of a 0.12-cc ionization chamber and a 3D water phantom. In the comparison of the rectangular and irregular fields, the difference in the off-center ratio (OCR) between the L-EPID dose and the conventional dose was approximately 3% in the steep-dose-gradient region (penumbra regions, >30%/cm) and approximately +/-0.5% in the gentle-dose-gradient region (5%/cm). On the other hand, the dose differences between the L-EPID and the measured doses were less than approximately 2% in the gentle-dose-gradient region. In addition, in the steep-dose-gradient region, the maximum difference was 30%. However, the differences in the distance-to-agreement (DTA) were less than approximately +/-1 mm and were unrelated to the dose gradient. These results suggest that dose verification by L-EPID is very useful in clinical applications.
引用
收藏
页码:171 / 177
页数:6
相关论文
共 50 条
  • [1] Using an electronic portal imaging device (EPID) for routine geometrical LINAC verifications
    Bejarano, M. Benito
    Beltran, F. Saez
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S253 - S254
  • [2] PORTAL DOSE MEASUREMENT IN RADIOTHERAPY USING AN ELECTRONIC PORTAL IMAGING DEVICE (EPID)
    HEIJMEN, BJM
    PASMA, KL
    KROONWIJK, M
    ALTHOF, VGM
    DEBOER, JCJ
    VISSER, AG
    HUIZENGA, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11): : 1943 - 1955
  • [3] The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device
    Mohammadi, Mohammad
    Bezak, Eva
    Reich, Paul
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (01): : 69 - 84
  • [4] Radiological thickness measurement using a liquid ionization chamber electronic portal imaging device
    Evans, PM
    Donovan, EM
    Partridge, M
    Bidmead, AM
    Garton, A
    Mubata, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (06): : N89 - N97
  • [5] Evaluation of rapid dose map acquisition of a scanning liquid-filled ionization chamber electronic portal imaging device
    Chang, JW
    Mageras, GS
    Ling, CC
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05): : 1432 - 1445
  • [6] Relative dosimetry using a scanning liquid-filled ion chamber (SLIC) electronic portal imaging device (EPID)
    Ortega, JFC
    Seoane, A
    Garrido, L
    Casals, J
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S415 - S415
  • [7] Transit Dosimetry for Verification of Dose Delivery Using Electronic Portal Imaging Device (EPID)
    Baek, T.
    Chung, E.
    Lee, S.
    Yoon, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 301 - 301
  • [8] Evaluation of quality assurance phantom for electronic portal imaging device (EPID)
    Das, I.
    Cheng, C.
    Chopra, K.
    Misic, V.
    Schule, E.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2371 - 2372
  • [9] In Vivo 3D Dose Verification for IMRT Using Electronic Portal Imaging Device (EPID)
    Chang, C.
    Anderson, N. G.
    Zhu, T. C.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Accuracy of the dose delivery in prostate cancer patients-using an electronic portal imaging device (EPID)
    Mahdavi, S. R.
    Gharehbagh, E. Jazayeri
    Mofid, B.
    Jafari, A. H.
    Nikoofar, A. R.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2017, 15 (01): : 39 - 47