A new dynamic myocardial phantom for the assessment of left ventricular function by gated single-photon emission tomography

被引:0
|
作者
Naoki Kubo
Koichi Morita
Chietsugu Katoh
Tohru Shiga
Masanori Konno
Eriko Tsukamoto
Yutaka Morita
Nagara Tamaki
机构
[1] Department of Radiological Technology,
[2] College of Medical Technology,undefined
[3] Hokkaido University,undefined
[4] Sapporo,undefined
[5] Department of Nuclear Medicine,undefined
[6] School of Medicine,undefined
[7] Hokkaido University,undefined
[8] Sapporo,undefined
[9] Department of Tracer Kinetics,undefined
[10] Hokkaido University,undefined
[11] Sapporo,undefined
[12] Correspondence to: Department of Nuclear Medicine,undefined
[13] School of Medicine,undefined
[14] Hokkaido University,undefined
[15] Kita-15,undefined
[16] Nishi-7,undefined
[17] Kita-ku,undefined
[18] Sapporo 060–8638,undefined
来源
关键词
Myocardium Phantom Wall motion Single-photon emission tomography;
D O I
暂无
中图分类号
学科分类号
摘要
Gated myocardial perfusion single-photon emission tomography (SPET) has been used for the measurement of left ventricular (LV) function and validated by means of comparison with other imaging modalities. We have designed a new dynamic myocardial phantom in order to validate the LV function as assessed by the use of gated myocardial perfusion SPET. The phantom consists of two half-ellipsoids (an endocardial surface and an epicardial surface) and a thorax. The myocardial space is filled with a radioactive solution. The endocardial surface moves continuously towards and away from the epicardial surface in the longitudinal axis to vary the LV volume [143 ml at end-diastole (ED), 107 ml at end-systole (ES)] and thickness (apex 8 mm at ED and 26 mm at ES, midplane 8 mm). The mean values of wall motion (WM) for the apical midplane region and the basal midplane region were 5 mm and 2 mm, respectively. Gated myocardial SPET was performed during 8 and 16 intervals. These projection data sets were processed using a Butterworth filter with an order of 5 and a critical frequency of 0.34 cycles/cm. LV function was calculated using the quantitative gated SPET (QGS) algorithm. The LV function values estimated by gated SPET during 16 intervals [22% for ejection fraction (EF), 3.7 mm for WM of the apical midplane, 1.7 mm for WM of the basal midplane] closely resembled actual LV functions [25% for EF, 5 mm for WM of the apical midplane, 2 mm for WM of the basal midplane]. However, the estimated values during 8 intervals were smaller than those during 16 intervals (19% for EF, 3.3 mm for WM of the apical-midplane, 1.1 mm for WM of the basal-midplane). The estimated LV volumes closely correlated with the actual volumes (r=0.99 for 16 intervals, r=0.95 for 8 intervals). Utilizing this phantom, LV function estimated using gated myocardial SPET can be compared with actual values.
引用
收藏
页码:1525 / 1530
页数:5
相关论文
共 50 条
  • [1] A new dynamic myocardial phantom for the assessment of left ventricular function by gated single-photon emission tomography
    Kubo, N
    Morita, K
    Katoh, C
    Shiga, T
    Konno, M
    Tsukamoto, E
    Morita, Y
    Tamaki, N
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (10) : 1525 - 1530
  • [2] QUANTITATIVE ASSESSMENT OF REGIONAL LEFT-VENTRICULAR FUNCTION BY GATED MYOCARDIAL SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY
    QURESHI, U
    HE, ZX
    VADUGANATHAN, P
    MAHMARIAN, JJ
    VERANI, MS
    CIRCULATION, 1995, 92 (08) : 2144 - 2144
  • [3] Assessment of left ventricular diastolic function by gated single-photon emission tomography: comparison with Doppler echocardiography
    Tetsuhiro Yamano
    Tomoki Nakamura
    Kenzo Sakamoto
    Takato Hikosaka
    Kan Zen
    Takeshi Nakamura
    Takahisa Sawada
    Akihiro Azuma
    Tsunehiko Nishimura
    Masao Nakagawa
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 1532 - 1537
  • [4] Assessment of left ventricular diastolic function by gated single-photon emission tomography: comparison with Doppler echocardiography
    Yamano, T
    Nakamura, T
    Sakamoto, K
    Hikosaka, T
    Zen, K
    Nakamura, T
    Sawada, T
    Azuma, A
    Nishimura, T
    Nakagawa, M
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (11) : 1532 - 1537
  • [5] Assessment of left ventricular mechanical dyssynchrony using electrocardiographically-gated myocardial perfusion single-photon emission computed tomography
    Niimi, T.
    Maeda, H.
    Nanasato, M.
    Sugimoto, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S335 - S335
  • [6] Validation of left ventricular function from gated single photon computed emission tomography by using a scintillator-photodiode camera: a dynamic myocardial phantom study
    Kubo, N
    Mabuchi, M
    Katoh, C
    Arai, H
    Morita, K
    Tsukamoto, E
    Morita, Y
    Tamaki, N
    NUCLEAR MEDICINE COMMUNICATIONS, 2002, 23 (07) : 639 - 643
  • [7] Single-photon emission computed tomography for the assessment of ventricular perfusion and function
    Gonzalez, Patricio
    Dussaillant, Gaston
    Gutierrez, Daniela
    Berrocali, Isabel
    Alay, Rita
    Otarola, Sonia
    REVISTA MEDICA DE CHILE, 2013, 141 (09) : 1136 - 1142
  • [8] Evaluation of left ventricular function and volume in patients with dilated cardiomyopathy: gated myocardial single-photon emission tomography (SPECT) versus echocardiography
    Berk, F
    Isgoren, S
    Demir, H
    Kozdag, G
    Ural, D
    Komsuoglu, B
    ANNALS OF SAUDI MEDICINE, 2005, 25 (03) : 198 - 204
  • [9] Myocardial stunning-induced left ventricular dyssynchrony on gated single-photon emission computed tomography myocardial perfusion imaging
    Jiang, Zhixin
    Tang, Haipeng
    Shi, Jianzhou
    Zhou, Yanli
    Wang, Cheng
    Li, Dianfu
    Shan, Qijun
    Zhou, Weihua
    NUCLEAR MEDICINE COMMUNICATIONS, 2018, 39 (08) : 725 - 731
  • [10] Limited performance of quantitative assessment of myocardial function by thallium-201 gated myocardial single-photon emission tomography
    Lee, Dong Soo
    Ahn, Ji Young
    Kim, Seok Ki
    Oh, Byung Hee
    Seo, Jung Don
    Chung, June-Key
    Lee, Myung Chul
    European Journal of Nuclear Medicine and Molecular Imaging, 2000, 27 (02) : 185 - 191