Twisting and constriction of a HeartMate II driveline in the area of a repair site

被引:0
|
作者
Sven Maier
Friedhelm Beyersdorf
Christoph Benk
机构
[1] Faculty of Medicine,Department of Cardiovascular Surgery, Heart Center Freiburg University
[2] University of Freiburg,undefined
来源
Journal of Artificial Organs | 2020年 / 23卷
关键词
Left ventricular assist device; Driveline repair; Driveline exchange;
D O I
暂无
中图分类号
学科分类号
摘要
The driveline’s durability is crucial for optimal long-term support with a left-ventricular assist device (LVAD). The incidence of percutaneous driveline fracture after HeartMate II LVAD implantation is low. For the first time, we describe a patient with an already repaired driveline and a massive constriction and twisting of the driveline in the area of the repair site. This dramatic finding necessitates a renewed exchange of the external part of the driveline by the manufacturer. Due to the increasing number of patients with elongated LVAD support, the stability of the driveline and possible repairs including the replacement of the driveline are becoming more and more important. Our case report describes a possible serious late complication after replacement of the driveline, shows possible risks for this development, and describes the necessity of a prophylactic X-ray examination of repaired drivelines to detect such complications as early as possible.
引用
收藏
页码:371 / 373
页数:2
相关论文
共 50 条
  • [1] Twisting and constriction of a HeartMate II driveline in the area of a repair site
    Maier, Sven
    Beyersdorf, Friedhelm
    Benk, Christoph
    JOURNAL OF ARTIFICIAL ORGANS, 2020, 23 (04) : 371 - 373
  • [2] Correction to: Twisting and constriction of a HeartMate II driveline in the area of a repair site
    Sven Maier
    Friedhelm Beyersdorf
    Christoph Benk
    Journal of Artificial Organs, 2021, 24 : 418 - 418
  • [3] Twisting and constriction of a HeartMate II driveline in the area of a repair site (vol 23, pg 371, 2020)
    Maier, Sven
    Beyersdorf, Friedhelm
    Benk, Christoph
    JOURNAL OF ARTIFICIAL ORGANS, 2021, 24 (03) : 418 - 418
  • [4] Driveline Angle Is Crucial to Prevent Exit Site Infection in Patients with HeartMate II
    Matsumoto, E.
    Fujita, T.
    Hata, H.
    Shimahara, Y.
    Nakajima, S.
    Sato, T.
    Seguchi, O.
    Yanase, M.
    Fukushima, N.
    Nakatani, T.
    Kobayashi, J.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (04): : S256 - S257
  • [5] Outcomes of modular cable exchange in HeartMate 3 patients versus conventional driveline repair in HeartMate II patients
    Merzah, Ali Saad
    Hanke, Jasmin S.
    Li, Tong
    Merzah, Fadi A. Saad
    Gabriel, Maria
    Derda, Anselm A.
    Homann, Katharina
    Haverich, Axel
    Schmitto, Jan D.
    Dogan, Gunes
    ARTIFICIAL ORGANS, 2023, 47 (02) : 380 - 386
  • [6] Percutaneous Driveline Fracture After Implantation of the HeartMate II Left Ventricular Assist Device: How Durable is Driveline Repair?
    Stulak, John M.
    Schettle, Sarah
    Haglund, Nicholas
    Dunlay, Shannon
    Cowger, Jennifer
    Shah, Palak
    Aaronson, Keith D.
    Pagani, Francis D.
    Maltais, Simon
    ASAIO JOURNAL, 2017, 63 (05) : 542 - 545
  • [7] Driveline angle is crucial for preventing driveline infection in patients with HeartMate II device
    Matsumoto, Yorihiko
    Fukushima, Satsuki
    Shimahara, Yusuke
    Yamashita, Kizuku
    Kawamoto, Naonori
    Kuroda, Kensuke
    Seguchi, Osamu
    Yanase, Masanobu
    Fukushima, Norihide
    Shimizu, Hideyuki
    Kobayashi, Junjiro
    Fujita, Tomoyuki
    JOURNAL OF ARTIFICIAL ORGANS, 2019, 22 (01) : 37 - 43
  • [8] Driveline angle is crucial for preventing driveline infection in patients with HeartMate II device
    Yorihiko Matsumoto
    Satsuki Fukushima
    Yusuke Shimahara
    Kizuku Yamashita
    Naonori Kawamoto
    Kensuke Kuroda
    Osamu Seguchi
    Masanobu Yanase
    Norihide Fukushima
    Hideyuki Shimizu
    Junjiro Kobayashi
    Tomoyuki Fujita
    Journal of Artificial Organs, 2019, 22 : 37 - 43
  • [9] Impact of an improved driveline management for HeartMate II and HeartMate 3 left ventricular assist devices
    Mutsuga, Masato
    Okumura, Takahiro
    Morimoto, Ryota
    Kondo, Toru
    Ito, Hideki
    Terazawa, Sachie
    Tokuda, Yoshiyuki
    Narita, Yuji
    Nishida, Kazuki
    Murohara, Toyoaki
    Usui, Akihiko
    ARTIFICIAL ORGANS, 2023, 47 (02) : 387 - 395
  • [10] Driveline infections in patients supported with a HeartMate II: Incidence, aetiology and outcome
    Bomholt, Tobias
    Moser, Claus
    Sander, Kaare
    Boesgaard, Soren
    Kober, Lars
    Olsen, Peter Skov
    Hansen, Peter Bo
    Mortensen, Svend-Aage
    Gustafsson, Finn
    SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2011, 45 (05) : 273 - 278